Search Results
Found 1 results
510(k) Data Aggregation
(260 days)
The Classic Stim is recommended for use for the following conditions:
• Relaxation of muscle spasms
• Prevention or retardation of disuse atrophy
• Increasing local blood circulation
• Muscle re-education
• Immediate post-surgical simulation of calf muscles to prevent venous thrombosis
• Maintaining or increasing range of motion
A portable NMS device for muscle re-education.
This document describes the 510(k) summary for the CARE STIM™ muscle stimulator, and it declares non-clinical testing was sufficient to show substantial equivalence to a predicate device.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (from predicate) | Reported Device Performance (CARE STIM™) |
---|---|
Output characteristics of the predicate device (Ortho Dx muscle stimulator). | Output characteristics are substantially equivalent. |
Intended Use of the predicate device. | Intended Use is substantially equivalent. |
Labeling of the predicate device. | Labeling is substantially equivalent. |
No new questions of safety and effectiveness compared to the predicate device. | No new questions of safety and effectiveness are posed. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not applicable. The study relies on bench testing comparisons to a predicate device, not a human test set.
- Data Provenance: Bench testing data comparing the output characteristics of the CARE STIM™ to the Ortho Dx. The country of origin and whether it's retrospective or prospective are not specified, but it would inherently be a prospective bench test.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Not applicable. Ground truth for device characteristics was established via engineering bench testing, not expert clinical assessment.
4. Adjudication Method for the Test Set
- Not applicable. This was a technical comparison, not a clinical study requiring adjudication of expert opinions.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
- Not applicable. This device is a muscle stimulator, not an AI-powered diagnostic tool. No MRMC study was conducted.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Not applicable. This is a physical medical device, not an algorithm. Bench testing was done on the device's electrical output.
7. The Type of Ground Truth Used
- Bench Test Comparison Records: The "ground truth" for the CARE STIM™ was the established output characteristics, intended use, and labeling of the predicate device (Ortho Dx). This was determined through non-clinical (bench) testing.
8. The Sample Size for the Training Set
- Not applicable. This is not a machine learning device and therefore does not have a training set.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. No training set was used.
Ask a specific question about this device
Page 1 of 1