510(k) Data Aggregation

    K Number
    K244025
    Manufacturer
    Date Cleared
    2025-01-28

    (32 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Argo Knotless GENESYS Anchor

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The biocomposite suture anchor is intended to reattach soft tissue to bone in orthopedic surgical procedures. The Argo Knotless® GENESYS™ Anchor may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization throughout the healing period.

    Device Description

    The Argo Knotless® GENESYS™ Anchor is an implantable bone anchor, that is supplied single use, sterilized via ethylene oxide (ETO) to a SAL of 10 °. The threaded anchor is manufactured of bioabsorbable material, and the suture eyelet is manufactured of PEEK Optima material. The device features a single use driver, a threaded anchor, a PEEK-Optima eyelet, a #2 UHMWPE, non-absorbable retention suture, and loader tab. The retention suture holds the PEEK eyelet in place on the driver and can be incorporated into the repair if desired. The suture eyelet can hold up to six (6) sutures or one (1) bioresorbable reinforced implant. These anchor configurations require a pre-prepared bone hole which can be created with Class I, Exempt instrumentation.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Argo Knotless® GENESYS™ Anchor, a medical device. This document focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed acceptance criteria and study results in the format typically found in clinical trial reports for software devices.

    Based on the information provided, here's what can be extracted and what is not present:

    Missing Information:

    The document does not describe acceptance criteria in the typical statistical sense (e.g., specific thresholds for sensitivity, specificity, or other performance metrics). It also does not describe a study that proves the device meets these types of acceptance criteria. The performance data mentioned refers to mechanical testing and user validation, not a clinical study to establish analytical or clinical performance of an AI/software device.

    Therefore, the following points cannot be answered from the provided text:

    • A table of acceptance criteria and the reported device performance: Not provided in terms of statistical performance metrics.
    • Sample sized used for the test set and the data provenance: Not applicable in the context of AI/software performance. The "test set" here refers to mechanical testing or cadaveric validation.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
    • Adjudication method for the test set: Not applicable.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This device is a physical bone anchor, not an AI/software product that assists human readers.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    • The type of ground truth used: Not applicable for an AI/software performance study. For the mechanical and user validation tests, the "ground truth" would be the physical properties or user feedback.
    • The sample size for the training set: Not applicable as this is not an AI/software device.
    • How the ground truth for the training set was established: Not applicable.

    What is present in the document about performance and validation:

    The document briefly mentions "performance data" in Section VII, but this refers to mechanical and user validation for a physical medical device.

    1. Acceptance Criteria and Reported Device Performance (As implied for a physical device, not statistical AI performance):

    While not explicitly stated as "acceptance criteria," the document indicates that the device's performance was evaluated to demonstrate substantial equivalence to its predicate. The implied acceptance criterion is that the new device performs as intended and similarly to the predicate device in terms of mechanical properties and user experience.

    Acceptance Criterion (Implied)Reported Device Performance
    Performance as intended (mechanical stability, insertion, pull-out)Performance Testing (Insertion, Cyclic and Ultimate Pull Displacement Resistance) completed.
    User validation for surgical proceduresUser Validation (Cadaver Medial Collateral Ligament/Lateral Ligament) completed.
    Packaging and labeling suitabilityPackaging and Labeling User Validation completed.

    2. Study that proves the device meets the (implied) acceptance criteria:

    The study refers to a variety of tests and validations performed:

    • Performance Testing: This would typically involve laboratory tests to measure parameters like insertion force, resistance to cyclic loading, and ultimate pull-out strength. The document doesn't provide specific quantitative results but states these tests were "completed."
    • User Validation (Cadaver Medial Collateral Ligament/Lateral Ligament): This indicates hands-on testing of the device's usability and effectiveness in a cadaveric model, simulating surgical conditions for the intended use (reattaching soft tissue to bone).
    • Packaging and Labeling User Validation: This would involve confirming that the packaging maintains sterility and integrity, and that the labeling is clear and accurate for users.

    Data Provenance and Sample Size for Test Set:

    • Sample Size: The specific sample sizes for these tests (e.g., number of anchors tested, number of cadavers, number of users for validation) are not provided in this summary.
    • Data Provenance: Not specified, but generally, such performance and cadaveric testing would be conducted in a laboratory or simulated surgical environment. It is not "data" in the sense of patient records from a specific country.

