Search Results
Found 1 results
510(k) Data Aggregation
(52 days)
CAPIOX RX15 HOLLOW, FIBER OXYGENATOR WITH/WITHOUT HARDSHELL RESERVOIR
The Capiox RX15 is intended to be used during open-heart surgical procedures requiring cardiopulmonary bypass for period up to 6 hours. The Capiox RX15 is for use with patients when required blood flow rate will not exceed 5 L/min (4 L/min if using the CXRX15RW30 product code or CXRX15RE30 Product code).
The Capiox RX15 Hardshell Reservoir is also intended for use in vacuum assisted venous drainage procedures, in post-operative chest drainage, and autotransfusion procedures to aseptically return the blood to the patient for blood volume replacement.
The design of the CAPIOX® RX15 Hollow Fiber Oxygenator with/without Hardshell Reservoir provides a semi-integral device whereby the oxygenator and heat exchanger are joined together, while the hardshell reservoir can be detached from the device assembly.
Here's a breakdown of the acceptance criteria and study information for the CAPIOX® RX15 Hollow Fiber Oxygenator with/without Hardshell Reservoir, based on the provided text:
Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Performance Evaluations) | Reported Device Performance |
---|---|
Gas Transfer | Compared to predicate devices; no clinically significant differences. |
Effects on Blood Components (Hemolysis) | Compared to predicate devices; no clinically significant differences. |
Pressure Drop | Compared to predicate devices; no clinically significant differences. |
Mechanical Integrity | Compared to predicate devices; no clinically significant differences. |
Static Priming Volume | Compared to predicate devices; no clinically significant differences. |
Heat Exchanger Performance | Compared to predicate devices; no clinically significant differences. |
Defoaming | Compared to predicate devices; no clinically significant differences. |
Filtration Efficiency | Compared to predicate devices; no clinically significant differences. |
Flow Rate | Compared to predicate devices; no clinically significant differences. |
Tubing Connection Strength | Compared to predicate devices; no clinically significant differences. |
Sterility Assurance Level (SAL) | In accordance with AAMI guidelines to provide an SAL of 10^-6. |
Ethylene Oxide Residues | Will not exceed maximum residue limits. |
Biocompatibility (blood-contacting materials) | Biocompatible as per ISO 10993, "Biological Evaluation of Medical Devices -- Part 1: Evaluation and Testing" for External Communicating Devices, Circulating Blood, Limited Exposure (≤ 24 hours) Contact Duration. |
Polymer Coating (X-Coating) | Evaluated in an in-vivo animal study; no adverse conditions noted. |
Study Details:
-
Sample size used for the test set and the data provenance:
- The document does not specify explicit sample sizes for each "in-vitro performance evaluation." Instead, it refers to "comparisons of the performance of the CAPIOX® RX15 Hollow Fiber Performance Oxygenator with/without Hardshell Reservoir and the predicate devices were conducted."
- The origin of the data is not explicitly stated as "country of origin," but the submitter information lists "Ashitaka Factory of Terumo Corporation 150 Maimaigi-cho Fujinomiya city, Shizuoka Pref. Japan 418-0015," suggesting the tests were likely conducted in Japan or under the supervision of the Japanese entity.
- All performance evaluations listed are described as "in-vitro performance evaluations," and there is an additional "in-vivo animal study" for the polymer coating. These would be considered prospective studies for the purpose of this submission.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided in the document. The studies performed are primarily objective engineering and biological performance tests, not subjective evaluations requiring expert interpretation of results for ground truth.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- This information is not applicable/provided as the performance evaluations are objective measurements (e.g., gas transfer rates, pressure drops, etc.) rather than subjective assessments that would require adjudication.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. This device is a medical oxygenator, not an AI-assisted diagnostic or interpretive tool that would involve human readers.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, the performance evaluations listed are standalone device evaluations. The device's performance is measured directly against predefined criteria or compared to predicate devices, without human in-the-loop performance as part of the primary evaluation.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The "ground truth" for the in-vitro performance evaluations is based on engineering and biological performance specifications and measurements. For example, gas transfer is measured quantitatively, and biocompatibility is assessed against established ISO standards.
- For the biocompatibility of the polymer coating, an in-vivo animal study was conducted, where the "ground truth" would be the observed biological response or lack thereof in the animal model.
-
The sample size for the training set:
- This is not applicable as the device is a medical hardware device (oxygenator) and not an AI/machine learning algorithm that requires a training set.
-
How the ground truth for the training set was established:
- This is not applicable as there is no training set for this type of medical device submission.
Ask a specific question about this device
Page 1 of 1