Affixus Tibial and Antegrade Femoral Nailing System
The Affixus® Tibial Nailing System is indicated for temporary stabilization and fixation of tibial fractures and osteotomies including proximal. metaphyseal. and distal shaft fractures. closed fractures, open fractures, pathologic fractures, non-unions, malunions and deformity corrections.
The Affixus® Antegrade Femoral Nailing System is indicated for temporary stabilization and fixation of femoral fractures and osteotomies including proximal, metaphyseal, and distal shaft fractures, trochanteric and subtrochanteric fractures, closed fractures, pathologic fractures, non-unions, malunions and deformity corrections.
The Affixus Tibial Nailing System and the Affixus Antegrade Femoral (AF) Nailing System are long bone nailing systems that offer implants designed to treat a range of tibial and femoral fractures from simple to complex, with versatile locking options. They restore the shape of preinjured bone and are available in a variety of lengths and diameters to meet assorted anatomical needs. Screws and End Caps are included in the system. All implants are made of Ti-6Al-4V. The Affixus platform of instrumentation is designed to provide options and flexibility for many intraoperative approaches while maintaining ease of use and commonality.
This is a 510(k) premarket notification for a medical device, specifically the Affixus Tibial and Antegrade Femoral Nailing System. These notifications are submitted to the FDA to demonstrate that a device is substantially equivalent to a legally marketed predicate device, meaning it is at least as safe and effective.
The provided text does not contain the detailed information necessary to complete all sections of your request regarding acceptance criteria and a study that proves the device meets those criteria, particularly in the context of an AI/algorithm-based device. This document is for a traditional medical implant, not an AI software as a medical device (SaMD).
Therefore, many of the requested fields are not applicable or cannot be extracted from the given text. However, I can provide information based on what is available.
Here's the breakdown of what can be inferred and what is missing:
Acceptance Criteria and Device Performance Study
As this is a 510(k) for a physical medical implant (intramedullary fixation rod), the concept of "acceptance criteria" and "device performance" in the context of an AI/algorithm is not directly applicable in the same way. Instead, the "acceptance criteria" can be interpreted as demonstrating substantial equivalence to predicate devices through various non-clinical tests. The "device performance" is demonstrated by meeting the standards and benchmarks set by these tests.
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria Category (Implied) | Specific Tests Performed (Reported Device Performance) |
---|---|
Mechanical Strength & Durability | - Screw Disengagement |
- Construct Bending Fatigue Strength | |
- Nail Static Bending Strength/Stiffness per ASTM F1264 Annex 1 | |
- Nail Bending Fatigue Strength per ASTM F1264 Annex 3 | |
- Nail Static Torsional Stiffness per ASTM F1264 Annex 2 | |
- Bone Screw Bending Strength per ASTM F1264 Annex 4 | |
Screw Properties | - Locking Mechanism Torque |
- Screw Maximum Torque to Failure and Breaking Angle per ASTM F543 Annex 1 | |
- Screw Insertion/Removal Torque per ASTM F543 Annex 2 | |
- Screw Pullout Strength per ASTM F543 Annex 3 | |
- Screw Self Tapping Performance per ASTM F543 Annex 4 | |
Biocompatibility/Safety | - MR Compatibility Evaluation per ASTM F2182, ASTM F2052, F2119, ASTM 2213 |
Overall Equivalence | - Demonstrated similar intended use and technological characteristics to predicates, with differences not raising new questions of safety/effectiveness. |
- Proposed device is at least as safe and effective as legally marketed predicate devices. |
Explanation of Performance: The document explicitly states: "The information provided herein demonstrates that: any differences do not raise different questions of safety and effectiveness; and the proposed device is at least as safe and effective as the legally marketed predicate devices." This is the ultimate "performance" metric for a 510(k) submission, achieved by passing the listed non-clinical tests and demonstrating material, design, packaging, and sterilization similarity to predicate devices. The specific values or pass/fail thresholds for each test are not provided in this summary, but the successful submission implies they were met.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: Not specified in the provided text. For mechanical testing of implants, sample sizes typically adhere to relevant ASTM standards.
- Data Provenance: Not specified in the provided text. These are non-clinical lab tests, so country of origin for "data" is typically the testing lab's location. The tests are prospective in the sense that they are specifically conducted for the submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. This is a physical implant, not an AI/algorithm requiring expert-established ground truth on image data or other patient data. The "ground truth" for these tests are the defined mechanical properties and physical characteristics as per the ASTM standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. This is a physical implant. The mechanical tests are objective measurements, not subject to adjudication by human interpreters.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This is for a physical implant. No AI component is mentioned, and therefore, no MRMC study or human reader improvement with AI assistance is relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is for a physical implant. No algorithm is mentioned.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
- Mechanical Test Standards: The "ground truth" for the non-clinical tests are the specified mechanical properties and performance requirements outlined in the referenced ASTM standards (e.g., ASTM F1264, ASTM F543, ASTM F2182, etc.). These standards define acceptable ranges for strength, fatigue, torque, etc., which the device must meet or exceed.
8. The sample size for the training set
- Not Applicable. This is for a physical implant; there is no "training set" in the context of an AI/machine learning model.
9. How the ground truth for the training set was established
- Not Applicable. As there is no training set, this question is not relevant.
Summary of what the document does indicate for this physical device:
The submission highlights that the Affixus Tibial and Antegrade Femoral Nailing System
is deemed substantially equivalent to predicate devices based on:
- Similar indications for use.
- Similar technological characteristics (materials, design features, packaging, sterilization).
- Successful completion of a battery of non-clinical mechanical and MR compatibility tests.
The goal of this 510(k) is to demonstrate that the new device does not raise new questions of safety or effectiveness compared to legally marketed predicate devices. The listed non-clinical tests are the evidence used to support this claim.