Search Results
Found 1 results
510(k) Data Aggregation
(102 days)
Advantage-C PEEK Cervical lnterbody Fusion Device
The Advantage-C PEEK Cervical Interbody Fusion Device is a cervical interbody fusion device indicated for use in skeletally mature patients with degenerative disease (DDD) with / without radicular symptoms at one level from C2 - T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The Advantage-C PEEK Cervical Interbody Fusion Device is intended for use with supplemental fixation systems and with autogenous bone graft.. The Advantage-C PEEK Cervical Interbody Fusion Device is supplied sterile and is intended for one-time use.
The proposed Intelivation Advantage-C Cervical Interbody Fusion Device is a sterile, single use implant grade polyetheretherketone (PEEK; ASTM F2026) device and Tantalum (ASTM F560), available in varied footprints and heights, designed for supplemental stabilization of the cervical spinal column in anterior cervical discectomy and fusion (ACDF) procedures. The device is intended for spinal fusion procedures at one level in skeletally mature patients with degenerative disc disease (defined as pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. The implant should be used with supplemental spinal fixation systems and with autogenous bone graft.
The provided text describes a 510(k) premarket notification for the "Advantage-C PEEK Cervical Interbody Fusion Device." This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than conducting extensive clinical efficacy studies typically required for a new device. Therefore, the information provided for acceptance criteria and studies will be geared towards non-clinical performance and biological compatibility.
Here's the breakdown of the information based on the provided text:
Acceptance Criteria and Study to Prove Device Meets Acceptance Criteria
1. A table of acceptance criteria and the reported device performance:
Requirement | Acceptance Criteria (Implicit) | Reported Device Performance |
---|---|---|
Subsidence Testing (ASTM F2267-04) | Characterization of subsidence as defined in ASTM F2267-04 | "Subsidence characterization as defined in ASTM F2267-04" (Implies successful characterization, but explicit "Pass" not stated) |
Monotonic Compressive and Torsional Strength (ASTM F2077-11) | Yield and/or ultimate force and moment characterization as defined in ASTM F2077-11 | PASS |
Dynamic Compressive and Torsional Strength (ASTM F2077-11) | Maximum runout (5x10^6 cycles) force or overall fatigue resistance as defined in ASTM F2077-11 | PASS |
Compressive and Torsional Wear Particle Analysis (ASTM F1877-16) | Wear particle size, shape, and distribution within the reported range of PEEK wear particles as defined in ASTM F1877-16 | PASS |
Sterilization Validation (ISO/AAMI/ANSI 11137-1:2006) | Sterilization of health care products. Radiation. Requirements for development, validation and routine control of a sterilization process. (Implicit: achieve SAL of 10^-6) | Pass (An SAL of 10^-6 is explicitly mentioned in the text for the method employed) |
LAL (Endotoxin) (ANSI/AAMI ST72:2019, USP, USP, EP 2.6.14 and JP 4.01) | Meet specified standards for endotoxin levels | Pass |
Biocompatibility (ISO 10993-1:2018) | Demonstrate device is safe for permanent contact (>30 days) implantation | Established for the predicate device (Refer to MAF #2735) |
Notes on Acceptance Criteria: For some tests (like subsidence), a "characterization" is performed, implying that the results are within expected limits for the material and design, even if an explicit "Pass" is not stated in the summary. The core acceptance criterion for a 510(k) is often "substantial equivalence" to a predicate device, which these non-clinical tests help to demonstrate.
2. Sample sizes used for the test set and the data provenance:
- Sample Size for Test Set: Not explicitly stated for any of the non-clinical tests. The document refers to "the proposed device" and "this testing" which implies standard sample sizes for each test method as per the referenced ASTM and ISO standards would have been used, but specific numbers are not given.
- Data Provenance: The studies are non-clinical (laboratory tests). No country of origin is mentioned for the data, but the submitter is Intelivation, LLC, based in Saint Simons Island, Georgia, USA. The studies are prospective in the sense that they were conducted specifically for this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This question is not applicable to the non-clinical performance and sterilization tests described. Ground truth, in the clinical sense, is not established for these types of engineering and material property tests. The "ground truth" for these tests is defined by the objective measurement criteria and methodologies specified in the referenced ASTM and ISO standards.
4. Adjudication method for the test set:
Not applicable. The tests are objective, laboratory-based physical and chemical evaluations following standardized protocols. There is no human adjudication process involved as there would be with clinical image interpretation or diagnosis.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This device is a physical interbody fusion implant, not an AI-assisted diagnostic or therapeutic tool. No MRMC study was conducted or is relevant for this product.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
Not applicable. This device is an implantable medical device, not an algorithm or software.
7. The type of ground truth used:
For the non-clinical tests, the "ground truth" is based on the objective measurements and performance specifications defined by the referenced industry standards (ASTM, ISO). For example, for strength tests, the ground truth is the measured yield/ultimate force or fatigue resistance compared against a predefined standard. For wear particle analysis, it's the measured size, shape, and distribution. For sterilization, it's the achievement of a specific sterility assurance level (SAL).
8. The sample size for the training set:
Not applicable. There is no concept of a "training set" for physical medical device performance testing in this context. The manufacturing process and quality control would involve internal testing, but not a "training set" in the machine learning sense.
9. How the ground truth for the training set was established:
Not applicable, as there is no "training set" for these non-clinical tests.
Ask a specific question about this device
Page 1 of 1