Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K954850
    Manufacturer
    Date Cleared
    1996-10-30

    (373 days)

    Product Code
    Regulation Number
    870.4350
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HF-6000 Oxygenator provides the gas exchange and temperature regulating capacity required to support a patient undergoing a cardiopulmonary bypass procedure for a period of up to six (6) hours.

    Device Description

    The Bard® William Harvey® HF-6000 Membrane Oxygenator is a single use gas exchange device with an integral venous side heat exchanger. Venous blood is pumped into the stainless steel heat exchanger, then is directed over the gas exchange hollow fibers. Oxygen and carbon dioxide transfer occur by diffusion as gas flows through the hollow fibers.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Bard® William Harvey® HF-6000 Membrane Oxygenator, based on the provided text.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance
    PerformanceISO/DIS 7199 (Performance Characteristics)Substantially equivalent to predicate devices.
    IntegrityISO/DIS 7199 (Integrity Testing)Met all criteria.
    Ex Vivo Animal StudyAbility to provide gas exchange and temperature regulation for up to 6 hours on partial bypass, with no significant adverse clinical, physiological, hematological, serum chemistry, gross pathology, or histopathology observations.Demonstrated ability to support an animal on partial bypass for up to six hours, with no meaningful adverse observations or concerns regarding safety. Concluded to be fully safe and effective.
    BiocompatibilityISO 10993 (Biological Evaluation of Medical Devices Part-1: Evaluation and Testing)Passed all required tests.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the numerical sample size for the test set used in the ISO/DIS 7199 performance and integrity testing or the biocompatibility testing. For the ex vivo animal study, it mentions supporting "an animal," implying a single animal or an unspecified small number.

    • Test Set Sample Size: Not explicitly stated for performance/integrity/biocompatibility. "An animal" for the ex vivo study.
    • Data Provenance: The document does not specify the country of origin. The ex vivo animal study is prospective in nature (conducting a new experiment) but described retrospectively in the submission. The ISO/DIS 7199 testing and biocompatibility testing are also laboratory-based studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to this device and its testing. The "ground truth" for an oxygenator's performance is established by objective, quantitative measurements against international standards (like ISO/DIS 7199) and physiological markers in animal models, not by expert interpretation of images or patient outcomes in the same way as, for example, an AI diagnostic tool.

    4. Adjudication Method for the Test Set

    This is not applicable as the tests are objective measurements against standards, not subjective assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This is not applicable. The device is an oxygenator, a medical device used in cardiopulmonary bypass, not an AI or imaging diagnostic tool that would involve human readers or MRMC studies.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This is not applicable. The device is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the device's performance and safety was established through:

    • Objective measurements against international standards: ISO/DIS 7199 for performance and integrity.
    • Physiological and histopathological data from an animal model: For the ex vivo animal study, assessing clinical outcomes, physiological responses, hematological and serum chemistry parameters, gross pathology, and histopathology.
    • Biocompatibility testing against international standards: ISO 10993.

    8. The Sample Size for the Training Set

    This is not applicable. The device is a physical medical device, not an AI model that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reason as #8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1