Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K191296
    Manufacturer
    Date Cleared
    2020-08-11

    (455 days)

    Product Code
    Regulation Number
    862.1215
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Pointe Scientific Creatinine Kinase (CK) Reagent Set

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the quantitative determination of creatine kinase activity in serum and plasma. Rx only.

    Measurements of Creatine Kinase are used in the diagnosis and treatment of myocardial infaction and muscle disease, such as progressive Duchenne-type muscular dystrophy.

    Device Description

    The Pointe Scientific Creatine Kinase (CK) Reagent Set consists of ready-to-use liguid reagents:

    • . CK R1 (buffer) contains: Imidazole buffer (pH 6.7) 100.0 mmol/L; NADP 2.0 mmol/L: HK (Baker's yeast) 2.5 KU/L: Glucose 20.0 mmol/L: Magnesium Acetate 10.0 mmol/L; EDTA 2.0 mmol/L and N-acetylcysteine (NAC) 20.0 mmol/L.
    • . CK R2 (enzyme reagent) contains: Imidazole buffer (pH 6.7) 100.0 mmol/L: ADP 2.0 mmol/L: AMP 5.0 mmol/L: Diadensosine pentaphosphate 10.0 mmol/L: Creatine phosphate 30.0 mmol/L; G6PDH (Baker's yeast) 1.5 KU/L and EDTA 2.0 mmol/L.

    The kinetic procedure presented is a modification of Szasz of the Rosalki technique, which optimizes the reaction by reactivation of CK activity with N-actyl-L-cysteine (NAC).

    Creatine Kinase specifically catalyzes the transphosphorylation of ADP to ATP. Through a series of coupled enzymatic reactions, NADPH is produced at a rate directly proportional to the CK activity. The method determines the NADPH absorbance increase per min at 340 nm.

    AI/ML Overview

    The provided document describes the analytical performance studies for the Pointe Scientific Creatine Kinase (CK) Reagent Set, which supports its substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and study details:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state "acceptance criteria" in a separate table for each test. Instead, it presents the study results and implies that these results were considered acceptable for demonstrating substantial equivalence. For some tests, like Linearity, an acceptable deviation is stated.

    However, based on the provided performance data, we can infer the acceptance criteria that the manufacturer likely aimed for to support their claims.

    Performance StudyImplied Acceptance Criteria (Inferred from results/general practice)Reported Device Performance (Pointe Scientific Creatine Kinase (CK) Reagent Set)
    Method Comparison (Serum vs. Predicate)High correlation (e.g., R > 0.975), acceptable biasCorrelation Coefficient: 0.9991, Regression: y = 1.041x - 5.2
    Method Comparison (Plasma vs. Predicate)High correlation (e.g., R > 0.975), acceptable biasCorrelation Coefficient: 0.9946, Regression: y = 1.032x - 0.4
    Precision (Controls & Patient Samples - Serum/Plasma)Low CV% for repeatability and total precision (typically 10% from control) up to clinically relevant concentrationsBilirubin, Ascorbic Acid, Hemoglobin, Intralipid showed no significant interference at high concentrations (e.g., Bilirubin 60 mg/dL, Hemoglobin 500 mg/dL)

    2. Sample size used for the test set and the data provenance:

    • Method Comparison (Serum):
      • Sample Size: 120 de-identified remnant serum samples. 4 samples were altered by mixing.
      • Data Provenance: Obtained from a commercial repository. Retrospective. Country of origin not specified, but likely within the US given the FDA submission.
    • Method Comparison (Plasma):
      • Sample Size: 123 de-identified remnant plasma samples. 2 samples were altered by mixing.
      • Data Provenance: Obtained from a commercial repository. Retrospective. Country of origin not specified.
    • Precision Studies:
      • Sample Size: 2 commercial quality controls, 3 serum pools, and 3 plasma pools. Each pool/control was tested in duplicate twice per day for 20 days (n=80 per sample type).
      • Data Provenance: Not explicitly stated, but common practice is for manufacturers to prepare pools or use commercially available controls.
    • Linearity/Assay Range Study:
      • Sample Size: A set of 12 serum samples and 12 plasma samples (prepared by admixture of high-level and low-level pools).
      • Data Provenance: Not explicitly stated, likely prepared in-house or from common available resources.
    • Detection Capability (LoB, LoD, LoQ):
      • LoB: Saline (1 sample * 20 repetitions * 3 days).
      • LoD: 20 low-level depleted serum samples and 20 depleted plasma samples.
      • LoQ: 5 low activity samples (each run in 8 duplicates over 5 days).
      • Data Provenance: Not explicitly stated, likely prepared in-house or from common available resources.
    • Dilution Recovery Studies:
      • Sample Size: Three contrived high-level samples for both serum and plasma.
      • Data Provenance: Not explicitly stated, likely prepared in-house.
    • Analytical Specificity (Endogenous Substances):
      • Sample Size: Not explicitly stated, but interference testing was conducted using "randomly selected serum and plasma samples ranging from 43 U/L to 268 U/L."
      • Data Provenance: Not explicitly stated.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is a diagnostic reagent for quantitative determination of an analyte (Creatine Kinase), not an imaging or qualitative diagnostic device requiring expert interpretation of results to establish ground truth.

