Search Results
Found 6 results
510(k) Data Aggregation
(405 days)
Mobility, MobiliT, BiPolar-i, OMNI MOD Hip System, OMNI K1 and K2 Hip Systems, OMNI Bipolar Heads, Corin Biolox
Delta Ceramic Heads, OMNI Delta Ceramic Heads
The BiPolar-i is intended for use in the following indications: Non-inflammatory degencrative joint disease including osteoarthritis and avascular necrosis in which the acetabulum does not require replacement, Treatment of non-union, femoral neck and trochanteric fractures of the proximal femur, Revision of failed partial hip replacements in which the acetabulum does not require replacement. The BiPolar-i is indicated for cementless use only.
The Trinity Acetabular System is indicated for use in non-intlammatory degenerative joint disease including osteoarthritis and avascular necrosis, rheumatoid arthritis, correctional deformity, developmental dysplasia of the hip (DDH) or congenital dysplasia of the hip (CDH). The Trinity Acctabular System is intended for cementless, single use only.
The indications for the Corin MctaFixM Hip Stem as a total hip arthroplasty, and when used in combination with a Corin hemi arthroplasty head, as a hip hemi-arthroplasty, include: Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis, Rheumatoid arthritis, Correction of functional deformity, Treatment of non-union and femoral neck fractures, Developmental dysplasia of the hip (DDH) and congenital dysplasia of the hip (CDH). The Corin MetaFixTM Hip Stem is indicated for cementless use only.
The indications for the MiniHip Stem as a total hip arthroplasty include: Non-inflammatory degenerative joint disease including osteoarthriis and avascular necrosis Rheumatoid arthritis Correction of functional deformity. Developmental dysplasia of the hip (DDH) and congenital dysplasia of the hiniHip Stem is indicated for cementless use only.
The indications for the TrinityTM Accabular System as a total hip arthroplasty include: Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis, Rheunatoid arthritis, Correction of functional deformity, Developmental dysplasia of the hip (DDH), and congenital dysplasia of the hip (CDH). The Trinity Acctabular System is intended for cementless, single use only.
The Trinity Dual Mobility System is intended for use in the following indications: 1. Non-inflammatory degenerative joint disease, including osteoarthritis & avascular necrosis 2. Rheumatoid Arthritis 3. Correction of functional deformity 4. Revision of previously failed total hip arthroplasty 5. Patients at increased risk of dislocation 6. Developmental dysplasia of the hip (DDH). The Trinity TM Dual Mobility System is indicated for cementless use only.
The indications for the Corin Trinity™ PLUS Accabular Shell as a total hip arthroplasty include: Non-inflammalory degenerative joint disease including ostoarthritis and avascular necrosis. Rheumatoid arthritis. Correction of functional deformity, Revision of previously failed total hip arthroplasty, Developmental dysplasia of the hip (DDH). The Trinity TM PLUS Acctabular Shell is indicated for cement less use only.
The MobiliT Cup, for cemented and cementless use, are indicated for primary replacement of the hip joint: - In degenerative pathologies: primary, secondary or post-traumatic osteoarthritis, rheumatoid arthritis - For patients who have a high risk of dislocation - In cases of necrosis of the femoral head - In cases of fracture of the neck of the femur - In cases of congenital luxation. The MobiliT Cup, for cemented and cementless use, are indicated for revision when the bone tissue remains sufficient after the removal of the previous acetabular cup. The cementless MobiliT standard Cup, with flanges or with flanges and hook are indicated for cementless use only. The cemented MobiliT Cup is indicated for cemented use only.
The Apex Hip System Bipolar Head is intended for use in combination with an Apex Hip System for uncemented primary or revision hemiarthroplasty of the hip. This prosthesis may be used for the following conditions, as appropriate: Femoral neck and trochanteric fractures of the proximal femur. Osteonecrosis of the femoral head, Revision procedures where other devices or treatments for these indications have failed.
The OMNI Hip system Ceramic Femoral Heads are intended for use in combination with the OMNI Hip System Stems as the femoral component in total hip replacement procedures. This ceramic head is intended to articulate with the OMNI Interface Acctabular System or bipolar component. This prosthesis is intended for single use may be used for the following conditions, as appropriate: Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; Rheumatoid arthritis; Correction of functional deformity; Congenital dislocation; Revision procedures where other treatments or devices have failed; Femoral neck and trochanteric fiactures of the proximal femur.
