Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K242547
    Manufacturer
    Date Cleared
    2025-05-19

    (265 days)

    Product Code
    Regulation Number
    868.5905
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AirFit F20 Mask System:

    The AirFit F20 mask has two product variants:

    • AirFit F20 mask is intended for single-patent reuse in the home environment.
    • AirFit F20 SLM (Sleep Lab Mask) variant is intended for multi-patient reuse in the hospital/institutional environment.

    Both masks are intended for patients weighing more than 66lb (30kg), who have been prescribed non-invasive CPAP or bi-level positive airway pressure (PAP) therapy. The Sleep Lab Mask is the only variant that is validated and intended for multi-patient reprocessing and must be reprocesssed if reused between patients.

    AirFit F20 NM Mask System:

    The AirFit F20 Non Magnetic mask has two product variants:

    • The AirFit F20 Non Magnetic variant is intended for single-patient reuse in the home environment.
    • The AirFit F20 Non Magnetic SLM (Sleep Lab Mask) variant is intended for multi-patient reuse in the hospital/institutional environment.

    This mask is for patients weighing more than 66 lb (30 kg), who have been prescribed non-invasive CPAP or bi-level positive airway pressure (PAP) therapy. The Sleep Lab Mask is the only variant that is validated and intended for multi-patient reprocessing and must be reprocessed if reused between patients.

    Device Description

    The AirFit F20 and AirFit F20 NM Mask Systems are both full-face masks intended for use with Continuous Positive Airway Pressure (CPAP) and/or Bi-Level devices. CPAP & Bi-Level devices are used for the treatment of Obstructive Sleep Apnea (OSA) and/or ventilatory support. Both treatments deliver pressurized air from a flow generator to the patient. Such delivery of pressurized air relies on the Patient Interface; commonly known as the mask. Full face masks are a common type of Patient Interface (Mask) and are known to efficaciously support the delivery of pressurized air by providing a seal around the patient's nose and mouth.

    The AirFit F20 mask system has two product variants:

    • AirFit F20 Mask: This is the home use variant that is intended for single patient re-use.
    • AirFit F20 SLM (sleep lab mask): This is the variant that is intended for multi-patient re-use and must be reprocessed if reused between patients using the disinfection guide.

    AirFit F20 (K170924) had a "For Her" variant which was cleared with aesthetic colour pops, the AirFit F20 Mask and AirFit F20 SLM also have the same "For Her" variant.

    The AirFit F20 NM mask system has two product variants:

    • AirFit F20 NM Mask: This is the home use variant that is intended for single patient re-use.
    • AirFit F20 NM SLM (sleep lab mask): This is the variant that is intended for multi-patient re-use and must be reprocessed if reused between patients using the disinfection guide.

    The AirFit F20 and AirFit F20 NM Mask Systems share a common design architecture as the predicate AirFit F20 (K170924), except that the AirFit F20 NM Mask System has non-magnetic clips.

    The AirFit F20 and AirFit F20 NM mask systems are made up of the following four main component assemblies: cushion, elbow, frame, and headgear. The cushion and headgear are available in various sizes to allow for adequate mask fit over the intended patient population.

    The AirFit F20 and AirFit F20 NM mask systems are prescription devices supplied non-sterile.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and associated summary pertain to the AirFit F20 Mask Systems, which are non-continuous ventilators (CPAP/Bi-Level masks). This type of device does not typically involve complex diagnostic algorithms or AI components that would necessitate the study structure outlined in your request (e.g., test sets, ground truth, expert adjudication, MRMC studies, standalone performance for AI).

    Instead, the clearance is based on demonstrating substantial equivalence to a predicate device (AirFit F20, K170924) through non-clinical performance and safety testing, adherence to recognized standards, and biocompatibility evaluations.

    Therefore, many of the specific questions you've asked cannot be directly answered from this document as they are not relevant to the type of device and clearance process involved.

    Here's an attempt to extract the relevant information and address your questions where possible:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a "table of acceptance criteria" with corresponding "reported device performance" in a format typically used for diagnostic algorithm validation. Instead, it compares specific design parameters and performance characteristics of the new devices (AirFit F20 Mask System and AirFit F20 NM Mask System) against the predicate device (AirFit F20, K170924). The "Comments" column often indicates that the performance is "Equivalent" or "Identical," implying that the new devices meet or perform similarly to the established safety and performance profile of the predicate, which serves as the de facto acceptance benchmark.

