510(k) Data Aggregation

    K Number
    K113836
    Date Cleared
    2012-09-24

    (271 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    BENEHOLD CHG TRANSPARENT FILM DRESSING (10X12CM) (8.5X12CM) (6.5X7 CM)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BeneHold CHG Transparent Film Dressing is intended to cover and protect catheter sites and to secure devices to the skin. Common applications include securing and covering IV catheters, other intravascular catheters and percutaneous devices.

    Device Description

    The BeneHold CHG Transparent Film Dressing is used to cover and protect catheter sites and to secure devices to the skin. The dressing consists of a transparent film with an acrylic adhesive which contains Chlorhexidine Gluconate (CHG), a well-known antiseptic agent with broad-spectrum antimicrobial and antifungal activity. The dressing is a barrier to liquid (waterproof), bacteria and yeast, and protects the IV site from outside contamination. The dressing absorbs between 8 and 21 ml of fluid depending on the dressing size. In vitro testing (barrier, kill time and log reduction) demonstrates that the BeneHold CHG Transparent Film Dressing has an antimicrobial effect against, and is a barrier to, the passage of a variety of gram-positive and gram-negative bacteria, yeast and fungus in the dressing. BeneHold CHG Transparent Film Dressing is transparent, allowing continual site observation, and is breathable, allowing good moisture vapor exchange.

    AI/ML Overview

    The provided 510(k) summary (K113836) for the BeneHold CHG Transparent Film Dressing describes performance testing, but does not provide specific acceptance criteria or detailed study results for all tests. It lists the tests performed but largely states that they demonstrated "conformance to design specifications" or that "testing indicates that the BeneHold CHG Transparent Film Dressing is substantially equivalent."

    Therefore, much of the requested information regarding specific acceptance criteria, detailed device performance, sample sizes, ground truth establishment, and expert involvement is not present in the provided text.

    Here's a breakdown of what can be extracted and what is missing:


    1. Table of Acceptance Criteria and Reported Device Performance

    Note: Specific acceptance criteria and quantitative performance metrics are generally not explicitly stated in this 510(k) summary. The summary broadly states that the device "conforms to design specifications" and demonstrated "antimicrobial effect" and "barrier" properties.

    Performance TestAcceptance Criteria (Not Explicitly Stated)Reported Device Performance (Summary)
    Fluid Handling Capacity(Not specified)Absorbs between 8 and 21 ml of fluid depending on dressing size.
    Static Absorption(Not specified)(Performed, no specific result mentioned beyond general conformance)
    Moisture Vapor Transmission Rate(Not specified)Breathable, allowing good moisture vapor exchange.
    Peel Adhesion(Not specified)(Performed, no specific result mentioned beyond general conformance)
    Liner Release(Not specified)(Performed, no specific result mentioned beyond general conformance)
    Finat Tack(Not specified)(Performed, no specific result mentioned beyond general conformance)
    Seal Strength(Not specified)(Performed, no specific result mentioned beyond general conformance)
    Die Migration(Not specified)(Performed, no specific result mentioned beyond general conformance)
    CHG/PCA Content(Not specified)(Performed, no specific result mentioned beyond general conformance)
    Antimicrobial Effectiveness(Not specified, but generally means significant reduction of pathogens)In vitro testing demonstrates an antimicrobial effect against, and is a barrier to, a variety of gram-positive and gram-negative bacteria, yeast and fungus in the dressing.
    Bacterial/Viral Barrier(Not specified)Barrier to liquid (waterproof), bacteria and yeast, and protects the IV site from outside contamination.
    Sterility(Not specified, but assumed to meet sterility standards)(Performed, no specific result mentioned beyond general conformance)
    Biocompatibility (Irritation)(Negligible irritation)Negligible irritation.
    Biocompatibility (Cell Lysis/Toxicity)(No evidence)No evidence the device causes cell lysis or toxicity.
    Biocompatibility (Sensitization)(No evidence)No evidence of delayed dermal contact sensitization.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size:
      • Test Set: Not specified in the document. The document refers to "in vitro testing" for antimicrobial effects and barrier properties, but the sample sizes (e.g., number of dressings, number of bacterial cultures) are not provided.
      • Training Set: Not applicable as this is a physical medical device, not an AI/ML diagnostic device with a training set.
    • Data Provenance: The studies are described as "in vitro testing." The country of origin of the data is not specified, but the submitter is "Avery Dennison Belgie, BVBA" (Belgium) and the contact person is in the US. The studies are described as performance tests, which are typically conducted in a laboratory setting for regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Not applicable/Not specified. This is a physical device performance study, not an imaging/diagnostic study requiring expert ground truth establishment in a clinical context. Laboratory testing for physical and chemical properties typically relies on standardized methods and measurements rather than expert consensus on a "ground truth" derived from clinical cases.

    4. Adjudication Method for the Test Set

    • Not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical or imaging studies where multiple readers interpret data and discrepancies need resolution. This document describes laboratory-based performance testing of a physical device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No. An MRMC study is not mentioned or implied by the described performance testing. This type of study is relevant for diagnostic devices (especially AI-powered ones) where human reader performance is being evaluated with and without assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This is not an algorithm or AI device.

    7. The Type of Ground Truth Used

    • In this context, the "ground truth" for the device's performance tests would be established through scientific measurements and standardized laboratory protocols. For example:
      • For fluid handling capacity, the ground truth is the measured volume of fluid absorbed.
      • For antimicrobial effectiveness, the ground truth is the measured log reduction of microbial population compared to controls based on established microbiological methods.
      • For biocompatibility, the ground truth is determined by standardized in vitro or in vivo tests for irritation, cytotoxicity, and sensitization.

    8. The Sample Size for the Training Set

    • Not applicable. This is a physical medical device, not an AI/ML diagnostic device.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. This is a physical medical device, not an AI/ML diagnostic device.