    • Ground Truth for Method Comparison: The predicate device's results (Beckman Coulter Creatine Kinase (CK-Nac) on the Beckman Coulter Olympus AU400 Clinical Chemistry Analyzer) serve as the "reference" or "ground truth" for comparison. This is a common method for demonstrating substantial equivalence for in vitro diagnostic (IVD) devices. No human experts are used to establish ground truth in this context; rather, the established method of the predicate device is the reference.
    • Ground Truth for other studies (Precision, Linearity, Detection, Dilution Recovery, Specificity): The "ground truth" is typically established by the analytical method itself, or by preparing samples with known concentrations/characteristics, without the need for human expert consensus.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    No adjudication method is relevant or mentioned as this device measures an objective biochemical marker. The 'ground truth' is the quantitative measurement itself or the reference method's result.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No. This is not applicable. The device is a diagnostic reagent for quantitative measurement of creatine kinase, not an AI-powered diagnostic system for image interpretation or a device requiring human readers/scorers. Therefore, MRMC studies are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device is a reagent set used on an automated chemistry analyzer (Shenzhen Mindray BA-800M Chemistry Analyzer). The performance studies presented are for the "algorithm only" in the sense that they demonstrate the analytical performance of the reagent on the analyzer without human interpretation of the raw data to generate the numerical result. The "human-in-the-loop" would be a lab technician performing the test and reviewing the results, but the analytical performance itself is standalone.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Method Comparison: The results obtained from the legally marketed predicate device (Beckman Coulter Creatine Kinase (CK-Nac) on the Beckman Coulter Olympus AU400 Clinical Chemistry Analyzer) served as the reference for comparison.
    • Precision, Linearity, Detection, Dilution Recovery, Analytical Specificity: Ground truth is based on the inherent analytical properties of the method/reagent including samples with known concentrations (e.g., prepared standards, spiked samples, qualified controls, or established reference material). Linearity was assessed against expected values based on known admixtures. Detection limits were determined using statistical methods on low-level and blank samples.

    8. The sample size for the training set:

    There is no "training set" in the context of this 510(k) submission for a diagnostic reagent. This device is not an AI/machine learning algorithm that requires training data. The studies performed are analytical validation studies.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this type of medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141265
    Manufacturer
    Date Cleared
    2014-07-18

    (64 days)

    Product Code
    Regulation Number
    862.1215
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ELITECH CLINICAL SYSTEMS ENVOY 500 CK REAGENT KIT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ENVOY®500 CK REAGENT KIT is intended for the quantitative in vitro determination of creatine kinase (CK) in buman serum and plasma using the ENVOY 500 Chemistry System.

    It is not intended for use in Point of Care settings.

    Creatine phosphokinase and its isoenzymes measurements are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.

    Device Description

    ENVOY CK REAGENT KIT is available as kit only. It consists of a bi-reagent R1 and R2 whose composition,

    for R1: 125 mmol/L Imidazole buffer, pH 6.10; 25 mmol/L D-Glucose; 25 mmol/L N-Acetyl-L-Cysteine; 12.5 mmol/L Magnesium acetate; 2.4 mmol/L NADP; 2.0 mmol/L EDTA; > 6800 U/L Hexokinase (microorganism);

    AI/ML Overview

    This document describes the performance characteristics and acceptance criteria for the ENVOY 500 CK REAGENT KIT, an in vitro diagnostic device for the quantitative determination of creatine kinase (CK) in human serum and plasma.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance CriteriaReported Device Performance
    Precision
    Within-run CV% (Level 1)Not explicitly stated (implied to be low)1.5%
    Within-run CV% (Level 2)Not explicitly stated (implied to be low)1.0%
    Within-run CV% (Level 3)Not explicitly stated (implied to be low)1.4%
    Total CV% (Level 1)Not explicitly stated (implied to be low)3.6%
    Total CV% (Level 2)Not explicitly stated (implied to be low)3.5%
    Total CV% (Level 3)Not explicitly stated (implied to be low)3.6%
    Linearity/Assay RangeAcceptable deviation from linearity of ±10%Linear from 10-1714 U/L (with automatic dilution up to 17140 U/L)
    Limit of Detection (LoD)Not explicitly stated (implied to be low for clinical utility)2 U/L
    Limit of Quantification (LoQ)Precision coefficient of variation of ≤ 15%5 U/L (with ≤ 15% CV)
    InterferenceAccepted bias of ±10% in sample pools with low (150 U/L) or high (1200 U/L) nominal activityNo significant interference for specific interferents up to tested concentrations (e.g., Hemoglobin up to 100 mg/dL, Triglycerides up to 3000 mg/dL, Bilirubin up to 30 mg/dL, Ascorbic acid up to 20 mg/dL, Acetylsalicylic acid up to 200 mg/dL, Acetaminophen up to 30 mg/dL)
    Method Comparison (Serum)Not explicitly stated (implied high correlation and agreement with predicate)y = 1.050x + 0 U/L, r = 0.998, Sy.x = 28 U/L (range: 14 to 1650 U/L)
    Method Comparison (Lithium Heparin Plasma)Not explicitly stated (implied high correlation and agreement with predicate)y = 1.020x + 3 U/L, r = 0.999, Sy.x = 21 U/L (range: 10 to 1660 U/L)
    On Board StabilityNot explicitly stated (implied to be sufficient for practical use)28 days
    Real-time (Shelf) StabilityStable until the expiry date stated on the labelFollowed for 14 months on 3 different lots

