Search Results
Found 1 results
510(k) Data Aggregation
(153 days)
The Ballast 088 Long Sheath is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
The Ballast 088 Long Sheath is a sterile, single lumen, variable stiffness Guide Catheter / Long Sheath that is designed for facilitating the introduction of appropriately sized interventional devices into target blood vessels in the peripheral, coronary, and neuro vasculature. The Ballast 088 Long Sheath consists of a lubricous inner liner made from Polytetrafluoroethylene (PTFE) reinforced by a stainless steel coil over the inner liner and stainless steel braid partially covering the inner coil layer. The outer jacket consists of thermoplastics ranging from thermoplastic polyurethane (TPU), polyether block amide (Pebax), and polyamide 12. The distal section of the Ballast 088 Long Sheath is coated with hydrophilic coating to provide lubricity during use. A Platinum/Iridium marker band is placed at the distal end of the catheter for maximum radiopacity. An introducer sheath, 9F hemostasis valve, 8F hemostasis valve adapter and a dilator are also provided with the device.
This document is a 510(k) summary for the Ballast 088 Long Sheath, a medical device. It aims to demonstrate substantial equivalence to a predicate device, the Neuron MAX™ System. The document focuses on performance and biocompatibility testing.
Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:
Important Note: This document describes a medical device (a catheter) and its mechanical, material, and biological performance, not an AI/ML-based device. Therefore, many of the requested categories related to AI/ML (e.g., ground truth, experts, MRMC studies, training set data) are not applicable to this type of device submission. The acceptance criteria here are typically physical and biological performance metrics for a hardware device.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a formal "acceptance criteria" table with pass/fail thresholds for each test. Instead, it describes general test methods and states that "All test samples passed testing" or "The subject device is comparable to the predicate device." For biocompatibility, it states "Non-cytotoxic," "Did not elicit sensitization response," etc.
However, based on the "Performance Bench Testing and Animal Testing Summary" (Table 2) and "Biocompatibility Test Summary" (Table 3), we can infer the acceptance criteria and reported performance for key tests:
| Acceptance Criteria Category/Test | Test Method Summary | Reported Device Performance (Implied Acceptance) |
|---|---|---|
| Mechanical/Physical Performance | ||
| Visual and Dimensional Inspection | Inspected dimensions (overall length, ID, OD, etc.). | All test samples passed testing. |
| Surface Contamination | Microscope inspection for particulates. | All test samples passed testing. |
| Catheter Compatibility | Verify compatibility with accessory devices, guidewires, dilators, inner guide/diagnostic catheters. | All test samples passed testing. |
| Simulated Use | In vitro performance in Bovine Aortic Arch vessel model. | All test samples passed testing. |
| Coating Integrity and Adherence | Dyed samples, simulated use, microscopic inspection. | All test samples passed testing. |
| Lubricity/Durability of Hydrophilic Coating | 25 friction cycles with max 75g force. | All test samples passed testing. |
| Dynamic Burst | MEDRAD Mark V Plus Injection System. | All test samples passed testing. |
| Liquid Leakage at 46 psi | Per ISO 10555-1 and ISO 594-2. | All test samples passed testing. |
| Air Leakage | Per EN ISO 10555-1 and ISO 594-2. | All test samples passed testing. |
| Catheter Static Burst Test | Hydraulic Burst/Leak tester per ISO 10555-1. | All test samples passed testing. |
| Tensile Strength | Instron tensile tester, per ISO 10555-1. | All test samples passed testing. |
| Corrosion Resistance | Immersion in NaCl solution, then boiling water; visual inspection. | All test samples passed testing. |
| Torque Strength | Number of turns-to-failures, compared to predicate. | All test samples passed testing. Subject device comparable to predicate. |
| Kink Resistance Test | Tested against predicate in simulated use; wrapping around pin and microscopic examination. | All test samples passed testing. Subject device comparable to predicate. |
| Catheter Stiffness Testing | Instron tensile tester. | All test samples passed testing. |
| Fracture Resistance and Flexing Test | Resistance to damage by flexing/fracture per ISO 11070. | All test samples passed testing. |
| Particulate Matter Characterization | Quantified after simulated use. | All test samples passed testing. |
| Luer Dimensional Inspection | Per ISO 80369-7. | All test samples passed testing. |
| Separation Force (Luer) | Per ISO 594-1 and ISO 594-2. | All test samples passed testing. |
| Unscrewing Torque (Luer) | Per ISO 594-1 and ISO 594-2; remained attached. | All test samples passed testing. |
| Resistance to Overriding (Luer) | Per ISO 594-1 and ISO 594-2; no override of threads. | All test samples passed testing. |
| Stress Cracking (Luer) | Per ISO 594-1 and ISO 594-2; no cracks. | All test samples passed testing. |
| Subatmospheric-pressure Air Leakage (Luer) | Per ISO 80369-7. | All test samples passed testing. |
| Animal Testing (In Vivo Performance) | ||
| Animal Testing (GLP) | Evaluate in vivo performance in acute porcine model; assess trackability, handling, radiopacity, catheter compatibility. Compared to predicate device. | All test samples passed testing. Trackability, handling, radiopacity, and compatibility comparable to predicate device. |
| Biocompatibility (ISO 10993 Series) | ||
| Cytotoxicity – MEM Elution Test | ISO 10993-5 (48 hours). | Non-cytotoxic (average grade of 0). |
| Sensitization - Guinea Pig Kligman Maximization Test | ISO 10993-10. | Did not elicit sensitization response (0% sensitization). |
| Irritation or Intracutaneous - Rabbits | ISO 10993-10. | Non-irritant (difference between test and control mean score was 0.0). |
