Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K971198
    Manufacturer
    Date Cleared
    1997-10-03

    (186 days)

    Product Code
    Regulation Number
    880.5130
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Babytherm 8004 / 8010 is an infant warmer system intended as a therapeutic device providing warmth for premature and full-term neonates and infants with a weight up to 16 lbs. It is intended for use in delivery rooms, operating theatres, maternity and obstetric units, neonatal wards, neonatal and paediatric intensive care units. Both Systems can be equipped with the optional phototherapy system for treatment of Hyperbilirubinemia.

    Device Description

    Babytherm 8004 / 8010 is a warmer system for infants that provides a combination of a heated mattress (Babytherm 8010, only), a radiant heater and an optional phototherapy system. The radiant heater is in a fixed position over the mattress and is equipped with an examination light. The phototherapy system can be installed by the manufacturer or by the biomedical engineer in the housing of the radiant heater. Babytherm 8004 provides warmth by a radiant heater in "manual mode" or with "skin temperature control". Babytherm 8010 provides warmth by a radiant heater in "manual mode" or with "skin temperature control". The Babytherm 8010 also has a heated gel mattress for conductive warmth.

    AI/ML Overview

    This document is a 510(k) Pre-market Notification for the Dräger Babytherm 8004 / 8010 infant warmer system. The document claims substantial equivalence to several predicate devices.

    Here's an analysis of the provided information regarding acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and the Reported Device Performance:

    The document presents a "Table of Comparison" which implicitly serves as the acceptance criteria by comparing the new device (Dräger Babytherm 8004 / 8010) against predicate devices (Babytherm 8000 WB, Radiant Heater RH 600, PT4000, and Microlite™). The "reported device performance" for the Dräger Babytherm 8004 / 8010 is listed within this table.

