Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K103159
    Date Cleared
    2011-01-11

    (77 days)

    Product Code
    Regulation Number
    870.5300
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    CARDIO MED MODEL A102, H104, O101, P103, AND Z100

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cardio Med DEFIBRILLATION/CARDIOVERSION/MONOPHASIC OR BIPHASIC PACING/ ECG ELECTRODES are indicated for use in external pacing, defibrillation and monitoring applications as a non-sterile, disposable device for single patient use only. The electrodes provide the conductive interface between the defibrillator and/or the external transcutaneous (noninvasive) cardiac pacemaker and the patient's skin. The electrode is intended for use on adult patients. When a patient requires defibrillation, cardioversion or external pacing, these electrodes will be applied to the patient and connected to the instrument. This device is intended for use on defibrillators whose output is classified as low power (360 joule maximum). Compatible with Cardiac Science (ANDERSON connectors), PHILIPS MEDICAL, PHYSIOCONTROL, and ZOLL MEDICAL models of monophasic and bi-phasic defibrillators, external pacemakers.

    Device Description

    These are single use, non-sterile, self stick defibrillator electrodes packaged in pairs. Effective electrode area = 98 cm2 (15.2 in2). They are radiotransparent. They come in various connector styles to match the specific defibrillator. The construction and materials employed are identical to the predicate. The patient contact material is a conductive adhesive hydrogel identical to the material used in the predicate devices. These pads meet the AAMI Standard ANSI/AAMI DF80:2003 and the connectors meet the FDA performance standard for touch proof ECG connectors.

    AI/ML Overview

    Acceptance Criteria and Device Performance Study for Cardio Med Defibrillation/Cardioversion/Monophasic or Biphasic Pacing/ECG Electrodes (K103154)

    1. Table of Acceptance Criteria and Reported Device Performance

    The Cardio Med Defibrillation/Cardioversion/Monophasic or Biphasic Pacing/ECG Electrodes were tested against the AAMI Standard ANSI/AAMI DF80:2003 for defibrillator pads.

    TestAcceptance Criteria (Pre-Pacing Limits)Acceptance Criteria (Post-Pacing Limits)Reported Device Performance (Pre-Pacing)Reported Device Performance (Post-Pacing)
    DC Offset$\leq$ 100 mV$\leq$ 100 mVPassedPassed
    Offset Instability and Noise$\leq$ 150 μV$\leq$ 150 μVPassedPassed
    AC Small Signal Impedance (10 Hz)$\leq$ 2,000 Ohms$\leq$ 2,000 OhmsPassedPassed
    AC Small Signal Impedance (30 kHz)$\leq$ 5 Ohms$\leq$ 5 OhmsPassedPassed
    AC Large Signal Impedance$\leq$ 2 Ohms$\leq$ 2 OhmsPassedPassed
    Defib Overload Recovery (at 4 seconds)$\leq$ 400 mV$\leq$ 400 mVPassedPassed
    Defib Overload Recovery (at 60 seconds)$\leq$ 300 mV$\leq$ 300 mVPassedPassed
    BiocompatibilityPassed Biocompatibility TestingN/APassedN/A
    Connector Compliance"Touch-proof" (FDA Performance Standard)N/APassedN/A

    2. Sample size used for the test set and the data provenance

    Sample Size: For the AAMI bench tests, three different lots were used, with six samples from each lot subjected to the tests. Additionally, the same testing routine (three lots, two samples each) was applied to a 36-month accelerated age shelf life test.
    Data Provenance: The data is reported from nonclinical bench testing performed by the manufacturer, Cardio Medical Products, Inc. This is prospective data generated specifically for the premarket notification. The country of origin of the data is not explicitly stated in the provided text but is implied to be within the United States, given the FDA submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This study involved bench testing against established engineering standards (AAMI, FDA performance standard). It did not involve human interpretation or subjective assessment of medical images or patient data. Therefore, there were no experts used to establish ground truth in the context of medical diagnosis or interpretation for this specific type of device testing. The "ground truth" was the objective measurement against the specified engineering limits.

    4. Adjudication method for the test set

    Not applicable. As this was bench testing against objective engineering standards, there was no subjective judgment or adjudication required. Each test result was either within the specified limits ("Passed") or not.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is a defibrillator electrode and does not involve AI or human interpretation of medical data in its primary function.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical medical device (defibrillator electrode), not an algorithm or AI software. Therefore, no standalone algorithm-only performance study was conducted.

    7. The type of ground truth used

    The ground truth used was based on objective engineering and performance standards established by the Association for the Advancement of Medical Instrumentation (AAMI Standard ANSI/AAMI DF80:2003) and FDA performance standards for ECG connectors. For biocompatibility, the ground truth was meeting established biocompatibility testing protocols for patient contact materials.

    8. The sample size for the training set

    Not applicable. This is a physical medical device. There was no machine learning or AI algorithm that required a "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there was no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1