    2. Sample Size Used for the Test Set and the Data Provenance

    • Precision Test Set: 80 measurements for each of 3 levels of samples (Level 1, Level 2, Level 3). Each level was measured two times per run, for two runs per day, for twenty operating days, on two instruments.
    • Linearity Test Set: 11 levels of patient pools (for serum).
    • Detection Limit Test Set: 15 measurements of 4 samples for LoD; 15 measurements of 4 samples for LoQ.
    • Interference Test Set: For each potential interferent, 2 serum sample pools (a low activity pool at 150 U/L and a high activity pool at 1200 U/L) were used. Aliquots of each pool were spiked with increasing interferent concentrations (e.g., 9 concentrations for triglycerides, 7 for unconjugated bilirubin, etc.). Each point was measured in triplicate per run.
    • Method Comparison (Serum) Test Set: 100 serum patient samples.
    • Method Comparison (Lithium Heparin Plasma) Test Set: 40 plasma specimens (in lithium heparin).
    • On Board Stability Test Set: At least 3 levels of sample (high/medium/low) were tested in duplicate at Day 0, and 4 activity levels were analyzed in duplicate for at least 30 days.

    Data Provenance: The document does not explicitly state the country of origin for the patient samples or if the data was retrospective or prospective. Given the submitter is ELITech Clinical Systems SAS, France, and the testing was conducted for regulatory submission in the US, it is plausible the studies were conducted in Europe or at contract research organizations. The nature of the studies (e.g., precision, linearity, interference, method comparison) generally involves prospective collection or commercially available control/patient samples for analytical validation.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This document describes the analytical performance of an in-vitro diagnostic reagent kit (ENVOY 500 CK REAGENT KIT). For such devices, "ground truth" generally refers to reference methods or measurements by highly calibrated instruments, not expert human interpretation. Therefore, the concept of "number of experts used to establish ground truth" and their qualifications is not applicable in this context. The ground truth for analytical studies is established by:

    • Reference materials or calibrators traceable to international standards (e.g., IFCC method for CK).
    • Predicate devices or established methods for method comparison studies.
    • Precise gravimetric or volumetric preparations for linearity and spiking studies.

    4. Adjudication Method for the Test Set

    The concept of an "adjudication method" (e.g., 2+1, 3+1) is typically relevant for studies involving subjective human interpretation of diagnostic images or data, where discrepancies between readers need to be resolved to establish ground truth for algorithm training or testing.

    For the analytical performance studies described for the ENVOY 500 CK Reagent Kit, no such adjudication method was used or is applicable. The measurements are quantitative and objective, following established laboratory protocols and guidelines (e.g., CLSI protocols). Discrepancies would be resolved through re-testing, investigation of instrument or reagent issues, or statistical analysis (e.g., outliers).


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document describes an in-vitro diagnostic reagent kit, not an AI-based diagnostic device intended for human reader assistance. Therefore, no MRMC comparative effectiveness study involving human readers or AI assistance was performed.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This document describes the performance of an in-vitro diagnostic reagent kit that functions on a chemistry analyzer (Envoy 500 Chemistry System). The device itself is the reagent kit, which works in conjunction with the instrument to provide quantitative results. It is not an algorithm or AI-based device. The performance described is inherently "standalone" in the sense that it measures the analytical capability of the reagent-instrument system to quantify CK levels without human intervention in the measurement process itself, beyond sample loading and system operation. No human-in-the-loop performance is relevant for the analytical measurement part of the device's function.


    7. The Type of Ground Truth Used

    The ground truth for the various analytical performance characteristics was established using:

    • Reference Methods/Standards: For traceability, the calibration factor for the ENVOY 500 CK Reagent Kit has traceability to the IFCC method (International Federation of Clinical Chemistry and Laboratory Medicine) recommendations for CK activity determination.
    • Predicate Device Comparison: For method comparison, the predicate device (ELITech Clinical Systems CK NAC SL on Selectra ProM analyzer) served as the reference for comparison using patient samples.
    • Prepared Samples/Known Concentrations:
      • For linearity, patient pools were prepared by spiking a serum pool and dilution to obtain 11 levels with equidistant activities (known relative concentrations).
      • For detection and quantification limits, samples were prepared by diluting patient samples to obtain specific activities (e.g., approximately 3.5 U/L for LoD, 5 U/L for LoQ).
      • For interference, serum sample pools were spiked with increasing concentrations of known interferents against control samples.
    • Statistical Analysis: Precision studies rely on repeated measurements to quantify variability, with "ground truth" derived from the calculated mean and statistical metrics.

    In summary, the ground truth is primarily based on reference methods (IFCC), comparison to a legally marketed predicate device, and precisely prepared samples with known (or highly characterized) concentrations/activities.


