Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K020484
    Device Name
    CARESIDE LDH
    Manufacturer
    Date Cleared
    2002-07-03

    (140 days)

    Product Code
    Regulation Number
    862.1440
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    CARESIDE LDH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro diagnostic use with the CARESIDE Analyzer to measure LDH activity from anti-coagulated whole blood, serum or plasma specimens to aid in the diagnosis and treatment of patients with certain liver diseases, heart diseases, and tumors of the lung, kidneys, and liver.

    Device Description

    CARESIDE LDH cartridges are used with the CARESIDE, Inc. CARESIDE Analyzer to measure LDH activity in anti-coagulated whole blood, serum or plasma specimens. The CARESIDE LDH cartridge, a single use disposable in vitro diagnostic test cartridge, delivers a measured volume of serum or plasma to a dry film to initiate the measurement of LDH activity. The patented film cartridge contains all reagents necessary to measure LDH activity.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the CARESIDE LDH device, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance Criteria (Implicit, based on predicate)Reported CARESIDE LDH PerformancePredicate Device (Vitros LDH DT Slides) Performance
    Detection LimitNot explicitly stated, implied to be comparable to predicate or better50 U/L100 U/L
    Reportable RangeNot explicitly stated, implied to be clinically useful50 - 650 U/L100 - 1750 U/L
    RecoveryNot explicitly stated (predicate data not available)Mean: 103%Not available
    LinearityAcceptable slope and correlation coefficientYielded slope and correlation coefficient within acceptable limitsNot available
    InterferenceNo significant interference at specified concentrationsNo significant interference observed at tested concentrations of interferents: Ascorbic Acid: 20 mg/dL, Bilirubin: 20 mg/dL, Triglycerides: 3000 mg/dLNot available
    Precision (Total CV)Not explicitly stated, implied to be clinically acceptable335 U/L, 7.2%649 U/L, 2.9%
    Relative AccuracyHigh correlation and minimal bias to a reference method (BM/Hitachi 902)Careside = 0.97 (BM/Hitachi 902) + 9.1 U/L, r = 0.99Not available

    Study Information

    1. Sample sizes used for the test set and data provenance:

      • The document does not explicitly state the sample size for the test set used for the performance characteristics reported (Detection Limit, Reportable Range, Recovery, Linearity, Interference, Precision, Relative Accuracy).
      • The document does not explicitly state the data provenance (e.g., country of origin, retrospective or prospective) for the test set. It can be inferred that the testing was performed by the manufacturer, CARESIDE, Inc., prior to submission.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in the document. The performance characteristics seem to be based on laboratory measurements against reference instruments or calibrated standards rather than expert consensus on diagnostic images or clinical outcomes.
    3. Adjudication method for the test set:

      • This information is not applicable or provided. The performance assessment for this in vitro diagnostic device does not involve adjudication by multiple experts in the way clinical diagnostic imaging studies might.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done. This device is an in vitro diagnostic assay, not an AI-assisted diagnostic imaging or interpretation tool for human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, the reported performance characteristics (Detection Limit, Reportable Range, Recovery, Linearity, Interference, Precision, Relative Accuracy) likely represent the standalone performance of the CARESIDE LDH system (cartridge + analyzer) as an in vitro diagnostic test. There is no human-in-the-loop component in the analytical measurement itself; the analyzer quantifies LDH activity.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth for the analytical performance appears to be established through:
        • Reference Methods/Instruments: For relative accuracy, it was compared to a "BM/Hitachi 902" analyzer.
        • Calibrated Standards: Linearity and recovery assays typically use samples with known, precisely measured concentrations.
        • Controlled Samples: Interference studies use samples spiked with known interferents at specific concentrations.
    7. The sample size for the training set:

      • The document does not directly mention a "training set" in the context of an AI/algorithm. For an in vitro diagnostic device like this, method development and calibration would be performed using various samples, but these are not typically referred to as a "training set" in the machine learning sense. The device uses lot-specific standard curves, which are established during manufacturing.
    8. How the ground truth for the training set was established:

      • Not applicable as there isn't a "training set" for an AI algorithm. For the device's internal calibration, the instrument uses "lot-specific standard curve" information, which is "bar-coded on each cartridge." This implies that during the manufacturing of each batch (lot) of cartridges, their performance is characterized against known standards to create a unique calibration curve that ensures accurate readings for that specific lot.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1