Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K093862
    Date Cleared
    2010-05-06

    (140 days)

    Product Code
    Regulation Number
    868.5925
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    CAREVENT HANDHELD CPAP SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CAREvent Handheld CPAP System is intended to provide CPAP to spontaneously breathing patients in the hospital and pre-hospital (EMS) environments.

    Device Description

    The proposed CAREvent Handheld CPAP System is a pneumatically powered device designed to provide CPAP (Constant Positive Airway Pressure) to spontaneously breathing patients who require respiratory support. The patient can breathe spontaneously through the device while the device provides constant positive airway pressure via a face mask. It has only one control for the adjustment of CPAP pressure from 0 to 20 cmH2O.

    The device may be used in pre-hospital transport applications including accident scene, emergency rescue vehicles, hospital transport applications including emergency, surgery, post anesthesia/ recovery and air transport via helicopter or fixed wing.

    The device is a restricted medical device intended for use by qualified medical personnel under the direction of a physician.

    The main accessories for the proposed device are Patient Circuit, Face Mask and Airway Pressure Manometer. They are all disposable for single-patient use.

    AI/ML Overview

    The CAREvent Handheld CPAP System (K093862) underwent non-clinical bench testing to demonstrate its safety and essential performance.

    1. Table of Acceptance Criteria and Reported Device Performance:

    ScopeStandard Applied / Acceptance CriteriaTest Result
    Safety & Essential PerformanceISO10651-5:2006 Product SpecificationsComply
    Patient Valve function after contaminationISO10651-5:2006 Clause 6.1.1Comply
    Function Test after ReassemblyISO10651-5:2006 Clause 6.1.3Comply
    Mechanical Shock (Drop Test/Immersion)ISO10651-5:2006 Clause 6.3.1, 6.3.3Comply
    Mechanical Shock (Splash-proof)ISO10651-5:2006 Clause 6.3.2Comply
    Delivered oxygen concentrationISO10651-5:2006 Clause 7.1.1Comply
    Inadvertent PEEP & Continuing Expiratory PressureISO10651-5:2006 Clause 7.1.3, 7.1.4Comply
    Dead SpaceISO10651-5:2006 Clause 7.1.5Comply
    Pressure limitationISO10651-5:2006 Clause 7.2.4, 7.2.5Comply
    Demand ValveISO10651-5:2006 Clause 7.2.9Comply
    Durability of MarkingsISO10651-5:2006 Clause 8.2.2Comply
    Inspiratory & Expiratory ResistanceISO10651-5:2006 Clause 7.1.2Comply
    Expiratory flow characteristicsAdequately to reduce carbon dioxide rebreathingComply
    Static pressure testing+/- 10%Comply
    Accuracy of pressure gauge+/- 4% of the full rangeComply
    EnvironmentalISO10651-5:2006 Clause 6.2Comply
    Vibration (Sinusoidal)IEC60068-2-6 Test FcComply
    Vibration (Random)IEC60068-2-36 Test FdbComply
    BumpIEC60068-2-29 Test EbComply

    2. Sample size used for the test set and the data provenance:

    The studies were non-clinical bench tests. The specific sample sizes for each test are not explicitly detailed in the provided summary, but it's implied that sufficient samples were tested to meet the "Comply" criteria for each standard. All testing was performed internally by O-Two Medical Technologies or by BET Service Inc. and Exova, indicating the data provenance is from laboratory testing of the device itself. The data is prospective in nature as it involves testing of the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. This was a non-clinical bench study assessing device performance against engineering standards and specifications, not clinical outcomes based on human expert assessment.

    4. Adjudication method for the test set:

    Not applicable. The tests were objective measurements against quantitative physical and performance standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a medical hardware device (CPAP system), not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This device is not an algorithm or AI system. Its performance as a standalone medical device was evaluated through the non-clinical bench tests.

    7. The type of ground truth used:

    The ground truth used for these non-clinical tests were the established industry standards (ISO 10651-5:2006, IEC60068-2-6, IEC60068-2-36, IEC60068-2-29) and the device's own product specifications, along with specific quantitative criteria (e.g., +/- 10% for static pressure, +/- 4% for pressure gauge accuracy, adequately reducing carbon dioxide rebreathing).

    8. The sample size for the training set:

    Not applicable. There is no concept of a "training set" for this type of non-clinical device performance testing.

    9. How the ground truth for the training set was established:

    Not applicable. As there is no training set, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1