    8. The Sample Size for the Training Set

    Not applicable. This device is an in-vitro diagnostic reagent kit, not a machine learning or artificial intelligence-based device that requires a "training set" in the conventional sense. The "training" of such a system involves chemical formulation, optimization of reaction conditions, and calibration procedures using calibrators or reference materials.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As stated above, this device does not utilize a "training set" in the context of machine learning. The "ground truth" for developing and optimizing the reagent kit and its associated methods would involve:

    • Chemical principles and stoichiometry: The understanding of the enzyme kinetics and chemical reactions being measured (e.g., the IFCC recommended method for CK).
    • Reference materials and calibrators: Used to ensure accuracy and traceability of the quantitative measurements.
    • Experimental optimization: Through iterative testing and modification of reagent concentrations, pH, and reaction conditions to achieve optimal performance characteristics (sensitivity, specificity, stability, etc.).
    Ask a Question

    Ask a specific question about this device

    K Number
    K123322
    Date Cleared
    2013-05-21

    (207 days)

    Product Code
    Regulation Number
    862.1770
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ACE BUN/UREA REAGENT, ACE CREATININE REAGENT, ACE URIC ACID REAGENT, ACE CK REAGENT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE BUN/Urea Reagent is intended for the quantitative determination of blood urea nitrogen (BUN) concentration in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. BUN measurements are used in the diagnosis and treatment of certain renal and metabolic diseases. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Creatinine Reagent is intended for the quantitative determination of creatinine concentration in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Uric Acid Reagent is intended for the quantitative determination of uric acid concentration in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Uric acid measurements are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions and of patients receiving cytotoxic drugs. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE CK Reagent is intended for the quantitative determination of creatine kinase activity in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Measurement of creatine kinase is used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Device Description

    In the ACE BUN/Urea Reagent assay, urea in serum is hydrolyzed in the presence of urease to yield ammonia and carbon dioxide. The ammonia formed then reacts in the presence of glutamate dehydrogenase with 2-oxoglutarate and NADH to yield glutamate and NAD. NADH absorbs strongly at 340 nm, whereas NAD+ does not. The initial rate of decrease in absorbance, monitored bichromatically at 340 nm/647 nm, is proportional to the urea concentration in the sample.

    In the ACE Creatinine Reagent assay, creatinine reacts with picric acid in an alkaline medium to form a red-orange colored complex, which absorbs strongly at 505 nm. The rate of complex formation, determined by measuring the increase in absorbance bichromatically at 505 nm/573 nm during a fixed time interval, is directly proportional to the creatinine concentration in the sample.

    In the ACE Uric Acid Reagent assay, uric acid in serum is oxidized by uricase to allantoin and hydrogen peroxide. The hydrogen peroxide then acts to oxidatively couple dichlorohydroxybenzene sulfonic acid and 4-aminoantipyrine in a reaction catalyzed by peroxidase, producing a red colored quinoneimine complex, which absorbs strongly at 505 nm. The amount of chromogen formed is determined by measuring the increase in absorbance bichromatically at 505 nm/610 nm, and is directly proportional to the uric acid concentration in the sample.

    In the ACE CK Reagent assay, serum creatine kinase initiates the conversion of creatine phosphate to creatine with the transfer of a phosphate group to adenosine diphosphate (ADP), forming ATP. The ATP is then used in the phosphorylation of D-glucose to form D-glucose-6-phosphate and ADP. This reaction is catalyzed by hexokinase. The enzyme glucose-6-phosphate dehydrogenase catalyzes the reduction of D-glucose-6-phosphate and nicotinamide adenine dinucleotide phosphate (NADP+). The series of reactions triggered by serum creatine kinase and ending in the formation of NADPH. NADPH strongly absorbs at 340 nm, whereas NADP+ does not. Therefore, the rate of conversion of NADP+ to NADPH can be determined by monitoring the increase in absorbance bichromatically at 340 nm/378 nm. This rate of conversion from NADP+ to NADPH is a function of the activity of CK in the sample.

    AI/ML Overview

    Here's a summary of the acceptance criteria and supporting studies for the Alfa Wassermann ACE Reagents (BUN/Urea, Creatinine, Uric Acid, CK), based on the provided 510(k) summary.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly derived from comparisons to a predicate device (Alfa Wassermann ACE K930104 reagents) and performance characteristics such as precision, accuracy (correlation/regression with predicate), linearity, detection limits, and interference. The reported device performance is from in-house studies and Point-of-Care (POL) studies.

    Note: The document does not explicitly state "acceptance criteria" numerical targets. Instead, it presents performance data for the candidate device, implying that the data's comparability to the predicate and established analytical standards is the basis for acceptance. I will present the reported performance, which demonstrates the device's meeting the necessary equivalency.