The indications for use of the OMNI Modular Hip Stems in hip arthroplasty include the following conditions, as appropriate: Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; Rheumatoid arthritis: Correction of functional deformity; Congenital dislocation: Revision procedures where other treatments or devices have failed; Femoral neck and trochanteric fractures of the proximal femur. The OMNI Modular Hip stems are indicated for cementless use only and single use implantation.
The indications for use of the OMNI Modular Hip Stems in hip arthroplasty include the following conditions, as appropriate: Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; Rheumatoid arthritis: Correction of functional deformity; Congenital dislocation: Revision procedures where other treatments or devices have failed; Femoral neck and trochanteric fractures of the proximal femur. The OMNI Modular Hip stems are indicated for cementless use only and single use implantation.
The Revival Modular Revision Hip Stem is indicated in revision surgery of femoral components, following failure of primary cemented or un-cemented prosthesis. The REVIVAL™ Hip Stem 100mm distal component is also indicated in primary total hip arthroplasty. The indications for the Revival TM Modular Revision Hip Stem include: Non-inflammatory degenerative joint disease including primary and secondary osteoarthritis. Rheumatoid arthritis, Correction of functional deformity, Treatment of non-union and femoral neck fractment of traumatic dislocations of the hip, Failures of osteotomy, Treatment of arthrodesis. The Revival ™ Revision Hip Stem is indicated for cementless, single use only.
TaperFirM Hip Stem is indicated for the relief of pain and restoration following the effects of femoral neck fracture, osteo, theumatory arthritis, post- traumatic disease effects, avascular necrosis and total hip revision. The Taper it Hip Stem is indicated for hemi-arthroplasty when used in combination with Corin hemiarthroplasty femoral heads. The TaperFitTM Hip Stem is indicated for cemented, single use only.
The indications for use of the K 1 Hip arthroplasty include the following conditions, as appropriate: Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis: Rheumatoid arthritis: Correction of functional deformity: Congenital disfocation: Revision procedures where other treatments or devices have failed; Femoral neck and trochanteric fractures of the proximal femur. The K 1 Hip Stem is indicated for cementless use only and single use implantation.
The indications for the TriFit CF Hip Stem as a total hip arthroplasty and as a hip hemiarthroplasty include: Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis, Rheumatoid arthritis, Correction of functional deformity, Treatment of non-union and femoral neck fractures, Developmental Dysplasia of the Hip (DDH), Previously failed hip surgery. The Trifit CF Hip Stem is indicated for cementless use only.
The indications for the Corin TriFit TSTM Hip atthroplasty, and when used in combination with a Corin hemi arthroplasty head, as a hip hemi-arthroplasty, include: Non-inflammatory degenerative joint disease including osteoadhritis and avascular necrosis, Rheumatoid arthritis, Correction of functional deformity, Treatment of non-union, femoral neck and trochanteric fractures of the proximal femur, Developmental dysplasia of the hip (DDH) or congenital dysplasia of the hip (CDH). The Tril'it TS Hip is intended for cementless use only.
The subject and predicate devices are one in the same and are comprised of several legally marketed Corin Ltd. hip products, which include OMNIIfe Sciences and Apex Surgical hip products. The subject devices include acetabular cups and liners, bone fixation screws, screw hole occluders, cemented and cementless femoral hip stems for primary and revision hip arthroplasty, fixation screws, modular necks, CoCrMo alloy and ceramic femoral heads, dual mobility acetabular systems, and bipolar heads. The purpose of this 510(k) is to notify the FDA of Corin's engineering assessment of the cross-compatibility of the subject devices, identification of conflicts, and updates to the product labeling. The subject hip devices components are manufactured from a variety of materials which include cobalt-chromium-molybdenum alloy, stainless steel alloy, unalloyed titanium, calcium phosphate (Bonit™ coating) Alumina Matrix Composite ceramic (Biolox Delta), and ultrahigh molecular weight polyethylene (UHMWPE), all of which conform to ASTM or ISO standards, or internal standards. The subject femoral hip stems and heads possess the same 12/14 taper design and reference system for determining head and neck offsets.