| Acute Systemic Toxicity - Mice | ISO 10993-11. | Non-toxic (no significant systemic toxicity or mortality). |
| Systemic Toxicity - Rabbit Pyrogen | ISO 10993-11. | Non-pyrogenic (no temperature rise >0.5℃). |
| Genotoxicity: Ames Test | ISO 10993-3; Salmonella typhimurium and Escherichia coli. | Non-mutagenic (no 2x or 3x increase in revertants). |
| Genotoxicity: Chromosomal Aberration Assay | ISO 10993; Chinese Hamster Ovary (CHO) cells. | Non-genotoxic. |
| Hemocompatibility – Hemolysis (Indirect/Direct) | ISO 10993-4 (ASTM Method). | Non-Hemolytic. |
| Hemocompatibility – Complement Activation | ISO 10993-4 (SC5b-9). | Passed (comparable to predicate). |
| Hemocompatibility – Partial Thromboplastin Time (PTT) | ISO 10993-4. | Passed (clotting times lengthened compared to predicate). |
| Hemocompatibility – Platelet and Leukocyte Count (PLC) | ISO 10993-4. | Passed (results for leukocyte and platelet counts within acceptable range, not statistically significant vs. reference material). |
| Hemocompatibility – Digital Optical Microscopy | 40x images for surface morphology comparison. | Passed (no significant differences in geometry, surface morphology, or particulate contamination vs. predicate). |
| Sterilization & Packaging | ||
| Original Sterilization Validation and Adoption | EO sterilization (100% EO) to SAL 10-6, ISO 11135. | Validation study demonstrated capability to sterilize to SAL 10-6. Subject device successfully adopted to original validation. |
| EO and ECH Residuals | Measured per ISO 10993-7:2008. | Residual traces below specified limits. |
| Bacterial Endotoxin Levels | LAL testing per FDA guidance, USP<85>, European Pharmacopeia BET 2.6.14. | < 2.15 EU/device. |
| Shelf Life | Accelerated aging (1 year). | All acceptance criteria met. Device will perform as intended for 1 year. |
| Packaging Integrity Test | Visual Inspection, Seal Strength (ASTM F88/F88M), Bubble Immersion (ASTM F2096-11), Seal Integrity (ASTM F1886/F1886M-16). Validated for 1-year accelerated aging. | All test samples passed testing. Adequate and effective protection and sterile barrier provided for 1 year. |
2. Sample Sizes Used for the Test Set and Data Provenance
- Sample Sizes: The document consistently states "All test samples passed testing" for the bench tests. However, it does not specify the exact number of samples used for each bench test. For animal testing, it refers to an "acute porcine model" but does not give the number of animals. For the biocompatibility tests, specific ISO standards are cited, which typically define minimum sample sizes for such tests, but the numbers are not explicitly stated in this summary.
- Data Provenance:
- Country of Origin: Not specified in the provided text. It's an FDA submission, implying testing acceptable to US regulatory standards, but the physical location of labs is not mentioned.
- Retrospective or Prospective: These concepts are not applicable to the bench and animal testing described. These are controlled laboratory experiments and prospective animal studies, not data collected from human patients over time.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not Applicable. This is a hardware medical device. Ground truth, in the context of AI/ML, refers to human expert annotations or pathology for diagnostic accuracy. For this device, "ground truth" implicitly comes from physical measurements against engineering specifications and biological reactions governed by established scientific principles in highly controlled lab and animal environments (e.g., a specific tensile strength value must be met, a material must not induce a specific biological reaction). There is no "opinion-based" ground truth from human readers/experts in the way there would be for an AI-diagnostic algorithm.
4. Adjudication Method for the Test Set
Not Applicable. As above, this is for a hardware device. Adjudication methods (like 2+1, 3+1) are used to resolve discrepancies in human interpretations or diagnoses within AI/ML performance studies, which is not relevant here.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not Applicable. This is a physical medical device, not an AI/ML diagnostic aid. Therefore, no MRMC study with human readers assisting AI was conducted.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
Not Applicable. This is a physical medical device. There is no algorithm to test in isolation.
7. The Type of Ground Truth Used
For this device, the "ground truth" is established by:
- Engineering Specifications/Standards: Each mechanical test (e.g., tensile strength, burst pressure, dimensions) has a defined acceptance range based on design specifications and relevant ISO or ASTM standards. "Passing" the test means meeting these predefined quantitative criteria.
- Biological Standards/Regulatory Guidelines: For biocompatibility, the ground truth is whether the device materials elicit specific biological responses (e.g., cytotoxicity, sensitization, irritation, genotoxicity, hemolysis) that fall within acceptable limits defined by ISO 10993 standards. "Non-cytotoxic," "Non-irritant," etc., are the "ground truth" outcomes for these tests.
- Comparative Performance to Predicate: For some tests (e.g., Torque Strength, Kink Resistance, Hemocompatibility), the "ground truth" for substantial equivalence is based on the subject device performing "comparable to" the already-cleared predicate device.
- Validated Protocols: The methods used (e.g., in vitro models, animal models) are validated scientific/engineering protocols providing the "truth" of the device's performance under specific conditions.
8. The Sample Size for the Training Set
Not Applicable. This is a hardware medical device. There is no "training set" in the context of machine learning. The design and manufacturing process are typically developed and refined through engineering iterations, not by "training" an algorithm on data.
9. How the Ground Truth for the Training Set Was Established
Not Applicable. As there is no training set for an AI/ML model, this question is irrelevant to this device description.
Ask a specific question about this device
Page 1 of 1