    CharacteristicAcceptance Criteria (Implied)Reported Device Performance (Candidate Device)
    Intended UseSame as predicate (quantitative determination in serum)BUN: Quantitative determination in serum and lithium heparin plasma.
    Creatinine: Quantitative determination in serum and lithium heparin plasma.
    Uric Acid: Quantitative determination in serum and lithium heparin plasma.
    CK: Quantitative determination in serum and lithium heparin plasma.
    (Extended to lithium heparin plasma compared to predicate, requiring performance studies in this matrix)
    PlatformsCompatible with ACE Clinical Chemistry SystemACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. (Expanded platforms compared to predicate)
    MethodPhotometric (Same as predicate)Photometric (Same as predicate)
    Calibration Stability7 days (BUN), 2 days (Creatinine), 30 days (Uric Acid)Same
    On-Board Stability30 days (BUN), 10 days (Creatinine), 30 days (Uric Acid), 25 days (CK)Same
    Sample TypeSerum (per predicate)Serum and lithium heparin plasma (Candidate device demonstrates equivalence in both)
    Sample Volume3 µL (BUN, Uric Acid), 20 µL (Creatinine), 5 µL (CK)Same
    Reaction Volume333 µL (BUN), 240 µL (Creatinine), 243 µL (Uric Acid), 170 µL (CK)Same
    Expected ValuesSame as predicateSame
    Measuring Range3-100 mg/dL (BUN), 0.33-25.0 mg/dL (Creatinine), 1.5-16.0 mg/dL (Uric Acid), 11-1350 U/L (CK)Same
    Sample StabilitySame as predicate (storage conditions)Same
    PrecisionLow, Mid, High %CV and SD comparable to predicate/clinical needsIn-House Serum/Plasma: Generally 0.98, Slope ~1, Intercept ~0)
    Creatinine: R > 0.99, Slope 1.003-1.050, Intercept -0.077 to 0.005.
    Uric Acid: R > 0.98, Slope 1.008-1.028, Intercept -0.29 to -0.09.
    CK: R > 0.99, Slope 0.978-1.006, Intercept -0.5 to 0.1. (See pages 8-9)
    Method Comparison (POL)Comparison to In-House ACE results: Slope, Intercept, Correlation (R) and Std Error Est. demonstrating equivalence to predicate system (e.g., R > 0.98, Slope ~1, Intercept ~0).BUN: R > 0.99, Slope 0.989-1.039, Intercept -0.1 to 1.4.
    Creatinine: R > 0.99, Slope 0.977-1.051, Intercept -0.085 to 0.037.
    Uric Acid: R > 0.99, Slope 0.936-1.034, Intercept 0.02 to 0.58.
    CK: R > 0.99, Slope 0.962-1.053, Intercept -16.5 to 1.1. (See pages 14-15)
    Detection Limits (LoB, LoD, LoQ)Low values demonstrating capability to measure analytes at clinically relevant low concentrations.BUN: LoB 1.53, LoD 1.97, LoQ 3.0 mg/dL.
    Creatinine: LoB 0.14, LoD 0.18, LoQ 0.33 mg/dL.
    Uric Acid: LoB 1.11, LoD 1.34, LoQ 1.50 mg/dL.
    CK: LoB 4.68, LoD 8.30, LoQ 11.0 U/L. (See page 16)
    LinearityWide linear range covering clinical needs, with high correlation.BUN: Linear to 100.0 mg/dL, R² 0.9991.
    Creatinine: Linear to 25.0 mg/dL, R² 0.9981.
    Uric Acid: Linear to 16.0 mg/dL, R² 0.9939.
    CK: Linear to 1350.0 U/L, R² 0.9975. (See page 16)
    InterferencesNo significant interference at specified levels of common interferents.Demonstrated no significant interference from icterus, hemolysis, lipemia/triglycerides, and ascorbic acid at clinically relevant concentrations for all four analytes. (See page 17)

    Studies Proving Acceptance Criteria:

    The studies are described under "Performance Data" and "Device Comparison with Predicate" sections of the 510(k) summary. These studies aim to demonstrate substantial equivalence to the previously cleared predicate device (Alfa Wassermann ACE BUN/Urea Reagent, ACE Creatinine Reagent, ACE Uric Acid Reagent, and ACE CK Reagents, K930104).

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set (Matrix Comparison: Serum vs. Plasma):

      • BUN: 95 pairs (ACE), 96 pairs (Alera), 51 pairs (Axcel)
      • Creatinine: 102 pairs (ACE), 102 pairs (Alera), 55 pairs (Axcel)
      • Uric Acid: 97 pairs (ACE), 95 pairs (Alera), 55 pairs (Axcel)
      • CK: 94 pairs (ACE), 96 pairs (Alera), 55 pairs (Axcel)
      • Data Provenance: The document states "In-House Precision" and "In-House Matrix Comparison". This typically implies that the data was generated within the manufacturer's laboratory or a testing facility under their control. The country of origin is not explicitly stated but is implicitly the US, given the 510(k) submission to the FDA. The data is retrospective, as it's being used to characterize reagent performance.
    • Test Set (POL - Method Comparison):

      • BUN: 53-54 samples per POL lab for comparison with In-House ACE.
      • Creatinine: 51 samples per POL lab for comparison with In-House ACE.
      • Uric Acid: 49 samples per POL lab for comparison with In-House ACE.
      • Creatinine Kinase: 48-50 samples per POL lab for comparison with In-House ACE.
      • Data Provenance: "POL - Method Comparison" indicates data from Physician Office Laboratories (POLs), likely external to the main testing facility but still considered part of the overall validation. The document refers to "In-House ACE (x) vs. POL 1 ACE (y)", "POL 2 ACE (y)", etc., indicating comparisons against internal reference methods. The data is retrospective.
    • Test Set (Detection Limits, Linearity, Interferences, Alera Precision): The sample sizes for these specific studies are not explicitly detailed in the provided summary beyond "Low level tested," "Upper level tested," and "number of replicates for precision measurements (i.e. '3.2, 4.0%') implies multiple measurements. These are likely in-house, retrospective studies.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. For in vitro diagnostic devices like these reagents, the "ground truth" is typically established by reference methods or validated comparative methods, often run on established clinical chemistry analyzers. The expertise lies in operating these reference instruments and ensuring proper laboratory practices, rather than expert interpretation of images or clinical cases.

