Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K110864
    Date Cleared
    2011-06-15

    (78 days)

    Product Code
    Regulation Number
    888.3060
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    BIOSPINE VBR SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aesculap Biospine VBR System is indicated for use in the thoracolumbar spine (T1 to L5) for partial or total replacement of a collapsed, or unstable vertebral body due to tumor or trauma (i.e. fracture). The Aesculap Biospine VBR System is intended for use with supplemental spinal fixation systems such as the Aesculap MACS TL or S4 Systems. The Aesculap Biospine VBR System may be used with bone graft.

    Device Description

    The Aesculap BioSpine VBR System is an adjustable vertebral body replacement device that is implanted into the vertebral body space to improve stability of the spine. The device can be adjusted to the exact length required by the patients anatomy affer implantation. Once it is adjusted to the desired length the column is mechanically locked in place by means of locking screws. Spikes on the end plates of the device improve the anchoring of the implant to the vertebral body. It is available in a variety of configurations to accommodate the anatomical requirements of different patients. Components are manufactured from titanium allov (Ti6Al4V) per ASTM F-136, and cobalt chrome (CoCr) per ASTM F1537.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Test Performed)Requirement / StandardReported Device Performance
    Static and Dynamic TorsionASTM F2077Meets or exceeds performance of predicate devices
    Static and Dynamic CompressionASTM F2077Meets or exceeds performance of predicate devices
    SubsidenceASTM F2267Meets or exceeds performance of predicate devices
    Wear DebrisASTM F2077 & ASTM F1877Meets or exceeds performance of predicate devices
    ExpulsionASTM Draft Standard F-04.25.02.02Meets or exceeds performance of predicate devices

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes used for each non-clinical test (e.g., number of devices tested for static torsion). It only generally states that "non-clinical testing was performed." The data provenance is prospective as the tests were conducted specifically for this 510(k) submission. No information on country of origin of the data is provided, but it's implied to be from a testing facility rather than patient data.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not applicable as the study is a non-clinical device performance study, not a medical imaging or diagnostic study that would require expert-established ground truth.

    4. Adjudication Method for the Test Set

    This information is not applicable as the study is a non-clinical device performance study, not a study involving human interpretation or consensus.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not conducted. This document describes non-clinical mechanical testing of a medical implant, not a study evaluating human reader performance with or without AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, a standalone algorithm performance study was not done. This document describes mechanical testing of a medical implant, not the performance of an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device performance study is the established performance requirements and methodologies defined by the referenced ASTM standards (ASTM F2077, ASTM F2267, ASTM F1877, and ASTM Draft Standard F-04.25.02.02) and the performance of legally marketed predicate devices. The goal was to demonstrate that the Aesculap BioSpine VBR System met or exceeded these established benchmarks.

    8. The Sample Size for the Training Set

    This information is not applicable. There is no "training set" as this is a non-clinical mechanical performance study of a physical device, not an algorithm that requires training data.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1