    In summary, the provided document is a regulatory submission for a physical medical device. It does not contain the specific information requested about acceptance criteria and study design as it pertains to the performance validation of an AI/software device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240090
    Manufacturer
    Date Cleared
    2024-03-05

    (53 days)

    Product Code
    Regulation Number
    888.3030
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Argo Knotless GENESYS Anchor

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Argo Knotless® GENESYS™ SP Anchor
    The biocomposite suture anchor is intended to reattach soft tissue to bone in orthopedic surgical procedures. The Argo Knotless® GENESYS™ SP Anchor may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaqed soft tissue, in conjunction with appropriate postoperative immobilization throughout the healing period.

    Argo Knotless® GENESYS™ Anchor
    The biocomposite suture anchor is intended to reattach soft tissue to bone in orthopedic surgical procedures. The Argo Knotless® GENESYS™ Anchor may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization throughout the healing period.

    Device Description

    Argo Knotless® GENESYS ™ SP Anchor: The Self-Punching (SP) Argo Knotless® GENESYS™ Anchor is an implantable bone anchor, that is supplied for single use, sterilized via ethylene oxide (ETO) to a SAL of 10 °. The threaded anchor is manufactured of bioabsorbable material, and the broaching tip (suture eyelet) is manufactured of titanium material. Each size features a single-use driver, the threaded anchor, a Titanium suture eyelet, a UHMWPE 1.0mm non-absorbable retention suture, and loader tab. The retention suture holds the titanium eyelet in place on the driver assembly and can be incorporated into the repair if desired. The suture eyelet can hold up to six (6) sutures.

    Argo Knotless® GENESYS ™ Anchor: The Non-Self-Punching Argo Knotless® GENESYS™ Anchor is an implantable bone anchor, that is supplied single use, sterilized via ethylene oxide (ETO) to a SAL of 10 °. The anchor configuration requires a pre-prepared bone hole. The threaded anchor is manufactured of bioabsorbable material, and the suture eyelet is manufactured of PEEK material. Each size features a single use driver, a threaded anchor, a PEEK suture eyelet, a #2 UHMWPE, non-absorbable retention suture, and loader tab. The retention suture holds the PEEK eyelet in place on the driver and can be incorporated into the repair if desired. The suture eyelet can hold up to six (6) sutures. These anchor configurations require a pre-prepared bone hole which can be created with Class I, Exempt instrumentation.

    Argo Knotless® GENESYS ™ Anchor: The Preloaded Argo Knotless® GENESYS™ Anchor is an implantable bone anchor, that is supplied single use, sterilized via ethylene oxide (ETO) to a SAL of 10 °. The threaded anchor is manufactured of bioabsorbable material, and the suture eyelet is manufactured of PEEK material. The eyelet is held on with a #2 UHMWPE, nonabsorbable retention suture and supplied with either a white/black or blue/blue colored non-absorbable Hi-Fi suture tape, a retention suture, and loader tab. These anchor configurations require a pre-prepared bone hole which can be created with Class I, Exempt instrumentation.

    Principle of Operation - soft tissue to bone fixation

    AI/ML Overview

    This document is an FDA 510(k) clearance letter and summary for a medical device (Argo Knotless GENESYS Anchor). It does not describe a study involving an AI/software as a medical device (SaMD) or diagnostic imaging. Therefore, the requested information about acceptance criteria, study design, ground truth establishment, and human-in-the-loop performance is not applicable to this document.

    The document focuses on demonstrating substantial equivalence of a physical medical device (suture anchor) to a previously cleared predicate device. The "Performance Data" section lists various engineering and material tests conducted on the physical device, not clinical performance or diagnostic accuracy studies.

    Here's why the questions are not applicable:

    • No AI/Software Component: The device described is a physical implantable suture anchor, not an AI or software-based diagnostic tool.
    • No Diagnostic Performance: The performance data mentioned (e.g., "Performance Testing," "Magnetic Resonance," "Transportation," "Corrosion Susceptibility") relate to the mechanical, material, and sterility aspects of the anchor, not its diagnostic accuracy in interpreting images or data.
    • No Human Reader Interaction: There's no scenario where human readers (like radiologists) would "improve with AI vs. without AI assistance" as this is not an imaging or diagnostic AI product.
    • No Ground Truth for AI: Ground truth for a physical device like a suture anchor is typically established through engineering specifications, material properties, and in-vitro or in-vivo (animal or human cadaver) biomechanical testing, not through expert consensus on images or pathology.

    Therefore, I cannot populate the table or answer the specific questions related to AI/SaMD performance criteria, study design, and ground truth. The provided text simply doesn't contain that type of information.