Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K970883
    Device Name
    CEDIA DIGOXIN II
    Date Cleared
    1997-05-21

    (72 days)

    Product Code
    Regulation Number
    862.3320
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CEDIA Digoxin II homogeneous enzyme immunoassay is for the quantitation of digoxin in human serum or plasma using automated clinical chemistry analyzers. Measurements are used in the diagnosis and treatment of digoxin overdose and in monitoring levels of digoxin to ensure proper therapy.

    Device Description

    The CEDIA® Digoxin II Assay is based on the bacterial enzyme ßgalactosidase, which has been genetically engineered into two inactive fragments. These fragments spontaneously reassociate to form fully active enzyme that, in the assay format, cleaves a substrate, generating a color change that can be measured spectrophotometrically. In the assay, digoxin in the sample competes with analyte conjugated to one inactive fragment of B-galactosidase for antibody binding site. If analyte is present in the sample, it binds to antibody, leaving the inactive enzyme fragments free to form active enzyme. If analyte is not present in the sample, antibody binds to analyte conjugated on the inactive fragment, inhibiting the reassociation of inactive ß-galactosidase fragments, and no active enzyme is formed.

    AI/ML Overview

    The presented document is a 510(k) Summary for the CEDIA® Digoxin II Assay. It describes the device, its intended use, and compares its performance characteristics to a predicate device, the Abbott TDx® Digoxin II Assay.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as distinct acceptance thresholds (i.e., "the device must achieve X performance"). Instead, the performance of the CEDIA® Digoxin II Assay is compared directly to the predicate device, the Abbott TDx® Digoxin II Assay, implying that substantial equivalence to the predicate's performance is the overarching acceptance criterion.

    FeatureAcceptance Criteria (Implied by Predicate Performance)Reported Device Performance (CEDIA® Digoxin II)
    PrecisionModified NCCLS (ng/mL):
    Level 1Within run %CV: 5.75, Total %CV: 7.67Within run %CV: 4.25, Total %CV: 5.44
    Level 2Within run %CV: 3.15, Total %CV: 3.98Within run %CV: 2.22, Total %CV: 3.58
    Level 3Within run %CV: 1.87, Total %CV: 1.91Within run %CV: 1.56, Total %CV: 2.34
    Lower Detection Limit0.2 ng/mL0.15 ng/mL
    Linearity Range0.0 - 4.0 ng/mL0.15 - 4 ng/mL
    Method Comparison
    vs Abbott TDx Digoxin(Implied high correlation)y = 0.94x + 0.08, r = 0.962 (N=200)
    vs Baxter Dade Stratus(Implied high correlation)y = 0.98x - 0.12, r = 0.955 (N=99)
    Deming's: y = 1.02x - 0.17, r = 0.955 (N=99)
    Interfering SubstancesNo interference at (within ±10% of baseline):
    Bilirubin20 mg/dL66 mg/dL
    Hemoglobin1000 mg/dL1000 mg/dL
    Triglyceride Lipemia2500 mg/dL100 mg/dL
    Total ProteinN/A (not provided for predicate)10 g/dL
    Rheumatoid FactorN/A (not provided for predicate)100 IU/mL
    Specificity (% Cross-reactivity)
    Digoxigeninup to 20060.5
    β-Acetyldigoxinnot tested77.0
    α-Acetyldigoxinnot tested75.3
    Gitalinnot tested2.1
    Digoxingenin-Mono-Digitoxisideup to 200102.5
    Digitoxin-Bis-Digitoxisideup to 20086.3
    Digitoxinup to 2001.5
    β-Methyldigoxin4.876.3
    3-Epe-Digoxigeninnot tested37.6
    3-Dehydrodigoxigeninnot tested82.6
    Epi-Digoxigenin-Glucuronidenot tested42.9

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Precision Test Set N: 120 for each of 3 levels (Level 1, Level 2, Level 3). (Total 360 individual measurements for precision, but N refers to the number of individual tests run for each level.)
    • Method Comparison Test Set N:
      • Vs Abbott TDx Digoxin: N=200
      • Vs Baxter Dade Stratus: N=99
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective or prospective). The samples are referred to as "human serum and plasma."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    There is no mention of experts being used to establish ground truth for this diagnostic assay. The "ground truth" for quantitative assays like this is typically established by comparative analysis against established, often reference, methods or instruments.

    4. Adjudication Method for the Test Set

    Not applicable. As this is an in vitro diagnostic (IVD) assay for quantitative determination, there is no mention of adjudication by human readers/experts for the test results. The device's output is a quantitative value, not subject to interpretation needing adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This document describes an in vitro diagnostic (IVD) assay, not an imaging device or AI algorithm intended for human-in-the-loop interpretation that would typically involve an MRMC study.

    6. Standalone Performance (Algorithm Only Without Human-in-the-Loop Performance)

    Yes, the performance characteristics presented (precision, linearity, detection limit, method comparison, interfering substances, specificity) represent the standalone performance of the CEDIA® Digoxin II Assay. There is no human-in-the-loop component mentioned in its operation for these performance metrics.

    7. Type of Ground Truth Used

    The "ground truth" for the performance studies appears to be based on:

    • Reference Methods/Predicate Devices: For method comparison, the results of the CEDIA® assay were compared against the Abbott TDx® Digoxin II Assay and the Baxter Dade Stratus, which serve as reference or predicate methods.
    • Standard Spiked Samples: For parameters like precision, linearity, and detection limit, it is implied that samples with known concentrations of digoxin were used, or that the "levels" described were standardized reference materials.
    • Known Concentrations of Interfering Substances: For interference and specificity studies, substances were tested at known concentrations.

    8. Sample Size for the Training Set

    The document does not specify a separate "training set" or its size. This is common for traditional in vitro diagnostic assays that are developed using biochemical principles rather than machine learning algorithms requiring explicit training data. The development and optimization process itself serves a similar function to "training," but it doesn't involve a distinct, quantifiable training set in the machine learning sense.

    9. How the Ground Truth for the Training Set Was Established

    As there is no explicit "training set" mentioned in the context of machine learning, this question is not directly applicable. The device's underlying principles are based on enzyme immunoassay chemistry, where performance is optimized through reagent formulation and assay design, not by training on a labeled dataset.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1