    4. Adjudication Method for the Test Set

    This concept is not applicable to this type of device. Adjudication methods (like 2+1, 3+1) are common in studies involving subjective interpretations (e.g., medical image analysis by radiologists) where discrepancies among readers need to be resolved to establish ground truth. For quantitative IVD reagents, the reference method provides a direct numerical result, not a subjective interpretation requiring adjudication.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    This is not applicable to this type of device. MRMC studies are used to assess the effectiveness of an AI system (or any diagnostic aid) for human readers, particularly in medical imaging. The current device is a diagnostic reagent, which directly measures chemical concentrations, not an AI intended to assist human interpretation of cases.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This is not applicable in the context of an IVD reagent. The "algorithm" here is the chemical reaction and photometric measurement itself. The performance data presented (precision, linearity, method comparison, etc.) is the standalone performance of the reagent on the specified analyzers, without human interpretive input altering the result.

    7. Type of Ground Truth Used

    The ground truth for all performance studies (precision, matrix comparison, method comparison, linearity) is established by comparison against a reference method or a substantially equivalent predicate method performed on existing, validated clinical chemistry analyzers (specifically, the predicate ACE Clinical Chemistry System and the candidate ACE, ACE Alera, and ACE Axcel systems themselves acting as the "reference" for their own performance claims, and for method comparisons, the "In-House ACE" results). This is a common and accepted approach for demonstrating substantial equivalence for IVD reagents.

    8. Sample Size for the Training Set

    This information is not provided and is generally not applicable in the way it is asked for AI/ML devices. These are chemical reagents, not AI/ML algorithms that require "training sets" in the conventional sense of machine learning. The development process would involve formulation, optimization, and internal testing to define assay parameters, which is a different concept than an AI training set.

    9. How the Ground Truth for the Training Set Was Established

    As stated above, the concept of a "training set" with established ground truth in the AI/ML sense is not applicable to these chemical reagents. The "ground truth" during their development and optimization would be based on established analytical chemistry principles and performance measurements against known standards or reference materials.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123802
    Manufacturer
    Date Cleared
    2013-03-01

    (81 days)

    Product Code
    Regulation Number
    862.1110
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    EASYRA TBIL REAGENT, EASYRA DBII REAGENT, EASYRA CK REAGENT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EasyRA TBIL reagent is intended for the quantitative measurement of Total Bilirubin in human serum and plasma of adults on the Medica EasyRA analyzer in clinical laboratories. Bilirubin measurements are used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall bladder block. For in vitro diagnostic use only.

    The EasyRA DBIL reagent is intended for the quantitative measurement of Direct Bilirubin in human serum and plasma of adults on the Medica EasyRA analyzer in clinical laboratories. Bilirubin measurements are used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall bladder block. For in vitro diagnostic use only.

    The EasyRA CK reagent is intended for the quantitative determination of Creatine Kinase (CK) in human serum and plasma, using the MEDICA "EasyRA Chemistry analyzer" in clinical laboratories. Measurements of CK are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy. For in vitro diagnostic use only.

    Device Description

    Not Found

    AI/ML Overview

    This is a 510(k) premarket notification for in-vitro diagnostic reagents, not an AI/ML device. Therefore, the requested information about acceptance criteria, study design, and ground truth for an AI/ML device is not applicable here.

    The document discusses the substantial equivalence of the EasyRA Total Bilirubin Reagent, EasyRA Direct Bilirubin Reagent, and EasyRA Creatinine Kinase Reagent to legally marketed predicate devices. The review is focused on the regulatory classification and general controls provisions for these diagnostic reagents.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113389
    Date Cleared
    2012-08-10

    (268 days)

    Product Code
    Regulation Number
    862.1770
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ACE CK REAGENT,ACE BUN/UREA REAGENT,ACE URIC ACID REAGENT,ACE CREATININE REAGENT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE BUN/Urea Reagent is intended for the quantitative determination of blood urea nitrogen (BUN) concentration in serum using the ACE Axcel Clinical Chemistry System. BUN measurements are used in the diagnosis and treatment of certain renal and metabolic diseases. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Creatinine Reagent is intended for the quantitative determination of creatinine concentration in serum using the ACE Axcel Clinical Chemistry System. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Uric Acid Reagent is intended for the quantitative determination of uric acid concentration in serum using the ACE Axcel Clinical Chemistry System. Uric acid measurements are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions and of patients receiving cytotoxic drugs. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE CK Reagent is intended for the quantitative determination of creatine kinase activity in serum using the ACE Axcel Clinical Chemistry System. Measurement of creatine kinase is used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Device Description

    The ACE Axcel Clinical Chemistry System consists of two major components, the chemistry instrument and an integrated Panel PC. The instrument accepts the physical patient samples, performs the appropriate optical or potentiometric measurements on those samples and communicates that data to an integral Panel PC. The Panel PC uses keyboard or touch screen input to manually enter a variety of data, control and accept data from the instrument, manage and maintain system information and generate reports relative to patient status and instrument performance. The Panel PC also allows remote download of patient requisitions and upload of patient results via a standard interface.

