Search Results
Found 3 results
510(k) Data Aggregation
(29 days)
Biomet Spine Fusion System
When used as a lumbar intervertebral body fusion device, the Biomet Spine Fusion System is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DDD") at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have had a previous non-fusion spinal surgery at the involved spinal level(s), and may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The Biomet Spine Fusion System is to be implanted via an anterior or posterior approach and is to be combined with supplemental fixation. Approved supplemental fixation systems include the Biomet Spinal Fixation System. The expandable interbody fusion devices are not indicated for vertebral body replacement
The Biomet Spine Fusion System implants are intervertebral body fusion/vertebral body replacement devices consisting of a rectangular or semi-rectangular shape and various heights and footprints. The devices have a hollowed out central area to accommodate bone graft. The upper and lower surfaces of the devices have a series of transverse grooves formed to improve stability and fixation once the device is inserted. The implants are available in a variety of sizes and configurations to approximate anatomical variation in different vertebral levels and/or patient anatomy. The implants are made of PEEK (OPTIMA LT1®) per ASTM F2026, titanium alloy (Ti-6A1-4V ELI) per ASTM F136, commercially pure titanium per ASTM F1580, and may include tantalum markers per ASTM F560. The expandable Biomet Spine Fusion System implants incorporate PEEK (OPTIMA LT1® per ASTM F2026) spacer components and titanium alloy (Ti-6Al-4V ELI per ASTM F136) endplate components that orient into predetermined lordotic angles of 0, 8 and 12 degrees. The expandable implants are available in 7 to 14mm heights, an 11mm width, and 25, 30 and 35mm lengths. The subject Biomet Spine Fusion System implant is provided sterile only.
This document describes a 510(k) premarket notification for the Biomet Spine Fusion System. The submission is to add a tantalum marker to the cam locking mechanism of the existing device. This is a modification to an existing device, not a new AI/ML-enabled device, and therefore does not include the detailed performance study information typically associated with AI/ML device submissions.
Here's a breakdown of the information provided, tailored to the requested categories where applicable:
1. A table of acceptance criteria and the reported device performance:
Acceptance Criteria | Reported Device Performance |
---|---|
Mechanical Performance: Demonstrate substantial equivalence to predicate device (K151483) as per FDA guidance and ASTM standards (ASTM F2077 for static/dynamic axial compression, static/dynamic shear compression, static torsion; ASTM F2267 for subsidence). | An engineering analysis demonstrated that no new mechanical testing was required for the modified device (adding a tantalum marker). The analysis demonstrated substantial equivalence of the new subject devices to the predicate device. The modified Biomet Spine Fusion System met all specified criteria. |
Mechanical testing on the predicate device (K151483) was conducted and demonstrated substantial equivalence to its own predicate devices. |
| Safety and Effectiveness: The modifications should not raise new safety or performance questions. | The modified Biomet Spine Fusion System met all specified criteria and did not raise new safety or performance questions. The minor differences in the new components (tantalum marker) do not raise any new issues of safety or effectiveness. |
| Intended Use & Indications for Use: Maintain the same intended use and indications for use as the predicate. | The modified Biomet Spine Fusion System implants included in this submission have the same intended use, indications, technological characteristics, and principles of operation as the previously cleared Biomet Spine Fusion System (K151483). |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Sample Size: Not applicable. For this specific 510(k) submission, there isn't a "test set" in the context of clinical data for an AI/ML device. The evaluation was primarily engineering analysis and mechanical testing comparison to a predicate device.
- Data Provenance: Not applicable. The "data" here refers to mechanical testing results and engineering analysis, not patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not applicable. This is not an AI/ML device requiring expert ground truth for image or clinical interpretation. The evaluation was based on engineering and mechanical testing standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. As above, there is no "test set" in the sense of clinical data requiring adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is not an AI/ML device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is not an AI/ML device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For the predicate device's mechanical testing, the "ground truth" would be the established ASTM standards (F2077, F2267) and the performance criteria outlined in FDA guidances (Spinal System 510(k)s, Class II Special Controls Guidance Document: Intervertebral Body Fusion Device). For the current submission, the ground truth is simply the established safety and performance of the predicate device (K151483) against which the minor modification is compared via engineering analysis.
8. The sample size for the training set:
- Not applicable. This is not an AI/ML device.