    Here's a summary derived from the comparison table (Pages 9-12 of the provided document):

    Design parameter or featurePredicate device: AirFit F20, K170924 (Performance/Criteria based on predicate)Modified device: AirFit F20 Mask System (Reported Performance)Modified device: AirFit F20 NM Mask System (Reported Performance)Acceptance Criteria (Implied)
    Indications for UseSpecified for CPAP/bi-level, >66lb (30kg), single-patient home, multi-patient hospital use.Equivalent to predicate, with variants for home/sleep lab use.Equivalent to predicate, with variants for home/sleep lab use.Equivalent to predicate
    Intended UseInterface for CPAP/bi-level devices.IdenticalIdenticalIdentical to predicate
    FDA Product CodeBZDBZDBZDIdentical to predicate
    Patient populationWeighing more than 66 lb (30 kg)Patients weighing more than 66lb (30kg)Patients weighing more than 66lb (30kg)Identical to predicate
    Environment of UseHome or health institutionHome or health institutionHome or health institutionIdentical to predicate
    Reprocessing claimsSingle patient re-use or multi-patient re-use.IdenticalIdenticalIdentical to predicate
    Sterility state as providedNon-sterileNon-sterileNon-sterileIdentical to predicate
    Validated reprocessing methodsHigh-Level Thermal disinfection, SterilizationHigh-Level Thermal disinfection, SterilizationHigh-Level Thermal disinfection, SterilizationIdentical to predicate
    Vent typeMulti-hole ventMulti-hole vent and diffuser ventMulti-hole vent and diffuser ventEquivalent to predicate's safety/efficacy profile
    PAP tubing connection pointISO 5356-1 (22mm)ISO 5356-1 (22mm)ISO 5356-1 (22mm)Identical to predicate
    Construction materialPolymeric, Textile, magnetsPolymeric, Textile, magnetsPolymeric, TextileEquivalent to predicate's safety/efficacy profile (NM variant removes magnets)
    Operating pressure range (cmH2O)3 - 403 – 403 – 40Identical to predicate
    SizesCushion (3), Headgear (3), Frame (1)IdenticalIdenticalIdentical to predicate
    Mask exhaust flow (Nominal) ISO 17510:2015 Annex BPressure (cm H2O): 3, 20, 40 -> Flow (L/min) 'Full Face' curve: 20.1, 55.5, 82.4Pressure (cm H2O): 3, 20, 40 -> Flow (L/min) 'Full Face' curve: 19, 54, 82Pressure (cm H2O): 3, 20, 40 -> Flow (L/min) 'Full Face' curve: 19, 54, 82Equivalent to predicate (minor differences within acceptable range for substantial equivalence)
    CO2 rebreathing performance (normal condition) ISO 17510:2015 Annex FPressure (cm H2O): 3, 5, 10 -> Relative CO2 increase Non-SLM: 14.8%, 7.5%, 6.8% SLM: 11.7%, 7.3%, 2.9%Pressure (cm H2O): 3, 5, 10 -> Relative CO2 increase Non-SLM: ≤ 11.4%, ≤ 11.4%, - SLM: 11.7%, 7.3%, 2.9%Pressure (cm H2O): 3, 5, 10 -> Relative CO2 increase Non-SLM: ≤ 11.4%, ≤ 11.4%, - SLM: 11.7%, 7.3%, 2.9%Equivalent (<20%) (as per ISO standard for substantial equivalence)
    CO2 rebreathing performance (single fault condition) ISO 17510:2015 Annex FFault (ISO 17510:2015) -> Relative CO2 increase Non-SLM: 32.3%, 39.5% SLM: 40.5%, 52.7%Fault (ISO 17510:2015) -> Relative CO2 increase Non-SLM: ≤ 53.1%, ≤ 53.1% SLM: 40.5%, 52.7%Fault (ISO 17510:2015) -> Relative CO2 increase Non-SLM: ≤ 53.1%, ≤ 53.1% SLM: 40.5%, 52.7%Equivalent (<60%) (as per ISO standard for substantial equivalence)
    Resistance to flow (Pressure drop across mask in cmH2O) ISO 17510:2015 Annex C@50 L/min: <0.8 cmH2O @100 L/min: <1.6 cmH2O@50 L/min: <0.8 cmH2O @100 L/min: <1.6 cmH2O@50 L/min: <0.8 cmH2O @100 L/min: <1.6 cmH2OEquivalent (Identical performance)
    Breathing Resistance (cmH2O) ISO 17510:2015 Annex EInspiratory: 0.44 @ 50L/min Expiratory: 0.60 @ 50L/minInspiratory: 0.6 @ 50L/min Expiratory: 0.80 @ 50L/minInspiratory: 0.6 @ 50L/min Expiratory: 0.80 @ 50L/minEquivalent (within acceptable limits for substantial equivalence)
    AAV Operating Pressures ISO 17510:2015 Annex DDe-activation: 1.5 cm H2O Activation: 1.0 cm H2ODe-activation: 2.0 cm H2O Activation: 1.5 cm H2ODe-activation: 2.0 cm H2O Activation: 1.5 cm H2OEquivalent (within acceptable limits for substantial equivalence)
    Flow generator setting on compatible ResMed CPAP and Bi-level flow generators"Full face""Full face""Full face"Identical to predicate
    SoundSound power level: 30 dBA Sound pressure level: 23 dBAMulti-hole vent: 30 dBA power, 22 dBA pressure. Diffuser vent: 21 dBA power, 13 dBA pressure.Multi-hole vent: 30 dBA power, 22 dBA pressure. Diffuser vent: 21 dBA power, 13 dBA pressure.Equivalent to predicate (Diffuser vent improves sound)
    Operating and storage temperatureOperating: 5°C to 40°C Storage: -20°C to +60°CIdenticalIdenticalIdentical to predicate
    Magnetic ClipsNumber: 4, Surface Strength: <400 mT, 50mm distance: <0.5 mTNumber: 4, Surface Strength: <400 mT, 50mm distance: <0.5 mTNumber: 0, Surface Strength: N/A, 50mm distance: N/AEquivalent (NM variant removes magnets for specific patient needs)

