Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K014153
    Date Cleared
    2002-03-15

    (87 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    COAPT SYSTMES ENDOTINE DEVICE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Endotine Device is intended for use in browplasty surgery. The Endotine device is specifically intended for use to fixate the sub-dermis to the cranial bone in browplasty procedures.

    Device Description

    The Endotine Device is a device fabricated from L-PLA that is intended for use in browplasty procedures. The device is molded with a post that is inserted directly into the cranial bone. A series of barbs along its face provides a feature that allows for the direct fixation the sub-dermis without the use of suture. The device is inserted into the bone with the use of an installation instrument. Additional instruments include a stepped drill and sterilization tray.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Endotine Device's acceptance criteria and studies:

    The provided document is a 510(k) summary for the Coapt Systems Endotine Device, seeking substantial equivalence to a predicate device. It focuses on the device's design, intended use, and a high-level overview of performance data. This type of document typically does not contain detailed acceptance criteria or extensive study results as would be found in a full study report or clinical trial.

    Based on the provided text, the following information can be extracted or inferred:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative acceptance criteria for shear strength or adequacy. Instead, it relies on a comparison to the predicate device.

    Acceptance CriterionReported Device Performance
    Bench Testing: Comparison of shear strength in-vitro to predicate deviceDemonstrated "Comparison of the shear strength of the Endotine Device In-Vitro compared to the predicate device." Implies the Endotine Device met or exceeded the predicate's shear strength, or was found to be comparable.
    Animal Testing: Adequacy for intended use"The testing demonstrated the acceptability of the Endotine Device and confirmed that the device functions adequately to meet its intended use."
    Tissue Interface: Comparison of device-to-tissue interface to predicate device"Comparison of the device to tissue interface of the Endotine Device to the predicate device." Implies comparability or acceptable performance.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes for the bench testing or animal testing.

    • Data Provenance: Not explicitly stated, but "In-Vitro" implies laboratory testing, and "Animal Testing" refers to an animal model. The country of origin for the data is not mentioned. The nature of these tests (bench, animal) suggests they were prospective studies designed to evaluate the device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This type of information is generally not applicable to bench or animal testing in the context of device performance directly related to mechanical properties or biological response in an animal model. Ground truth for these tests would be established through scientific measurements and observations, not expert consensus on human data.

    4. Adjudication Method for the Test Set

    Not applicable for bench or animal testing of mechanical properties and biological response.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done, nor would it be relevant for this type of device and the performance data presented. MRMC studies are typically used for diagnostic devices involving human interpretation of images or other data.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

    Not applicable. This device is a bioabsorbable fixation device, not an algorithm or AI system.

    7. Type of Ground Truth Used

    • Bench Testing: The ground truth would be precise engineering measurements of shear strength under defined conditions.
    • Animal Testing: The ground truth would be a combination of direct observation of device function, histological analysis of tissue response, and possibly mechanical testing of the fixation in the animal model.

    8. Sample Size for the Training Set

    Not applicable. This device is not an AI/ML algorithm that requires a "training set." The materials and design are established through engineering principles.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1