9. How the ground truth for the training set was established:
- Not applicable. This is not an AI/ML device.
In summary, this 510(k) is for a minor modification (adding a tantalum marker) to an existing, previously cleared spinal fusion device. The "study" proving the device meets acceptance criteria is an engineering analysis, which determined that the modification did not alter the mechanical performance or safety/effectiveness of the device and thus no new mechanical testing or clinical studies were required beyond what was done for the original predicate device.
Ask a specific question about this device
(59 days)
Biomet Spine Fusion System
When used as a lumbar intervertebral body fusion device, the Biomet Spine Fusion System is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DDD") at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have had a previous non-fusion spinal surgery at the involved spinal level(s), and may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The Biomet Spine Fusion System is to be implanted via an anterior or posterior approach and is to be combined with supplemental fixation. Approved supplemental fixation systems include the Biomet Spinal Fixation System. The expandable interbody fusion devices are not indicated for vertebral body replacement.
The Biomet Spine Fusion System implants are intervertebral body fusion/vertebral body replacement devices consisting of a rectangular or semi-rectangular shape and various heights and footprints. The devices have a hollowed out central area to accommodate bone graft. The upper and lower surfaces of the devices have a series of transverse grooves formed to improve stability and fixation once the device is inserted. The implants are available in a variety of sizes and configurations to approximate anatomical variation in different vertebral levels and/or patient anatomy. The implants are made of PEEK (OPTIMA LT1®) per ASTM F2026, titanium alloy (Ti-6A1-4V ELI) per ASTM F136, commercially pure titanium per ASTM F1580, and may include tantalum markers per ASTM F560. The expandable Biomet Spine Fusion System implants incorporate PEEK (OPTIMA LT1® per ASTM F2026) spacer components and titanium alloy (Ti-6Al-4V ELI per ASTM F136) endplate components that orient into predetermined lordotic angles of 0, 8 and 12 degrees. The expandable implants are available in 7 to 14mm heights, an 11mm width, and 25, 30 and 35mm lengths. The subject Biomet Spine Fusion System implant is provided sterile only.
The provided document describes a 510(k) premarket notification for the Biomet Spine Fusion System, indicating it is a medical device submission to the FDA. The document focuses on demonstrating substantial equivalence to a predicate device, rather than proving the device meets specific acceptance criteria through a clinical study. Therefore, much of the requested information regarding acceptance criteria, study design, ground truth, and expert involvement is not present in this type of regulatory submission. The performance assessment here is primarily based on mechanical testing, not human reader performance or clinical outcomes.
Here's the information that can be extracted and a clear statement about what is not available based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not provide a formal table of acceptance criteria with corresponding performance metrics in a clinical context. Instead, it states that:
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Meet or exceed predicate device performance in mechanical tests | "In all instances, the modified device met or exceeded predicate device performance, functioned as intended..." |
Function as intended | "...functioned as intended and therefore demonstrated substantial equivalence to the predicate device(s)." |
Comparable properties to previously cleared devices | "Performance data presented also demonstrated comparable properties to the previously cleared devices." |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not applicable in the context of this mechanical testing study. The performance data refers to results from physical tests of the device components.
- Data Provenance: Not applicable. The data is generated from mechanical testing of the device components, not from patient data, and thus does not have a country of origin or a retrospective/prospective classification in the typical sense of a clinical study.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. Ground truth, in this context, would relate to clinical outcomes or expert diagnoses, which are not part of this mechanical testing. The "ground truth" for mechanical testing would be the established engineering specifications and performance limits of the predicate device.
4. Adjudication method for the test set
Not applicable. Adjudication methods are relevant for clinical trials or studies involving expert assessment, which is not described here.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This submission is for a physical intervertebral body fusion device and does not involve AI or human image readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
No, a standalone (algorithm-only) performance evaluation was not done. This device is a physical implant, not a software algorithm.
7. The type of ground truth used
For the mechanical testing, the "ground truth" is established by engineering safety and performance standards (e.g., ASTM F2077, ASTM F2267) and the performance specifications of the legally marketed predicate devices. The goal is to demonstrate that the new device's mechanical properties are equivalent to or better than those of the predicate.