    2. Sample size used for the test set and the data provenance

    The document states that "Non-clinical verification and validation testing completed for the AirFit F20 and AirFit F20 NM Mask Systems demonstrated that the mask systems met all intended performance requirements." This primarily refers to bench testing and in-vitro studies.

    • Sample Size: Not explicitly stated as a number of "cases" or "patients" in the context of diagnostic performance. For physical device performance, testing involves multiple units of the product to ensure consistency and reliability, but specific sample sizes for tests (e.g., number of masks tested for exhaust flow, CO2 rebreathing, resistance) are not detailed in this summary.
    • Data Provenance: This is not a clinical study in the typical sense. The data comes from non-clinical bench testing conducted by the manufacturer (ResMed Pty Ltd, Australia, with a US office). It is not retrospective or prospective patient data from a particular country, but rather laboratory test results.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable. The device is a medical mask system, not a diagnostic algorithm that requires expert interpretation for a "ground truth." The "ground truth" for performance is established by recognized international standards (e.g., ISO 17510:2015 for mask performance, ISO 10993 for biocompatibility) and engineering specifications. Compliance is measured against these standards, not expert consensus on medical images or diagnoses.

    4. Adjudication method for the test set

    This question is not applicable for the same reasons as #3. Performance is determined by objective measurements against established standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. This clearance is for a physical medical device (CPAP/Bi-Level mask), not an AI-powered diagnostic tool. There are no "human readers" or "AI assistance" in the context described for this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is based on:

    • Compliance with international standards: For physical performance (e.g., CO2 rebreathing, exhaust flow, resistance, AAV pressures, sound) as defined by ISO 17510:2015 and other relevant ISO standards.
    • Biocompatibility testing: According to ISO 10993 series and ISO 18562 series.
    • Mechanical integrity testing: To ensure durability after various environmental stressors (home cleaning, transport, storage, operation, free fall, reprocessing).
    • Cleaning and reprocessing validation: Against standards like ISO 17664-1, ISO 17664-2, ST98, and AAMI TIR12.

    Essentially, the ground truth is objective measurement against established engineering and safety standards.

    8. The sample size for the training set

    This question is not applicable. There is no "training set" as this is not a machine learning or AI algorithm.

    9. How the ground truth for the training set was established

    This question is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1