Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K182799
    Manufacturer
    Date Cleared
    2019-03-08

    (157 days)

    Product Code
    Regulation Number
    888.3660
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Arthrex Univers II Shoulder Prosthesis System: Titanium Humeral Heads

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arthrex UNIVERS II Shoulder Prosthesis is indicated in replacement(s) when conditions including severe pain or significant disability resulting from degenerative, theumatic disease, or injury of the glenohumeral joint; nonunion humeral head fractures of long duration: irreducible 2- and 4-part proximal humeral fractures; of the humeral head, or other difficult clinical management problems where arthrodesis or resectional arthroplasty is not acceptable.

    The glenoid components are intended fixation in the joint and must only be used with appropriate bone cement.

    The Arthrex Titanium Humeral Head is indicated for patients with suspected cobalt alloy sensitivity. The wear properties of Titanium and Titanium alloys are inferior to that of cobalt alloy. A Titanium humeral head is not recommended for patients who lack suspected material sensitivity to cobalt alloy.

    Device Description

    The Titanium humeral heads are intended to be used in the Arthrex Univers II Shoulder Prosthesis System. The humeral heads were originally cleared under predicate K071032 as cobalt-chromium (CoCr) devices. The design of the titanium humeral heads is identical to the CoCr heads cleared under predicate K071032. The only difference between the proposed device and the predicate is the material. The titanium humeral heads will provide an alternative to CoCr for use on patients with nickel allergies.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA regarding a medical device, the Arthrex Univers II Shoulder Prosthesis System: Titanium Humeral Heads. This document describes the device, its intended use, and the basis for its substantial equivalence to a predicate device.

    However, the provided text does not contain information about a study proving the device meets acceptance criteria in the context of an AI-powered diagnostic device, nor does it refer to acceptance criteria or performance metrics such as accuracy, sensitivity, specificity, or AUC, or details about test set size, expert adjudication, or MRMC studies.

    The "Performance Data" section of the document mentions:

    • "Fatigue testing, pull-off testing, and wear testing were conducted to evaluate the strength and fatigue properties of the proposed titanium humeral heads."
    • "Additionally, an engineering analysis was performed to address the substantial equivalence of hardness, adhesion, surface roughness, scratch resistance, young's modulus, and articulating congruency compared to the predicate device."
    • "Bacterial endotoxin per EP 2.6.14/USP was conducted to demonstrate that the device meets pyrogen limit specifications."

    This indicates that the performance data gathered relates to the physical and material properties of the orthopedic implant (a shoulder prosthesis), and its biocompatibility, not the performance of an AI or diagnostic algorithm.

    Therefore, I cannot fulfill your request for information regarding acceptance criteria and performance of an AI device based on this document. The information you've asked for (e.g., sample size for test/training sets, expert qualifications, MRMC studies, ground truth establishment) is specific to the validation of AI/ML-based diagnostic devices, which is not what this FDA notification is about.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1