I am sorry, but the provided text does not contain information about the acceptance criteria and study proving device performance as typically seen in a medical device submission beyond general statements of substantial equivalence. The document is a 510(k) summary for hip replacement components, primarily focusing on cross-compatibility of various existing devices.
The text outlines:
- Device identification and manufacturer information.
- List of numerous predicate devices.
- Detailed descriptions of the indications for use for many hip components (e.g., Corin BiPolar-i, Trinity™ Acetabular System, MetaFix™ Hip, MiniHip™, Trinity™ Dual Mobility, MobiliT™ Cup, OMNI Bipolar Head, OMNI Delta Ceramic Femoral Head, OMNI MOD Hip System, OMNI K1/K2 Hip Systems, Revival™ Modular Hip Stem, TaperFit™, TriFit™ CF/TS Hip). These indications primarily relate to non-inflammatory degenerative joint disease, rheumatoid arthritis, correction of functional deformity, avascular necrosis, fractures, and revision surgeries.
- A "Performance Data" section which describes the types of engineering analyses and bench testing performed to establish cross-compatibility (e.g., ceramic head burst testing, head pull-off, fretting-corrosion, impingement testing, range of motion assessment, comparison of taper geometries, fatigue strength assessment, and contact stress/wear potential).
However, it does not provide:
- A specific table of acceptance criteria and reported device performance for the types of tests mentioned (e.g., what burst pressure was required vs. achieved). It only lists the types of tests done.
- Sample sizes used for test sets or data provenance.
- Number of experts and their qualifications for establishing ground truth (as this pertains to clinical studies, which are not detailed here for performance).
- Adjudication method.
- Multi-reader multi-case (MRMC) comparative effectiveness study information.
- Standalone algorithm performance (as this is not an AI/algorithm-driven device).
- Type of ground truth used (again, this is not a diagnostic device with "ground truth" in the typical sense).
- Sample size for training set.
- How ground truth for the training set was established.
The document's purpose is to demonstrate substantial equivalence of a range of hip components, including their cross-compatibility when used together. The "performance data" refers to the engineering and bench testing conducted to ensure this compatibility rather than clinical performance against specific metrics as one would find for a diagnostic or AI-driven device.
Ask a specific question about this device
(154 days)
Biolox delta Ceramic Heads, Biolox delta Option Ceramic Heads
Biolox® delta Ceramic Heads and Biolox® delta Option Ceramic Heads are indicated for use in total hip replacement with cemented or non-cemented femoral and acetabular components in cases of:
- Non-inflammatory degenerative joint disease including osteoarthritis, avascular necrosis and traumatic arthritis.
- Rheumatoid arthritis.
- Correction of functional deformity.
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques.
- Revision procedures where other treatment or devices have failed.
Specific indications for compatible components that can be used with the above modular heads include:
Salvage/Oncology Hip and Total Femur System components are also indicated for cases of ligament deficiency, tumor resection, trauma and revision of unsuccessful osteotomy or arthrosis. (K974558, K002757, K0213871) *
Interlocking hip stems are indicated for non-cemented application trauma, fracture, oncology or other situations where severe proximal bone loss may compromise the fixation and stability of a standard-type hip replacement prosthesis. (K990830, K042774)*
*Note - for the USA only
The Biolox® delta component is a traditional, one piece ceramic head indicated for primary hip arthroplasty. The material for the device is Zirconia-Platelet Toughened Alumina (ZPTA) 75% Alumina, 24% Zirconia and 1% Platelet. The highly polished spherical surface articulates with a polyethylene liner acetabular component. The modular head attaches to a metallic femoral stem with a Biomet Type I taper. The ceramic heads are available in three sizes, with several offsets.
The Biolox® delta Option component is a modular ceramic head with a Type I or 12/14 adapter sleeve, indicated for primary or revision hip arthroplasty. The material for the device is Zirconia-Platelet Toughened Alumina (ZPTA) 75% Alumina, 24% Zirconia and 1% Platelet. The highly polished spherical surface articulates with a polyethylene liner acetabular component. The modular head attaches to any Biomet metallic femoral stem with a Type 1 taper or a 12/14 taper. using the associated adapter sleeve.
