Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K230723
    Date Cleared
    2023-05-18

    (63 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    C510G/JUSHA-C510G/C510/JUSHA- C510 LCD Monitor

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    C510G/JUSHA-C510G/C510/JUSHA-C510 LCD Monitor is intended to be used in displaying and viewing digital images, including standard and multi-frame digital mammography, for view, and diagnosis by trained medical practitioners. It is specially designed for breast tomosynthesis applications.

    Device Description

    C510G/JUSHA-C510G/JUSHA-C510/C510 LCD Monitor is the display system with the high resolution (2560*2048), high luminance (1150 cd/m2), and 16-bit (65536 level), built-in DICOM standard LUT. In particular, C510G has ambient brightness adapting, real-time DICOM automatic calibration, full-screen brightness equalization and presence induction system, with these this display can automatic adjustment according to different requirements in order to achieve the best results.
    The product is consisted of the following components:

    • 21.3" Color-TFT LCD Panel
    • DMX0704AR1/Main Board/REV: 1.2
    • C510G LCD Monitor software
    • Power Adapter
    • Data Cable.
    AI/ML Overview

    The provided text describes the regulatory clearance for the Nanjing Jusha Display Technology Co., Ltd's C510G/JUSHA-C510G/C510/JUSHA-C510 LCD Monitor. This device is a medical display intended for viewing and diagnosing digital images, including mammography and breast tomosynthesis applications.

    Crucially, the document states: "The subject of this premarket submission, LCD Monitor, did not require clinical studies to support substantial equivalence." and "The subject of this premarket submission, C510G, does not require animal or clinical studies to support substantial equivalence."

    Therefore, the following points regarding acceptance criteria and study proving performance are not applicable as they relate to clinical studies or studies using ground truth established by experts/pathology/outcomes data. The device's clearance was based on demonstrating substantial equivalence to a predicate device (JUSHA-M550G; K190848) through non-clinical bench testing and adherence to relevant standards.

    Based on the provided document, here's what can be extracted and what cannot:


    1. A table of acceptance criteria and the reported device performance:

    The document describes performance measurements that were conducted as part of bench testing, but it does not specify explicit "acceptance criteria" values against which these measurements were compared. Instead, it states that the device "meets all performance standards" and "complies with" certain IEC standards. The comparison table shows the specifications of the proposed device against the predicate.

    Performance Measurement Conducted (Bench Testing)Proposed Device (C510G) Performance (as reported)Implicit Acceptance Criteria / Standard Met
    Measurement of the angular dependency of luminance response in horizontal, vertical and diagonal directionsData not explicitly detailed, assumed to meet standardsTGI18 guideline (implied, not explicitly stated as a pass/fail value)
    Measurement of the luminance non-uniformity characteristics of the display screenData not explicitly detailed, assumed to meet standardsTGI18 guideline (implied, not explicitly stated as a pass/fail value)
    Measurement of the chromaticity non-uniformity characteristics of the display screenData not explicitly detailed, assumed to meet standardsTGI18 guideline (implied, not explicitly stated as a pass/fail value)
    Measurement of small-spot contrast ratioData not explicitly detailed, assumed to meet standardsNot specified
    Measurement of temporal responseData not explicitly detailed, assumed to meet standardsNot specified
    Performance data on luminance stabilityData not explicitly detailed, assumed to meet standardsNot specified
    Electrical safety and electromagnetic compatibility (EMC)Complies with IEC 60601-1 and IEC 60601-1-2IEC 60601-1 and IEC 60601-1-2 standards
    Key Display Specifications (from Product Comparison Table):Proposed Device (C510G) SpecificationPredicate Device (JUSHA-M550G) Specification (for comparison, not acceptance criteria)
    Resolution2560 x 2048 / 2048 x 25602560 x 2048 / 2048 x 2560 (Same)
    Contrast Ratio2000:11700:1 (Better)
    DICOM calibrated luminance500 cd/m21000 cd/m2 (Different panels with different brightness, but same 5MP resolution for Mammo diagnosis)
    Pixel Pitch0.165 x 0.165 mm0.165 x 0.165 mm (Same)
    DICOM LUT16-bit: 6553616-bit: 65536 (Same)

    The following information cannot be provided from the given document because the device did not require clinical studies for its substantial equivalence determination:

    2. Sample size used for the test set and the data provenance: Not applicable (no clinical test set).
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable (no clinical test set requiring ground truth).
    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable (no clinical test set requiring adjudication).
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable (no AI component, no MRMC study).
    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable (this is a display monitor, not an algorithm).
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable (no clinical study or ground truth required).
    8. The sample size for the training set: Not applicable (no training set for a display monitor).
    9. How the ground truth for the training set was established: Not applicable (no training set).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1