    In the ACE BUN/Urea Reagent assay, urea in serum is hydrolyzed to yield ammonia and carbon dioxide in the presence of urease. The ammonia formed then reacts with 2-oxoglutarate and NADH in the presence of glutamate dehydrogenase to yield glutamate and NAD. Two moles of NADH are oxidized for each mole of urea present. NADH absorbs strongly at 340 nm, whereas NAD+ does not. The initial rate of decrease in absorbance, monitored bichromatically at 340 nm/647 nm, is proportional to the urea concentration in the sample.

    In the ACE Creatinine Reagent assay, creatinine reacts with picric acid in an alkaline medium to form a red-orange colored complex, which absorbs strongly at 505 nm. The rate of complex formation, determined by measuring the increase in absorbance bichromatically at 505 nm/573 nm during a fixed time interval, is directly proportional to the creatinine concentration in the sample.

    In the ACE Uric Acid Reagent assay, uric acid in serum is oxidized by uricase to allantoin and hydrogen peroxide. The hydrogen peroxide then acts to oxdatively couple dichlorohydroxybenzene sulfonic acid and 4-aminoantipyrine in a reaction catalyzed by peroxidase, producing a red colored quinoneimine complex, which absorbs strongly at 505 nm. The amount of chromogen formed, determined by measuring the increase in absorbance bichromatically at 505 nm/610 nm, is directly proportional to the uric acid concentration in the sample.

    In the ACE CK Reagent assay, serum creatine kinase initiates the conversion of creatine phosphate to creatine with the transfer of a phosphate group to adenosine diphosphate (ADP), forming ATP. The ATP is then used in the phosphorylation of D-glucose to form D-glucose-6-phosphate and ADP. This reaction is catalyzed by hexokinase. The enzyme glucose-6-phosphate dehydrogenase catalyzes the reduction of D-glucose-6-phosphate and nicotinamide adenine dinucleotide phosphate (NADP+). The series of reactions triggered by serum creatine kinase and ending in the formation of NADPH. NADPH strongly absorbs at 340 nm, whereas NADP+ does not. Therefore, the rate of conversion of NADP+ to NADPH can be determined by monitoring the increase in absorbance bichromatically at 340 nm/378 nm. This rate of conversion from NADP+ to NADPH is a function of the activity of CK in the sample.

    The ACE BUN/Urea Reagent consists of a single reagent bottle. The reagent contains alpha-ketoglutarate, urease, glutamate dehydrogenase, adenosine diphosphate (ADP), nicotinamide adenine dinucleotide and reduced (NADH).

    The ACE Creatinine Reagent consists of two reagent bottles (Sodium Hydroxide Reagent and Picric Acid Reagent). The Sodium Hydroxide Reagent (R1) contains sodium hydroxide. The Picric Acid Reagent (R2) contains picric Acid.

    The ACE Uric Acid Reagent consists of a single reagent bottle. The reagent contains 4-aminoantipyrine, dichlorohydroxybenzene sulfonic acid, peroxidase and uricase.

    The ACE CK Reagent consists of two reagent bottles (Buffer and Substrate). The Buffer Reagent (R1) contains: imidazole buffer, glucose, N-acetyl-cysteine, magnesium acetate, EDTA, NADP and hexokinase. The Substrate Reagent (R2) contains: creatine phosphate, ADP, AMP, diadenosine pentaphosphate, EDTA and glucose-6-phosphate dehydrogenase.

    AI/ML Overview

    This 510(k) summary describes the analytical performance of the Alfa Wassermann ACE BUN, Creatinine, Uric Acid, and CK Reagents when used with the ACE Axcel Clinical Chemistry System. The study aims to demonstrate substantial equivalence to a predicate device by evaluating precision, accuracy, and detection limits.

    1. Table of Acceptance Criteria (Implied) and Reported Device Performance

    The acceptance criteria for this type of device are generally understood to be that the performance of the new device (ACE Axcel System with new reagents) should be comparable to or better than a legally marketed predicate device (Alfa Wassermann ACE Clinical Chemistry System). While explicit numerical acceptance criteria are not strictly stated as "acceptance criteria" but rather as "reported performance," the goal is to show the device performs within acceptable analytical limits for clinical chemistry assays and is strongly correlated with the predicate.