This document is a 510(k) premarket notification for a medical device called "Biolox® delta Ceramic Heads and Biolox® delta Option Ceramic Heads." This is a regulatory submission to the FDA (Food and Drug Administration) for medical device clearance, not a study report detailing acceptance criteria for an AI/software device.
Therefore, the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert adjudication, MRMC studies, and standalone performance for an AI/software device cannot be extracted from this document.
Here's why and what kind of information is present instead:
- Device Type: This document pertains to physical orthopedic implants (ceramic hip components), not a software or AI device. The regulatory pathway (510(k)) is for demonstrating "substantial equivalence" to a legally marketed predicate device, primarily focusing on materials, design, manufacturing processes, and indications for use.
- Focus of the Submission: The submission highlights cumulative changes to already cleared devices, specifically:
- Manufacturing site transfer
- Change of sterilization sub-contractor
- Labeling changes (including Instructions For Use and labels)
- Packaging changes
- Inclusion of additional compatible products
- Performance Data: The document explicitly states:
- Non-Clinical Testing: Includes "Product Compatibility Testing" and "Pyrogenicity Testing."
- Clinical Testing: "None provided as a basis for substantial equivalence."
- Acceptance Criteria for THIS device: For this type of physical device and submission (510(k) for changes), acceptance criteria would relate to engineering specifications, material properties, biocompatibility, sterilization efficacy, and functional performance (e.g., mechanical strength, wear resistance) as demonstrated by bench testing against established standards or predicate device performance. These specific criteria are not detailed in this summary, but rather the types of testing performed are mentioned.
In summary, this document does not contain the information requested about acceptance criteria and a study proving an AI/software device meets those criteria.
Ask a specific question about this device
(187 days)
TGM Biolox delta Ceramic Heads
TGM Biolox® delta Ceramic Heads are indicated for use in total hip replacement with cemented or non-cemented femoral and acetabular components of compatible TGM Medical systems in cases of:
A. Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis.
B. Revision of failed femoral head replacement, cup arthroplasty or other hip procedures.
C. Proximal femoral fractures.
D. Avascular necrosis of the femoral head.
E. Non-union of proximal femoral neck fractures.
F. Other indications such as congenital dysplasia, arthrodesis conversion, coxa plana, coxa vara, coxa valea, developmental conditions, metabolic and tumorous conditions, osteomalacia, pseudarthrosis conversion, and structural abnormalities.
The Biolox® delta Ceramic Heads are modular femoral components used in hip arthroplasty. The head attaches to a femoral stem via a 12/14 Morse taper and articulates with a polyethylene acetabular component. The ceramic material is Transition-Toughened-Platelet Alumina (TTPA) consisting of 75% Alumina, 24% Zirconia and 1% Platelet. The heads are available in three diameters (28, 32 & 36mm) and a variety of offsets.
I am sorry, but based on the provided document, there is no information available regarding acceptance criteria or a study that proves a device meets such criteria to address your detailed request about performance, sample size, ground truth, or expert involvement in a medical device study.
The document is a 510(k) premarket notification summary for "TGM Biolox® delta Ceramic Heads." It primarily focuses on demonstrating substantial equivalence to predicate devices for marketing authorization, not on providing detailed performance studies against specific acceptance criteria.
Specifically, under "Clinical testing," the document explicitly states: "Clinical testing was not required to demonstrate substantial equivalence in this premarket notification."
Therefore, I cannot extract the information required to populate a table of acceptance criteria, describe a study on device performance, or detail aspects like sample size, ground truth, or expert involvement, as these were not part of the submission described.
Ask a specific question about this device
(25 days)
BIOLOX DELTA CERAMIC HEADS WITH 100KGY E-POLY ACETABULAR LINERS
Biolox® delta Ceramic Heads are indicated for use in total hip replacement with cemented or non-cemented femoral and acetabular components in cases of:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- Rheumatoid arthritis.
- Correction of functional deformity.
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques.
- Revision procedures where other treatment or devices have failed.
Specific indications for compatible components that can be used with the above modular heads include:
Salvage/Oncology Hip and Total Femur System components also indicated for cases of ligament deficiency, tumor resection, trauma and revision of unsuccessful osteotomy or arthrosis.
Interlocking hip stems are indicated for non-cemented application in cases of revision, trauma, fracture, oncology or other situations where severe proximal bone loss may compromise the fixation and stability of a standard-type hip replacement prosthesis.