8. The sample size for the training set
Not applicable. This is not a machine learning or AI device that would have a training set. The "training" in product development for such a device refers to engineering design and testing iterative processes, not data training.
9. How the ground truth for the training set was established
Not applicable, as there is no training set in the context of machine learning. The design and testing of the device are based on established engineering principles, material science, and regulatory guidance for spinal implants.
Ask a specific question about this device
(176 days)
Biomet Spine Fusion System
When used as a lumbar intervertebral body fusion device, the Biomet Spine Fusion System ("Biomet Spine Fusion System") is intended for spinal fusion procedures to be used with autogenous bone graft in skeletally mature patients with degenerative disc disease ("DDD") at one or two contiguous spinal levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. These DDD patients may have had a previous non-fusion spinal surgery at the involved spinal level(s), and may have up to Grade 1 spondylolisthesis at the involved level(s). The Biomet Spine Fusion System is to be implanted via an anterior or posterior approach and is to be combined with supplemental fixation. Approved supplemental fixation systems include the Biomet Spine Spinal Fixation System. The expandable interbody fusion devices are not indicated for vertebral body replacement
The Biomet Spine Fusion System implants are intervertebral body fusion/vertebral body replacement devices consisting of a rectangular or semi-rectangular shape and various heights and footprints. The devices have a hollowed out central area to accommodate bone graft. The upper and lower surfaces of the devices have a series of transverse grooves formed to improve stability and fixation once the device is inserted. The implants are available in a variety of sizes and configurations to approximate anatomical variation in different vertebral levels and/or patient anatomy. The implants are made of PEEK (OPTIMA LT1®) per ASTM F2026, titanium alloy (Ti-6A1-4V ELI) per ASTM F136, commercially pure titanium per ASTM F1580, and may include tantalum markers per ASTM F560. The expandable Biomet Spine Fusion System implants incorporate PEEK (OPTIMA LT1® per ASTM F2026) spacer components and titanium alloy (Ti-6Al-4V ELI per ASTM F136) endplate components that orient into predetermined lordotic angles of 0, 8 and 12 degrees. The expandable implants are available in 7 to 14mm heights, an 11mm width, and 25, 30 and 35mm lengths. The Biomet Spine Fusion System is intended to be sterile.
This document describes the Biomet Spine Fusion System, an intervertebral body fusion device. The provided text is a 510(k) summary submitted to the FDA, which focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets acceptance criteria through a standalone clinical study.
Therefore, many of the requested elements (acceptance criteria, specific performance metrics, sample sizes, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, and training set information) are not available from this regulatory submission. This document primarily details mechanical testing and comparison to predicate devices.
Here's a breakdown of the information that is available and what is not available based on the provided text:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria (Implied by testing standards) | Reported Device Performance (from text) |
---|---|
Static and dynamic axial compression (ASTM F2077) | Met or exceeded predicate device performance |
Static and dynamic shear compression (ASTM F2077) | Met or exceeded predicate device performance |
Static torsion (ASTM F2077) | Met or exceeded predicate device performance |
Subsidence (ASTM F2267) | Met or exceeded predicate device performance |
Particulate wear analysis (ASTM F1877) | Met or exceeded predicate device performance |
Note: The specific numerical acceptance criteria for "met or exceeded predicate device performance" are not provided in this document.
2. Sample size used for the test set and the data provenance
- Sample Size: Not applicable. The "tests" mentioned are mechanical performance tests of the device itself, not clinical trials on patient data.
- Data Provenance: Not applicable. These are laboratory mechanical tests on the device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. Ground truth as typically understood in a clinical study (e.g., diagnosis, outcome) is not relevant for the mechanical tests described.
4. Adjudication method for the test set
- Not applicable. Mechanical tests do not involve adjudication by experts.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC comparative effectiveness study was done. This device is an intervertebral body fusion system, not an AI or imaging diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is a medical implant, not an algorithm. The performance tests are for the physical device.
7. The type of ground truth used
- Not applicable in the clinical sense. For mechanical tests, the "ground truth" would be the engineering specifications and performance benchmarks derived from the relevant ASTM standards and predicate devices.
8. The sample size for the training set
- Not applicable. There is no "training set" as this is not an AI or machine learning device.
9. How the ground truth for the training set was established
- Not applicable.
Ask a specific question about this device
Page 1 of 1