CHEMTRUE SINGLE/MULTI-PANEL DRUG SCREEN DIP CARD/ CASSETTE TESTS
The Chemtron Biotech, Inc.'s Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card Tests are rapid lateral flow immunoassays for the qualitative detection of up to six of the following drugs in a variety of combinations in human urine. The designated cutoff concentrations of each drug and the calibrators used for these drugs are as follows:
Analyte Abbreviation Calibrator Cutoff Concentration
Benzodiazepines BZO Oxazepam 300 ng/mL
Barbiturates BAR Secobarbital/Pentobarbital 300 ng/mL
Ecstasy MDMA/XTC d,l-Methylenedioxymethamphetamine 500 ng/mL
Methadone MTD Methadone 300 ng/mL
Opiates OPI/MOR Morphine 2000 ng/mL
Oxycodone OXY Oxycodone 100 ng/mL
The Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card Tests are intended for the qualitative detection of drugs of abuse for health care professionals, in vitro diagnostic and Over-The-Counter (OTC) use.
The BAR.BZO and OXY assay will yield preliminary positive results when BAR, BZO, and OXY is ingested at or above therapeutic doses. There are no uniformly recognized drug levels for Barbiturate, Benzodiazepine and Oxycodone in urine. The Chemtrue® Single/Multi-Panel Drug Screen Cassette and Div Card Tests shows the drug was or was not present at the cutoff level. This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method for most drugs (HPLC is the preferred confirmatory method for tri-cyclic antidepressants). Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.
The Drugs of Abuse (DOA) Screen Panels are one-step lateral flow immunoassays in which chemically labeled drugs (drug-protein conjugates) compete for limited antibody binding sites with drugs that may be present in urine. The test device consists of up to six test strips placed in separate panels of a plastic holder. On each test strip, a drug-protein conjugate is striped on the test band of the membrane - known as the test region (T) and the drug antibody-colloidal gold conjugate pads are placed at one end of the membrane (opposite in morphine). In the absence of drugs in the urine, the solution of the colored antibody-colloidal gold conjugates move along with the sample solution upward chromatographically by capillary action across the membrane to the immobilized drug-protein conjugate zones on the test band region. The colored antibody-gold conjugates then complexes with the drug-protein conjugates to form visible lines. Therefore, the formation of the visible precipitant in the test band occurs when the test urine is negative for the drug. If any drug is present in the urine, the drug/metabolite antigen competes with the drugprotein conjugates on the test band region for the limited antibody on the colored drug antibodycolloidal gold conjugate pad. When a sufficient amount of drug is present in the urine, the drug will saturate the limited antibody binding sites and the colored antibody-colloidal gold conjugate cannot bind to the drug-protein conjugate at the test strip. Therefore, absence of the color band on the test region indicates a preliminary positive result.
A control band with a different antigen/antibody reaction is added to the membrane strip at the control region (C) to indicate that the test has performed properly. This control line is manufactured as a built-in internal control of the test device and should always appear regardless of the presence of drug or metabolite. If the control line does not appear the test cassette should be discarded. The presence of this colored band in the control region also serves 1) as verification that adequate specimen volume is added (flooding, if too much urine is added, or no flow, due to insufficient urine volume), 2) the test device is properly functioning, and 3) as reagent control.
