510(k) Data Aggregation

    K Number
    K161742
    Manufacturer
    Date Cleared
    2016-11-15

    (144 days)

    Product Code
    Regulation Number
    888.3660
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Aequalis PerFORM Reversed, Aequalis PerFORM+ Reversed Glenoid

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Aequalis PerFORM Reversed & Aequalis PerFORM+ Reversed Glenoid are indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and non-repairable rotator cuff-tear with pain disabled by:

    • · Rheumatoid arthritis
    • · Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
    • · Correction of functional deformity
    • · Fractures of the humeral head
    • · Traumatic arthritis
    • · Revision of the devices if sufficient bone stock remains.
    Device Description

    The Aequalis™ PerFORM Reversed and Aequalis™ PerFORM+ Reversed Glenoid are intended to replace the glenoid part of the scapulohumeral joint as part of a reverse shoulder prosthesis. Aequalis™ PerFORM Reversed and Aequalis™ PerFORM+ Reversed glenoid component assembly must be used in association with a Tornier humeral component:

    • Humeral implants Aequalis Reversed, Aequalis Reversed Fracture or Aequalis Adjustable ● Reversed Shoulder
    • . or Humeral implants Aequalis Ascend Flex Shoulder System in reverse configuration
    • or humeral implants Aequalis Reversed FX2 ●

    The glenoid component assembly includes a modular system including a baseplate with central screw or press-fit post, peripheral anchoring screws and a glenosphere. The baseplates are available in various design configurations including: standard, lateralized, half-wedge augment and full-wedge augment. The glenospheres are available in various design configurations including: standard, lateralized and eccentric. The press-fit post may be used with the standard baseplate in lieu of the central anchoring screw. The baseplates and the press-fit posts are manufactured using additive manufacturing technology.

    AI/ML Overview

    The document provided does not describe an acceptance criteria or study that proves the device meets specific performance metrics in the way a clinical or standalone study for an AI/algorithm-based device would.

    Instead, this is an FDA 510(k) Premarket Notification for a medical device (Aequalis PerFORM Reversed and Aequalis PerFORM+ Reversed Glenoid – a shoulder prosthesis). The purpose of a 510(k) submission is to demonstrate "substantial equivalence" to a legally marketed predicate device, not necessarily to meet pre-defined performance acceptance criteria as would be typical for a diagnostic or AI device.

    Here's how the information provided relates to your request, with an emphasis on what's not relevant for a typical AI/algorithm acceptance criteria study:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Not explicitly stated as pass/fail numeric targets for performance metrics like sensitivity, specificity, or AUC, as these are not relevant for a mechanical prosthesis 510(k). The "acceptance criteria" here is that the device is "substantially equivalent" to predicate devices, meaning it is as safe and effective.
    • Reported Device Performance: The document lists "Performance Data" which consists of a series of non-clinical bench tests (mechanical testing). These demonstrate the physical properties and durability of the prosthesis. Examples include:
      • Compressive Strength and Elastic Modulus Testing of Titanium Porous Metal Verification
      • Titanium Porous Structure Stereological Evaluation
      • Titanium Porous Structure Mechanical Verification
      • PerFORM Reversed Fixation Pull-out Test
      • Fatigue Test
      • Range of Motion Analysis
      • PerFORM Reversed Taper Pull-Off Test
      • Glenoid Loosening Test (100,000 cycles)
      • PerFORM+ Reversed Augment Baseplate compared to the Exactech Equinoxe Reverse Augmented Baseplate Analysis
      • PerFORM Reversed Lateralization Comparison
      • PerFORM Reversed and PerFORM+ Reversed Peripheral Screw Angulation
      • Direct Metal Laser Sintering (DMLS) Process Validation
      • Endotoxin (