    Feature / Acceptance CriteriaDräger Babytherm 8004 / 8010 Reported PerformanceEquivalent Predicate DeviceNotes/Differences
    Intended UseInfant warmer system providing warmth for premature and full-term neonates and infants up to 16 lbs. Use in delivery rooms, neonatal wards, NICUs. Optional phototherapy for Hyperbilirubinemia.Babytherm 8000 WB, Radiant Heater RH 600, PT4000, Microlite™Combines functions of multiple predicate devices.
    Physical Size
    Height (mm)1820 to 2385 with height adjustmentBabytherm 8000 WB (1220 to 1515)New device is taller.
    Length (mm)1315Babytherm 8000 WB (1070)New device is longer.
    Width (mm)Like Babytherm 8000 WB (750)Babytherm 8000 WB (750)Same.
    Weight135 kg maxBabytherm 8000 WB (85 kg max)New device is heavier.
    Mattress Size (mm)Like Babytherm 8000 WB (750 x 490)Babytherm 8000 WB (750 x 490)Same.
    Bassinet TiltLike Babytherm 8000 WB (± 10° from horizontal)Babytherm 8000 WB (± 10° from horizontal)Same.
    Case MaterialLike Babytherm 8000 WB (PS-TSG)Babytherm 8000 WB (PS-TSG)Same.
    Case ColorLike Babytherm 8000 WB (blue)Babytherm 8000 WB (blue)Same.
    Power Input
    VoltageLike Babytherm 8000 WB (120 VAC, 60 Hz)Babytherm 8000 WB (120 VAC, 60 Hz)Same.
    Total Power (W)670 W (8004) / 790 W (8010) + 120 W (height adjust) + 440 W (Phototherapy)Babytherm 8000 WB (135 W + 560 W height adjust), RH 600 (735 W), PT4000 (120 W), Microlite™ (300 W)Higher total power due to combined functionality.
    Fuses2 x 15 ABabytherm 8000 WB (2 x 1.6 A / 2 x 6.3 A), RH 600 (2 x 6.3 A), PT4000 (2 x 1.6 A), Microlite™ (2 x 4 A)Higher fuse rating, likely due to increased power draw.
    Leakage Current< 0.5 mABabytherm 8000 WB (< 0.2 mA), RH 600 (< 0.1 mA), PT4000 (< 0.2 mA), Microlite™ (< 0.1 mA)Higher leakage current, but still within acceptable safety limits (presumably IEC standards).
    Heated Mattress (Babytherm 8010 only)
    TypeLike Babytherm 8000 WB (4 mm Aluminium plate with heating foil)Babytherm 8000 WBSame.
    PowerLike Babytherm 8000 WB (24 VAC / 120 W)Babytherm 8000 WBSame.
    Setting Temp. RangeLike Babytherm 8000 WB (30.0°C - 38.0°C)Babytherm 8000 WBSame.
    Measurement RangeLike Babytherm 8000 WB (5.0°C - 45.0°C)Babytherm 8000 WBSame.
    AccuracyLike Babytherm 8000 WB (± 0.5 °C)Babytherm 8000 WBSame.
    Applied PartLike Babytherm 8000 WB (BF)Babytherm 8000 WBSame.
    Radiant Heater
    TypeLike RH 600 (ceramic)RH 600Same type of ceramic heater, but different types and improved geometry.
    PowerLike RH 600 (600 W)RH 600Same. Two ceramic heaters of 300W each.
    ControlLike RH 600 (manual or servo)RH 600Same.
    Distance heater - mattress80 cmRH 600 (70 cm default, possible 60cm to 80 cm)Fixed distance specified for the new device.
    Exam light1 x 11 W, 1 x 19 WRH 600 (1 x 35 W)Different exam light configuration.
    Manual settingLevel 0 to 10, 30 mW/cm² max. at 80 cmRH 600 (Level 0 to 10, 23 mW/cm² max. at 70 cm)Improved maximum irradiance due to better reflector geometry.
    Setting Temp. Range35°C - 37.5 °CRH 600 (35°C - 37°C)Improved performance (wider range).
    Measurement Range15°C to 42 °CRH 600 (33°C - 38°C)Extended measurement range.
    AccuracyLike RH 600 (± 0.3 °C)RH 600Same.
    Applied PartBFRH 600 (B)Improved safety requirement (BF vs B).
    Power FailActuates if there is a loss of supply voltageBabytherm 8000 WB, RH 600Same.
    Phototherapy Option
    TypeIntegrated in radiant warmer housing, fixed distance to patient.PT4000 (separate device), Microlite™ (separate device)Integrated design, fixed distance (80 cm).
    Light source6 halogen bulbsPT4000 (fluorescent bulbs), Microlite™ (3 halogen bulbs)Different number and type of bulbs.
    Power input440 WPT4000 (120 W), Microlite™ (300 W)Higher power input.
    Distance to mattress80 cmPT4000 (variable, 30 cm min), Microlite™ (variable)Fixed distance.
    On/Off switchYesPT4000, Microlite™Same.
    Effective Irradiance12 µW/cm²/nm at 80 cm distance (425..475 nm)PT4000 (13 µW/cm²/nm at 40 cm, 400..500 nm), Microlite™ (17 µW/cm²/nm at 43 cm, 400..500 nm)Different irradiance values and specified wavelength range due to fixed distance.
    Bulb life1000 hoursPT4000, Microlite™Same.
    Hour meterDual timing system for patient exposure time and bulb life.PT4000, Microlite™Same.
    Control System
    TypeMicroprocessor based, digitalBabytherm 8000 WB, RH 600Same.
    Self-CalibratingYesBabytherm 8000 WB, RH 600Same.
    Setpoint-ReadoutDigital, 3 digitBabytherm 8000 WB, RH 600Same.
    Power OnGreen LEDBabytherm 8000 WB, RH 600 (Red LED)Change in LED color to fulfill IEC 601 requirement.
    Safety System
    OvertemperatureFor mattress: like Babytherm 8000 WB (40.0°C). For skin temp.: 39.0°C in skin mode or manual mode if sensor connected.Babytherm 8000 WB (40.0°C)Improved safety feature for radiant heater.
    Overheat ProtectionYesBabytherm 8000 WB, RH 600Same.
    Probe FailFor mattress: like Babytherm 8000 WB. For radiant heater: like RH 600.Babytherm 8000 WB, RH 600Same.
    Set point deviationFor mattress: like Babytherm 8000 WB (± 1°C). For skin temp.: like RH 600 but user-adjustable (0.3℃ to 1.0℃).Babytherm 8000 WB (± 1°C), RH 600 (± 0.5°C)User-adjustable deviation for skin temp.
    Shut off temperatureLike Babytherm 8000 WB (42.0°C)Babytherm 8000 WBSame.
    15 Minute AlarmActuates after 15 minutes in manual mode with level > 3, and heater output is reduced to level 2.RH 600 (Actuates after 15 minutes in manual mode with level > 5)Alarm condition modified based on IEC 601-2-21 requirements (trigger at >10 mW/cm²).
    Visible AlarmYesBabytherm 8000 WB, RH 600Same.
    Audible AlarmYesBabytherm 8000 WB, RH 600Same.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    The document is a 510(k) Pre-market Notification, which focuses on demonstrating substantial equivalence to legally marketed predicate devices. It does not describe a clinical study with a test set, sample size, or data provenance in the way a clinical trial report would. The "testing" referred to is against international standards, not a specific patient cohort for performance evaluation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    Not applicable. As this is not a study involving human interpretation or diagnostic accuracy, there is no "ground truth" established by experts in this context. The document relies on engineering specifications and adherence to international standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. There is no test set or human adjudication described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a medical apparatus (infant warmer), not a diagnostic or AI-powered system that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    Not applicable. The "ground truth" in this context is the adherence to established international safety and performance standards (like IEC 601-2-21) and the functional specifications of predicate devices.

    8. The sample size for the training set:

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable. There is no training set for this type of device.

    Summary of the Study and Device Performance:

    The "study" described in the 510(k) summary is not a traditional clinical trial but rather an engineering and regulatory comparison. The Dräger Babytherm 8004 / 8010 device demonstrates its "acceptance criteria" and "performance" by:

    • Claiming Substantial Equivalence: The primary method is to show that the new device is substantially equivalent in intended use, technological characteristics, safety, and effectiveness to several legally marketed predicate devices.
    • Meeting International Standards: The document explicitly states that the Babytherm 8004 / 8010 "fulfill at least the same international standards as the predicate devices of Dräger and has been tested according to these standards" (e.g., IEC 601-2-21). This adherence to standards is the basis for demonstrating safety and effectiveness.
    • Detailed Comparison Table: The provided "Table of Comparison" outlines various technical specifications and features, highlighting similarities and differences with predicate devices. Where differences exist (e.g., increased height, improved irradiance, wider temperature range, enhanced safety features), the document often explains these as "improved performance" or modifications to meet current standards.

    In essence, the "study" is a regulatory submission that leverages an engineering comparison and compliance with established medical device standards to argue for the device's safety and effectiveness. It does not involve patient data collection or clinical outcomes in the manner of a clinical trial.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1