The acceptance criteria for the Chemtrue® Single/Multi-Panel Drug Screen Dip Card / Cassette Tests are implicitly defined by the reported agreement with GC/MS values in the OTC accuracy study, with a target of "greater than 97.8% accuracy" for lay users and "≥97.8% agreement" with GC/MS. The study demonstrates that lay users can perform and interpret the results correctly with the specified accuracy, and that the device's performance is substantially equivalent to the GC/MS reference method.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
Drug Analyte | Acceptance Criteria (Agreement with GC/MS - Implied) | Reported Performance (Dip Card - Overall Agreement) | Reported Performance (Cassette - Overall Agreement) |
---|---|---|---|
Barbiturates (BAR) | ≥ 97.8% | 98.9% | 100% |
Benzodiazepines (BZO) | ≥ 97.8% | 100% | 100% |
Ecstasy (MDMA) | ≥ 97.8% | 100% | 100% |
Methadone (MTD) | ≥ 97.8% | 97.8% | 97.8% |
Opiates (OPI 2000) | ≥ 97.8% | 100% | 100% |
Oxycodone (OXY) | ≥ 97.8% | 97.8% | 98.9% |
Overall for lay-users | > 97.8% accurate | > 97.8% accurate | > 97.8% accurate |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The study involved a total of 200 OTC lay-users. 100 lay-users tested the Dip Card format, and 100 lay-users tested the Cassette format. Within each format, the results table shows varying numbers of samples tested for each drug, typically with 30 samples in each category (e.g., "no drug present," "GC/MS Negative," "Near cutoff negative," etc.), suggesting at least ~90 positive and ~90 negative samples per drug per format for GC/MS comparison, plus samples in the near-cutoff ranges. The exact total number of unique urine samples is not explicitly stated but implies a substantial number.
- Data Provenance: The data was generated from "three (3) independent sites" through an "OTC accuracy study." The nationality of the participants is not explicitly stated but, given the FDA submission, it can be inferred to be from the United States. The study appears to be prospective as it involved selecting and having lay-users perform tests.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- The ground truth for the test set was established using Gas Chromatography / Mass Spectrometry (GC/MS), which is referred to as "the preferred confirmatory method."
- There were no human experts explicitly described as establishing the ground truth for the test results themselves. The "experts" in this context would be the technicians or laboratory personnel operating and interpreting the GC/MS results, whose qualifications are not specified but are implicitly assumed to be trained professionals for performing such confirmatory chemical analyses.
4. Adjudication Method for the Test Set
- No explicit adjudication method (e.g., 2+1, 3+1) for the interpretation of the test device was used in the lay-user study. The lay-users themselves interpreted the results, and their interpretation was compared against the GC/MS ground truth.
- Discordant results were identified and reported with their GC/MS values, but no process of expert re-adjudication of the test device result itself is described. The agreement percentage directly reflects the lay-user interpretation versus GC/MS.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC comparative effectiveness study was done to compare human readers with AI assistance versus without AI assistance. This device is a rapid diagnostic test with a visual interpretation by the user (lay-user in this case), not an AI-powered diagnostic system that assists human readers.
6. Standalone Performance Study
- Yes, a standalone study was done, specifically the "OTC accuracy study." This study evaluated the performance of the algorithm (i.e., the rapid immunoassay device) when interpreted by a human (lay-users) without any additional human-in-the-loop assistance beyond the instructions provided. The device's performance was compared directly to objective GC/MS results. The results presented in Tables 1 and 2 show the direct agreement rates of the device's output (interpreted by lay-users) with the GC/MS reference.
7. Type of Ground Truth Used
- The type of ground truth used was chemical confirmatory method: Gas Chromatography / Mass Spectrometry (GC/MS). For tri-cyclic antidepressants, HPLC is mentioned as the preferred method, but the drugs in this submission primarily relied on GC/MS.
8. Sample Size for the Training Set
- The document does not explicitly state a sample size for a training set. This device is a rapid immunoassay, which does not typically involve a "training" phase in the same way machine learning or AI models do. The device's formulation and design are developed, and then its performance is validated through studies like the one described. The "Analytical sensitivity (Cut-off characteristics), precision (reproducibility), Accuracy (Method comparison study with clinical samples), specificity and stability study data were established in K111322," which was the predicate device. These earlier studies would have involved samples for the initial characterization and validation but are not detailed here as a "training set."
9. How the Ground Truth for the Training Set Was Established
- As there's no explicitly defined "training set" for an AI or machine learning algorithm, this question isn't directly applicable in the context of this immunoassay device. The ground truth for the development and validation of the immunoassay (e.g., for setting cut-off concentrations, verifying specificity) would have been established through controlled studies using spiked samples and clinical samples, with reference methods such as GC/MS. The document refers to "Performance Data" and states that "Analytical sensitivity (Cut-off characteristics), precision (reproducibility), Accuracy (Method comparison study with clinical samples), specificity and stability study data were established in K111322" (the predicate device's submission), implying that these data, likely using GC/MS or similar reference methods, formed the basis for establishing the device's characteristics.
