Search Results
Found 1 results
510(k) Data Aggregation
(60 days)
The Boston Sight PD Prosthetic Device for daily wear is indicated for the management of a distorted corneal surface that:
- precludes satisfactory spectacle lens correction
- demonstrates significant improved rigid contact lens corrected vision
- is incapable of wearing traditional corneal lenses because of the inability to achieve adequate lens centration/stability and/or tolerance to physical contact with a lens
Causes of corneal distortion include corneal degeneration (e.g. keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration), corneal dystrophy (e.g. lattice dystrophy, Reis-Bucklers dystrophy), and scarring from surgery (e.g. corneal transplant, LASIK, radial keratotomy), infection, or trauma.
The BostonSight PD Prosthetic Device for daily wear is also indicated for therapeutic use in eyes with ocular surface disease from dry eye (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, limbal stem cell deficiency (e.g.Stevens-Johnson syndrome, chemical and thermal burns, radiation), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for distorted cornea or ocular surface disease, the BostonSight PD Prosthetic Device may incidentally provide correction of refractive error.
The BostonSight PD Prosthetic Device may be cleaned and disinfected using a chemical (not heat) care system.
The BostonSight PD Prosthetic Device is a daily wear, prosthetic device for the ocular surface lathe cut from one of the following fluoro-silicone acrylate materials:
- · roflufocon D (supplied by Contamac Ltd.)
- · roflufocon E (supplied by Contamac Ltd.)
- · oprifocon A (supplied by Bausch & Lomb, Inc.)
- · hexafocon B (supplied by Bausch & Lomb, Inc.)
The BostonSight PD Prosthetic Device is designed to vault over the cornea and rest on the conjunctiva overlying the sclera, resulting in a tear reservoir between the back surface of the prosthetic device and the corneal surface. The tear reservoir masks optical distortions from an irregular corneal surface, and in combination with the device itself, protects the ocular surface from an adverse external environment, including but not limited to dysfunctional evelids and margins. The design parameters are customized to allow for tear exchange underneath the device.
The BostonSight PD Prosthetic Device may be shipped "dry" or "wet" in a polypropylene contact lens case. The primary container for shipping the BostonSight PD Prosthetic Device is the Bausch & Lomb Frequent Replacement Contact Lens Case, with clearance under 510(k) K896685. When shipped "wet". The BostonSight PD Prosthetic Device manufactured from material supplied by Bausch & Lomb, inc. may be packaged and shipped in Boston Advance Comfort Formula Conditioning Solution (K974466) or Boston SIMPLUS Multi-Action solution (K024289). The BostonSight PD Prosthetic Device manufactured from material supplied by Contamac, Ltd. may be packaged and shipped "wet" in in the OPTIMUM by Lobob Cleaning and Disinfecting Storage solution, with clearance under 510(k) K014162.
This document is a 510(k) Summary of Safety and Effectiveness for the BostonSight PD Prosthetic Device, a rigid gas permeable contact lens. As such, it does not contain a study or data proving the device meets acceptance criteria. Instead, it asserts substantial equivalence to predicate devices based on device characteristics, indications for use, and a reliance on prior clearances for materials and clinical performance of similar devices.
Therefore, most of the requested information regarding acceptance criteria, sample sizes, expert involvement, and statistical analyses cannot be extracted from this document. However, I can provide the available information:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state acceptance criteria in a quantitative manner for performance. Instead, it relies on substantial equivalence to predicate devices, implying that if the new device is sufficiently similar to the legally marketed predicate devices, it is considered safe and effective. The table below summarizes the comparison to predicate devices, which implicitly defines the "performance" as being comparable in key design and material aspects.
| Characteristic | BostonSight PD Prosthetic Device (Subject Device) | BostonSight IC Corneal & Scleral Lens (K153066) (Predicate Device) | Boston® Scleral (itafluorofocon B) RGP Contact Lens (P860022/S40) (Predicate Device) |
|---|---|---|---|
| Indication for Use | Therapeutic use in eyes with ocular surface disease from dry eye, limbal stem cell deficiency, disorders of the skin, neurotrophic keratitis, and corneal exposure. May incidentally correct refractive error. | Correction of refractive error in aphakic and non-aphakic persons. For management of irregular corneal conditions (keratoconus, pellucid marginal degeneration, post-keratoplasty or LASIK). | Significantly reduced vision due to distorted corneal surface: 1. precludes satisfactory spectacle lens correction 2. demonstrates significantly improved RGP contact lens corrected vision 3. incapable of wearing traditional corneal lenses Also for ocular surface disorders benefiting from physical protection and saline bath. |
| Device Classification | Class II, Lenses, Rigid Gas Permeable, Daily Wear, HQD | Class II, Lenses, Rigid Gas Permeable, Daily Wear, HQD | Class II, Lenses, Rigid Gas Permeable, Daily Wear, HQD |
| Production Method | Lathe-cut | Lathe-cut | Lathe-cut |
| FDA Group # | Group # 3 Fluoro Silicone Acrylate | Group # 3 Fluoro Silicone Acrylate | Group # 3 Fluoro Silicone Acrylate |
| Water Content | <1% | <1% | <1% |
| UV Absorber/Blocker | YES | YES | YES |
| Materials | roflufocon D, roflufocon E, oprifocon A, hexafocon B | (Implied similar types, as it's the primary predicate for materials, but not explicitly listed in this table.) | itafiuorofocon B |
| Physical Properties | (Specific Dk, refractive index, etc. are listed for each material on page 6, considered comparable to predicates.) | (Not detailed in this summary table, but assessed in K153066 which is referenced) | (Not detailed in this summary table, but assessed in P860022/S40 which is referenced) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This document does not describe a new study with a test set. Instead, it relies on prior clearances for predicate devices and materials. Therefore, no information on sample size for a test set or data provenance is available here.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable, as no new study with a test set requiring ground truth establishment is described in this submission.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a contact lens, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable, as no new study with a test set requiring ground truth establishment is described in this submission. The "ground truth" for contact lens performance and safety is typically established through clinical trials submitted for PMA devices or through extensive post-market experience for predicate devices.
8. The sample size for the training set
Not applicable, as no new clinical study with a training set is described.
9. How the ground truth for the training set was established
Not applicable.
Summary based on the provided document:
The BostonSight PD Prosthetic Device gained 510(k) clearance by asserting substantial equivalence to existing legally marketed predicate devices. The "study" proving it meets acceptance criteria is primarily an analysis of its characteristics (materials, design, intended use) compared to previously cleared devices.
Specifically, the submission cross-references:
- Non-clinical studies and part of the clinical assessment: BostonSight IC Corneal & Scleral Lens (K153066).
- Therapeutic indications for use and device design: Boston® Scleral (itafluorofocon B) RGP Contact Lens (P860022/S40).
- Clinical studies for specific materials:
The document specifies that the new device does not raise different questions of safety and effectiveness than the identified predicate devices. Therefore, direct "acceptance criteria" and "device performance" in the context of a new, standalone study (as is often the case for AI/diagnostic devices) are not presented.
Ask a specific question about this device
Page 1 of 1