The 100 kGy E-Poly™ Acetabular Liners are intended for cemented and uncemented applications with the following indications for use:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- Rheumatoid arthritis.
- Correction of functional deformity.
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques.
- Revision of previously failed total hip arthroplasty.
Biolox® delta Ceramic Heads are designed to be the bearing surface of a total hip joint replacement. The highly polished spherical surface articulates with a polyethylene liner acetabular component. The modular head attaches to a metallic femoral stem with a Biomet Type I taper.
The 100 kGy E-Poly™ acetabular liners for the Biolox® delta Ceramic Heads are a larger size of the previously cleared 100 kGy E-Poly™ liners, K070364 and K070399.
This document is a 510(k) summary for a medical device (Biolox® delta Ceramic Heads with 100kGy E-Poly™ Acetabular Liners) and does not contain information about acceptance criteria or a study proving that the device meets those criteria, as typically found in clinical studies of AI/software devices.
Instead, this document describes:
- Device Description and Indications for Use: The purpose and intended applications of the hip replacement components.
- Substantial Equivalence Claim: The device's technological similarity to previously cleared predicate devices (K042091, K051411, K061312 for ceramic heads; K070364, K070399 for acetabular liners).
- Non-Clinical Testing Confirmation: It states that all parameters of the "Guidance Document for the Preparation of PreMarket Notifications for Ceramic Ball Hip Systems" were met. This is a crucial point regarding "acceptance criteria" in this context; however, the specific acceptance criteria (e.g., mechanical strength thresholds, wear rates, biocompatibility standards) and the detailed results of the non-clinical tests are not provided in this summary.
Therefore, I cannot provide a response filling in the requested table and study details (sample size, experts, adjudication, MRMC, standalone performance, ground truth, training set) because the provided text does not contain this information.
The 510(k) summary focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics and established non-clinical testing guidelines, rather than presenting a clinical study with new performance metrics against specific acceptance criteria for a novel AI/software device.
Ask a specific question about this device
(27 days)
36MM BIOLOX DELTA CERAMIC HEADS
Biolox® delta Ceramic Heads are indicated for use in total hip replacement with cemented or non-cemented femoral and acetabular components in cases of:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
- Rheumatoid arthritis
- Correction of functional deformity
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques.
- Revision procedures where other treatment or devices have failed.
Specific indications for compatible components that can be used with the above modular heads include:
Salvage/Oncology Hip and Total Femur System components are also indicated for cases of ligament deficiency, tumor resection, trauma and revision of unsuccessful osteotomy or arthrosis.
Interlocking hip stems are indicated for non-cemented application in cases of revision, trauma, fracture, oncology or other situations where severe proximal bone loss may compromise the fixation and stability of a standard type hip replacement prosthesis.
Biolox® delta Ceramic Heads are designed to be the bearing surface of a total hip joint replacement. The highly polished spherical surface articulates with a polyethylene liner acetabular component. The modular head attaches to a metallic femoral stem with a Biomet Type I taper. This submission covers 36mm diameter heads with neck lengths of -3, 0, +3 and +6.
The provided text describes a 510(k) submission for a medical device (36mm Biolox® delta Ceramic Heads), not a study evaluating an AI device or a device that meets acceptance criteria through a performance study. Therefore, most of the requested information cannot be extracted from this document.
However, I can extract information related to non-clinical testing and the absence of clinical testing.
Here's what can be extracted:
- Acceptance Criteria and Reported Device Performance: This document states that "All parameters of the 'Guidance Document for the Preparation of PreMarket Notifications for Ceramic Ball Hip Systems' were met for the devices contained in this 510(k)." This indicates that the device met the non-clinical acceptance criteria outlined in the relevant guidance document. However, the specific parameters and reported performance values are not detailed in this summary.
Table of Acceptance Criteria and Reported Device Performance (Based only on the provided text):
Acceptance Criteria Category | Specific Criteria (from document) | Reported Device Performance (from document) |
---|---|---|
Non-Clinical Testing | Parameters in "Guidance Document for the Preparation of PreMarket Notifications for Ceramic Ball Hip Systems" | "All parameters of the 'Guidance Document for the Preparation of PreMarket Notifications for Ceramic Ball Hip Systems' were met." |
Information that CANNOT be extracted from the provided text:
- Sample sizes used for the test set and the data provenance: Not applicable, as no clinical or AI-based performance study is mentioned.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
- Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a hip implant, not an AI-assisted diagnostic device.