Ask a specific question about this device
(14 days)
CHEMTRUE SINGLE/MULTI-PANEL DRUG SCREEN DIP CARD/CASSETTE TESTS
The Chemtron Biotech, Inc.'s Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card Tests are rapid lateral flow immunoassays for the qualitative detection of up to six of the following drugs in a variety of combinations in human urine. The designed cutoff concentrations and the calibrators used for these drugs are as follows:
Analyte Abbreviation Calibrator Cutoff Concentration
Amphetamine AMP d-Amphetamine 1000 ng/mL
Cocaine COC Benzoylecgonine 300 ng/mL
Marijuana THC 11-nor-Δ9-THC9-COOH 50 ng/mL
Methamphetamine MET d-Methamphetamine 1000 ng/mL
Opiates OPI/MOR Morphine 300 ng/mL
Phencyclidine PCP Phencyclidine 25 ng/mL
The Chemtrue® Single/Multi-Panel Drug Screen Cassette and Dip Card Tests are intended for the qualitative detection of drugs of abuse for health care professional, in vitro diagnostic and Over-the-Counter (OTC) use. These assays provide only a preliminary result. A more specific alternative chemical method must be used in order to obtain a confirmed assay result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory methods.
Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly when preliminary positive results are indicated.
The Drugs of Abuse (DOA) Screen Panels are one-step lateral flow immunoassays in which chemically labeled drugs (drug-protein conjugates) compete for limited antibody binding sites with drugs that may be present in urine. The test device consists of up to six test strips placed into separate panels of a plastic holder. On each test strip, a drug-protein conjugate is striped on the test band of the membrane - known as the test region (T) and the drug antibody-colloidal gold conjugate pads are placed at one end of the membrane (opposite in morphine). In the absence of drugs in the urine, the solution of the colored antibody-colloidal gold conjugates move along with the sample solution upward chromatographically by capillary action across the immobilized drug-protein conjugate zones on the test band region. The colored antibody-gold conjugates then complexes with the drug-protein conjugates to form visible lines. Therefore, the formation of the visible precipitant in the test band occurs when the test urine is negative for the drug. If any drug is present in the urine, the drug/metabolite antigen competes with drug-protein conjugates on the test band region for the limited antibody on the colored drug antibody-collooidal gold conjugate pad. When a sufficient amount of drug is present in the urine, the drug will saturate the limited antibody binding sites and the colored antibody -colloidal gold conjugate cannot bind to the drug-protein conjugate at the test region of the test strip. Therefore, absence of the color band on the test region indicates a preliminary positive result.
A control band with a different antibody reaction is added to the membrane strip at the control region (C) to indicate that the test has performed properly. This control line is manufactured as a builtin internal control of the test device and should always appear regardless of the presence of drug or metabolite. If the control line does not appear the test cassette should be discarded. The presence of this colored band in the control region also serves 1) as verification that adequate specimen volume is added (flooding, if too much urine is added, or no flow, due to insufficient urine volume), 2) the test device IS properly functioning, and 3) as reagent control.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Device Performance
The core acceptance criteria for this device are its accuracy of detection for various drugs of abuse when compared to a confirmatory method (GC/MS). The study aimed to demonstrate that lay-users could perform the testing and interpret the results correctly with a high level of agreement.
Table 1: Acceptance Criteria and Reported Device Performance (Summary from provided text)
Analyte | Cutoff Concentration | Acceptance Criteria (Implied) | Reported Performance (Agreement with GC/MS) |
---|---|---|---|
Amphetamine | 1000 ng/mL | High agreement (e.g., >95%) | 98.9% (Dip Card), 100% (Cassette) |
Cocaine | 300 ng/mL | High agreement (e.g., >95%) | 100% (Dip Card), 100% (Cassette) |
Marijuana (THC) | 50 ng/mL | High agreement (e.g., >95%) | 100% (Dip Card), 100% (Cassette) |
Methamphetamine | 1000 ng/mL | High agreement (e.g., >95%) | 100% (Dip Card), 100% (Cassette) |
Opiates (Morphine) | 300 ng/mL | High agreement (e.g., >95%) | 100% (Dip Card), 98.9% (Cassette) |
Phencyclidine | 25 ng/mL | High agreement (e.g., >95%) | 98.9% (Dip Card), 98.9% (Cassette) |
Overall | - | >98.9% total correlation | 98.9% total correlation |
Note: The document explicitly states "Correlation studies produced a 98.9% total correlation when compared to the GC/MS methodology" as the overarching performance metric, suggesting this was the primary acceptance criterion. The individual agreement percentages for each drug further support the overall claim.
