    Device Name :

    CHISON DIAGNOSTIC ULTRASOUND SYSTEM,CONVEX PROBE,LINEAR PROBE,TRANSVAGINAL MICRO-CONVEX PROBE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The system is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Abdomen, Cardiac, Small Organ (Thyroid, parotid, submaxillary gland, testes and breast.). Peripheral Vascular, Transvaginal, Transrectal, Musculo-skeletal (Conventional and Superficial), Pediatric, Fetal, OB/Gyn and Urology.

    Device Description

    The CHISON iVis20/iVis30/i3/ Q1/Q2/Q3/Q5 ultrasound system is an integrated preprogrammed color doppler ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications. The CHISON ultrasound system can be configured as a roll-around model on wheels (iVis20/iVis30/i3/ Q1/Q2/Q3/Q5). These systems are designed with the latest technology, using the same quality procedure as ultrasound systems which have been available in the market for years. This CHISON ultrasound system is a general purpose, software controlled, diagnostic ultrasound system. Its basic function is to acquire ultrasound echo data and display the image in B-Mode (including Tissue Harmonic Imaging), M-Mode, Pulsed (PW) Doppler Mode, Color Doppler Mode, Power Doppler Mode, Directional Power Doppler Mode, or a combination of these modes.

    AI/ML Overview

    Acceptance Criteria and Study for CHISON iVis Series & Q Series, i3 Diagnostic Ultrasound Systems

    This document describes the acceptance criteria and study information for the CHISON iVis Series & Q Series, i3 Diagnostic Ultrasound Systems, extracted from the provided 510(k) summary.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document is a 510(k) premarket notification summary. For medical devices seeking substantial equivalence, performance characteristics are typically compared against an existing "predicate device" rather than predefined numerical acceptance criteria in the same way as an AI/ML device. The "acceptance criteria" here are implicitly that the new device performs comparably to the predicate device, GE Voluson E6/E8/E8 Expert (K101236), across various imaging modes and clinical applications.

    The reported device performance is that the CHISON iVis Models and Q Models and i3 are "substantially equivalent in safety and effectiveness" to the predicate. This equivalence is based on the systems having the "same intended uses and basic operating modes," "same gray-scale and Doppler capabilities," and acoustic output levels "below the applicable FDA limits."

    The 510(k) summary itself does not contain specific quantitative performance metrics (e.g., sensitivity, specificity, accuracy) for image quality or diagnostic effectiveness, as would often be the case for AI/ML devices or novel technologies. Instead, the focus is on demonstrating that the new device meets the same safety and performance standards as the predicate device already on the market.

    Table: Comparison to Predicate Device (Implied Acceptance Criteria)

    Feature/Performance AspectAcceptance Criteria (Comparable to Predicate: GE Voluson E6/E8/E8 Expert (K101236))Reported Device Performance (CHISON iVis Series & Q Series, i3)
    Intended UsesSame as predicate deviceSame as predicate device
    Basic Operating ModesSame as predicate device (B-Mode, M-Mode, Pulsed Doppler, Color Doppler, Power Doppler, Directional Power Doppler)Same as predicate device
    Gray-scale CapabilitiesComparable to predicate deviceSame capabilities as predicate device
    Doppler CapabilitiesComparable to predicate deviceSame capabilities as predicate device
    Acoustic Output LevelsBelow applicable FDA limitsBelow applicable FDA limits
    Product Safety StandardsCompliance with relevant standards (NEMA UD 2, NEMA UD 3, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 10993-1)Designed to meet listed product safety standards

    2. Sample Size Used for the Test Set and Data Provenance

    The provided 510(k) summary does not contain information about a specific test set, its sample size, or data provenance (e.g., country of origin, retrospective/prospective study design).

    The submission is for a diagnostic ultrasound system aiming for substantial equivalence to a predicate device. This type of submission typically relies on device design specifications, performance testing (e.g., electrical safety, acoustic output, electromagnetic compatibility), and a comparison of technical features and intended uses to the predicate, rather than a clinical study with a "test set" in the context of AI/ML or novel diagnostic algorithms.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Since the 510(k) summary does not describe a clinical study with a test set requiring ground truth established by experts, there is no information provided regarding the number or qualifications of experts for this purpose.

    4. Adjudication Method for the Test Set

    As there is no mention of a clinical study or a test set requiring expert adjudication for ground truth, no adjudication method is described in the provided document.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was performed, nor does it provide any effect size of human readers improving with or without AI assistance. The device is a diagnostic ultrasound system, not explicitly described as incorporating AI for interpretation.

    6. Standalone (Algorithm Only) Performance Study

    The document does not describe a standalone (algorithm only) performance study. The device is a complete ultrasound system, and its performance is evaluated in the context of its overall functionality and comparison to a predicate device, not as a standalone algorithm.

    7. Type of Ground Truth Used

    Given the nature of a 510(k) for a diagnostic ultrasound system (not an AI algorithm or a novel diagnostic modality), the concept of "ground truth" in the context of a clinical test set is not directly applicable or described. The submission focuses on demonstrating that the device's technical specifications, imaging capabilities (B-Mode, M-Mode, Doppler), and safety aspects are equivalent to those of a legally marketed predicate device.

    8. Sample Size for the Training Set

    The provided 510(k) summary does not contain any information about a training set or its sample size. This is consistent with a traditional medical device submission for substantial equivalence, which does not typically involve machine learning model training as a documented component of the submission.

    9. How Ground Truth for the Training Set Was Established

    Since no training set is mentioned or implied, there is no information on how ground truth for a training set was established.