    Reagent (Analyte)Performance MetricImplied Acceptance Criteria (Comparison to Predicate)Reported Device Performance (ACE Axcel vs. ACE Clinical Chemistry System)
    ACE BUN/UreaPrecision (Within-run CV)0.975 (strong correlation)0.9963 (lab), 0.9982 to 0.9988 (POL)
    Accuracy (Slope CI)Close to 1 (e.g., 0.95-1.05)0.995 to 1.028 (lab), 0.983 to 1.039 (POL)
    Accuracy (Intercept CI)Close to 0 (e.g., -5 to 5)-0.3 to 0.6 (lab), -0.7 to 1.6 (POL)
    Detection LimitClinically relevant low level1.1 mg/dL
    ACE CreatininePrecision (Within-run CV)0.975 (strong correlation)0.9998 (lab), 0.9994 to 0.9998 (POL)
    Accuracy (Slope CI)Close to 1 (e.g., 0.95-1.05)0.975 to 0.983 (lab), 0.961 to 1.027 (POL)
    Accuracy (Intercept CI)Close to 0 (e.g., -0.1 to 0.1)-0.022 to 0.010 (lab), -0.136 to 0.001 (POL)
    Detection LimitClinically relevant low level0.19 mg/dL
    ACE Uric AcidPrecision (Within-run CV)0.975 (strong correlation)0.9958 (lab), 0.9858 to 0.9961 (POL)
    Accuracy (Slope CI)Close to 1 (e.g., 0.95-1.05)1.023 to 1.060 (lab), 0.972 to 1.054 (POL)
    Accuracy (Intercept CI)Close to 0 (e.g., -0.5 to 0.5)-0.18 to 0.07 (lab), -0.31 to 0.28 (POL)
    Detection LimitClinically relevant low level1.13 mg/dL

    Note: Acceptance criteria are implied based on typical expectations for clinical chemistry assays and the intent to demonstrate substantial equivalence to a predicate device. Specific numerical targets for acceptance were not explicitly stated in the provided text, but the strong correlation and low CVs indicate meeting such criteria.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • ACE BUN/Urea Reagent:

      • Accuracy (Correlation Study): 113 samples (clinical laboratory), and patient correlation studies at three Physician Office Laboratory (POL) sites (number of samples not explicitly stated for POL, but implied to be sufficient for regression analysis).
      • Precision: Four BUN levels over 22 days (laboratory study), and three POL sites over 5 days (levels not specified for POL).
      • Data Provenance: Not explicitly stated, but clinical laboratory and Physician Office Laboratory (POL) settings are mentioned, suggesting human serum samples. Whether these were retrospective or prospective is not specified, but typically, method comparison studies use prospective or collected retrospective clinical samples.
    • ACE Creatinine Reagent:

      • Accuracy (Correlation Study): 136 samples (clinical laboratory), and patient correlation studies at three POL sites (number of samples not explicitly stated for POL).
      • Precision: Four creatinine levels over 22 days (laboratory study), and three POL sites over 5 days (levels not specified for POL).
      • Data Provenance: Not explicitly stated, but clinical laboratory and POL settings are mentioned, suggesting human serum samples.
    • ACE Uric Acid Reagent:

      • Accuracy (Correlation Study): 106 samples (clinical laboratory), and patient correlation studies at three POL sites (number of samples not explicitly stated for POL).
      • Precision: Four uric acid levels over 22 days (laboratory study), and three POL sites over 5 days (levels not specified for POL).
      • Data Provenance: Not explicitly stated, but clinical laboratory and POL settings are mentioned, suggesting human serum samples.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. This is an in vitro diagnostic (IVD) device for quantitative measurement of chemical analytes (BUN, Creatinine, Uric Acid, CK) in serum. The 'ground truth' for such devices is established by a reference method or a legally marketed predicate device, not by expert interpretation of images or clinical findings.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, this is an IVD device for quantitative chemical analysis. Adjudication methods are typically used for qualitative or interpretive diagnostic devices where human expert disagreement might occur (e.g., radiology, pathology). Here, the comparison is directly numerical between the candidate device and the predicate device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This is an IVD device for laboratory chemical analysis, not an imaging or interpretive diagnostic device that involves human readers or AI assistance in interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in a sense. The described studies evaluate the performance of the algorithm/system only (the ACE Axcel Clinical Chemistry System with the new reagents) in quantifying the analytes in serum. The performance data (precision, accuracy, detection limit) are intrinsic to the device's analytical capability, independent of human interpretation of the results for the purpose of generating the values themselves. While trained personnel operate the system, the analytical performance is measured as a standalone function of the device.

    7. The Type of Ground Truth Used

    The "ground truth" for the accuracy studies was established by comparing the results from the Alfa Wassermann ACE Axcel Clinical Chemistry System (the new device, 'y') to a legally marketed predicate device, the Alfa Wassermann ACE Clinical Chemistry System ('x'). This is a common method for IVD substantial equivalence, where the predicate is considered the accepted reference for performance. For detection limits, it would typically involve analyzing samples with known, very low concentrations of the analytes or diluting higher concentration samples to determine the lowest measurable level.

    8. The Sample Size for the Training Set

    The provided text describes performance validation studies, not the development or training of an algorithm in the machine learning sense. Therefore, there is no "training set" for an algorithm to learn from in this context. The study focuses on verifying the performance of the already-developed reagent and instrument system.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the machine learning sense for this type of IVD device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K971396
    Device Name
    CK REAGENT
    Date Cleared
    1997-05-29

    (44 days)

    Product Code
    Regulation Number
    862.1215
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    CK REAGENT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1