Study Details
-
Sample size used for the test set and the data provenance:
- Sample Size: The study evaluated 200 OTC (Over-the-Counter) lay-users. The tables provided show results for various drug concentrations, including "no drug present", "GC/MS Negative (-50% to -25% cutoff)", "Near cutoff negative (-25% cutoff to cutoff)", "Near cutoff positive (cutoff to +25% cutoff)", "GC/MS Positive (+25% to +50% cutoff)", and "GC/MS Positive (+50% to 200% cutoff)". For each of these categories and for each drug, there were typically 30 samples, indicating a total of 180 samples per drug, and thus 1080 samples across all 6 drugs for each device type (Dip Card and Cassette). The document states "blind-labeled spiked urine correlation study," implying the urine samples were prepared and then tested.
- Data Provenance: The study used "spiked urine" samples, meaning the drug concentrations were artificially introduced into urine. The samples were tested by lay-users from "three (3) sites". It's not explicitly stated if these sites were in a specific country, but given the company's address (San Diego, CA, USA) and the FDA submission, it's highly probable the study was conducted in the USA. The study design is prospective in the sense that new tests were run according to a predefined protocol.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of experts: Not applicable in the traditional sense of human expert consensus. The ground truth was established by a laboratory method.
- Qualifications: The gold standard for establishing ground truth was Gas Chromatography / Mass Spectrometry (GC/MS) methodology. This is a highly accurate analytical chemistry technique performed by trained laboratory personnel.
-
Adjudication method for the test set:
- Adjudication: Not applicable. The ground truth was determined by GC/MS, so there was no need for human expert adjudication to resolve discrepancies in ground truth. The study compared the device's reading to the GC/MS result directly.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: No, an MRMC comparative effectiveness study was not performed in the context of human readers improving with or without AI assistance. This device is a rapid, visually-read immunoassay, not an AI-powered diagnostic tool requiring interpretation by experts in an MRMC setting. The study assessed the ability of "200 OTC lay-users" to correctly interpret the visual results of the device.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Standalone Performance: No, a standalone algorithm-only performance was not done. The device is a "visually-read" immunoassay where human users (specifically "lay-users" for OTC use) interpret the presence or absence of colored lines. The reported performance includes the human interpretation component.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Type of Ground Truth: The ground truth was established using an analytical reference method: Gas Chromatography / Mass Spectrometry (GC/MS). This is a highly accurate chemical method for identifying and quantifying substances in a sample. The document also mentions Liquid Chromatography / Mass Spectrometry (LC/MS) as an alternative confirmatory method.
-
The sample size for the training set:
- Training Set Sample Size: The document does not explicitly state a training set size. For this type of immunoassay device, there isn't typically a "training set" in the machine learning sense. The device's performance characteristics (e.g., analytical sensitivity, precision, specificity, stability) were established and likely optimized during product development, but this process isn't described as using a distinct "training set" in the filing. The "K102203" submission is referenced for "Other performance data" such as analytical sensitivity and precision, which would have been part of the internal development and validation, not a separate training set as understood in AI/ML.
-
How the ground truth for the training set was established:
- Training Set Ground Truth: As there's no explicitly defined "training set" in the context of this immunoassay's submission, the method for establishing its ground truth is not detailed via this document. However, the foundational analytical sensitivity and specificity of immunoassays are typically established by running known concentrations of analytes and interferents and confirming these with reference methods like GC/MS or LC/MS during the assay's development and internal validation.