- If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable for a traditional medical device submission without AI or clinical outcome studies detailed.
- The sample size for the training set: Not applicable.
- How the ground truth for the training set was established: Not applicable.
Summary of what was found about the "study" (non-clinical testing):
- Clinical Testing: "None provided" - explicitly stated that no clinical testing was performed for this 510(k) submission.
- Non-Clinical Testing: "All parameters of the 'Guidance Document for the Preparation of PreMarket Notifications for Ceramic Ball Hip Systems' were meet for the devices contained in this 510(k)." This indicates regulatory compliance through in-vitro or bench testing as per guidance documents, not an AI or clinical performance study. The specific "study" is therefore the non-clinical testing performed to satisfy the guidance document, but details of this testing are not provided in the summary.
Ask a specific question about this device
(234 days)
BIOLOX DELTA CERAMIC HEADS
Biolox® delta Ceramic Heads are indicated for use in total hip replacement with cemented or non-cemented femoral and acetabular components in cases of:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
- Rheumatoid arthritis
- Correction of functional deformity
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques.
- Revision procedures where other treatment or devices have failed.
Specific indications for compatible components that can be used with the above modular heads include:
Salvage/Oncology Hip and Total Femur System components are also indicated for cases of ligament deficiency, tumor resection, trauma and revision of unsuccessful osteotomy or arthrosis.
Interlocking hip stems are indicated for non-cemented application in cases of revision, trauma, fracture, oncology or other situations where severe proximal bone loss may compromise the fixation and stability of a standard type hip replacement prosthesis.
Biolox® delfa Ceramic Heads are designed to be the bearing surface of a total hip joint replacement. The highly polished spherical surface articulates with a polyethylene liner acetabular component. The modular head attaches to a metallic femoral stem with a Biomet Type I taper. This submission covers 28mm diameter heads with neck lengths of -3, 0, +3 and +5 and 32mm diameter heads with neck lengths of -3, 0, +3 and +6.
The Biomet Biolox® delta Ceramic Heads device is a modular ceramic head intended for use in total hip replacement. The provided documentation does not include any clinical or standalone performance studies against a set of acceptance criteria. Instead, the submission relies solely on non-clinical testing to demonstrate substantial equivalence to predicate devices.
Here's a breakdown based on the provided information:
1. A table of acceptance criteria and the reported device performance:
Since no clinical study was performed or reported, a table of acceptance criteria and reported device performance related to diagnostic accuracy, sensitivity, specificity, or similar metrics (which are typical for AI-powered devices) cannot be created from the provided text.
The closest information provided is about meeting non-clinical testing parameters:
Acceptance Criterion (Non-Clinical) | Reported Device Performance |
---|---|
All parameters of the "Guidance Document for the Preparation of PreMarket Notifications for Ceramic Ball Hip Systems" were met | All parameters were met. |
This indicates that the device met the established engineering and materials performance standards for ceramic hip systems, likely related to strength, wear, biocompatibility, etc.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
Not applicable. No clinical test set or data from human subjects was used for this 510(k) submission. The testing was entirely non-clinical (laboratory/mechanical testing).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
Not applicable. No clinical test set requiring expert ground truth establishment was conducted. The ground truth for the non-clinical testing would be derived from established engineering standards and measurements.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable. No clinical test set requiring adjudication was conducted.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. No MRMC study was done, as this is a physical medical device (ceramic hip head), not an AI-powered diagnostic tool. The submission states, "Clinical Testing: None provided."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This is a physical implantable device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
For the non-clinical testing, the "ground truth" would be the established performance specifications and material properties defined in the "Guidance Document for the Preparation of Pre-Market Notifications for Ceramic Ball Hip Systems." This would involve physical measurements and tests conforming to recognized standards (e.g., ISO standards for implants).
8. The sample size for the training set:
Not applicable. There was no "training set" in the context of machine learning or AI algorithms, as this is a physical device.
9. How the ground truth for the training set was established:
Not applicable, as there was no training set.
Ask a specific question about this device
Page 1 of 1