Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K223503
    Date Cleared
    2023-01-19

    (58 days)

    Product Code
    Regulation Number
    862.1825
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Access 25(OH) Vitamin D Total

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access 25(OH) Vitamin D Total assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of total 25-hydroxyvitamin D [25(OH) vitamin D] levels in human serum and plasma using the DxI Access Immunoassay Analyzers. Results are to be used as an aid in the assessment of vitamin D sufficiency.

    Device Description

    The Access 25(OH) Vitamin D Total assay is a competitive binding immunoenzymatic assay. The Access 25(OH) Vitamin D Total assay consists of the reagent pack and calibrators. Other items needed to run the assay include substrate and wash buffer. The Access 25(OH) Vitamin D Total assay reagent pack, Access 25(OH) Vitamin D Total assay calibrators, along with the UniCel Dxl Wash Buffer II are designed for use with the Dxl 9000 Access Immunoassav Analyzer in a clinical laboratory setting.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Access 25(OH) Vitamin D Total assay, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Target or Range)Reported Device Performance (Access 25(OH) Vitamin D Total on Dxl 9000 Access)
    Method Comparison: R²≥ 0.900.97
    Method Comparison: Slope1.00 ± 0.101.05 (with 95% CI of 0.99 – 1.10)
    Precision (Within-laboratory total % CV for > 15.0 ng/mL)Not explicitly stated as acceptance criteria, but implied to be within acceptable clinical limits.Ranged from 4.3% to 9.0%
    Precision (Within-laboratory total SD for ≤ 15.0 ng/mL)Not explicitly stated as acceptance criteria, but implied to be within acceptable clinical limits.Ranged from 0.09 to 2.2
    LinearityLinear throughout the analytical measuring interval.Linear throughout the analytical measuring interval of approximately 7.0-120 ng/mL.
    Limit of Blank (LoB)Not explicitly stated as acceptance criteria, but implied that the determined LoB should be low enough for clinical utility.2.5 ng/mL
    Limit of Detection (LoD)Not explicitly stated as acceptance criteria, but implied that the determined LoD should be low enough for clinical utility.4.5 ng/mL
    Limit of Quantitation (LoQ)≤ 20% within-lab CV7.0 ng/mL (with ≤ 20% within-lab CV)

    2. Sample Size Used for the Test Set and Data Provenance

    • Method Comparison: 150 serum samples. The provenance of these samples (country of origin, retrospective/prospective) is not specified in the provided text.
    • Precision: Multiple serum samples were tested in duplicate. The exact number of distinct serum samples beyond Sample 1-6 (which represent different concentration levels) is not specified, but each of the 6 samples was tested with N=80 replicates. The provenance is not specified.
    • Limit of Blank (LoB): 120 replicates of a zero-analyte sample. The provenance is not specified.
    • Limit of Detection (LoD): 8 to 11 serum samples containing low levels of Vitamin D analyte. These samples were tested over five days with one run per day and nine replicates per run for each pack lot, resulting in ≥ 40 replicates minimally required for LoD estimation for each sample on each pack lot tested. The provenance is not specified.
    • Limit of Quantitation (LoQ): 10-13 serum samples containing low levels of Vitamin D analyte. Samples were tested in replicates of nine per run with one run per day and five total days on each pack lot and instrument, resulting in a minimum of 40 replicates for each sample on each pack lot. The provenance is not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This device is an in vitro diagnostic (IVD) assay for quantitative determination of a biomarker (25-hydroxyvitamin D). The ground truth for such assays is established through analytical methods and reference standards, not typically by human expert consensus or pathology review in the way a diagnostic imaging AI might be evaluated.

    Therefore:

    • Number of experts: Not applicable.
    • Qualifications of experts: Not applicable.

    The assay is traceable to NIST-Ghent ID-LC-MS/MS, which serves as a highly accurate reference method for establishing the true concentration of Vitamin D in samples.

    4. Adjudication Method for the Test Set

    Not applicable for this type of IVD assay. The performance is assessed by comparing results directly to reference methods or statistical metrics, not through expert adjudication of qualitative or semi-quantitative interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an automated in vitro diagnostic assay, not an AI-assisted diagnostic imaging tool that would involve human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the studies described (Method Comparison, Precision, Linearity, LoB, LoD, LoQ) evaluate the performance of the device (assay on the analyzer) in a standalone fashion, without human interpretation or intervention beyond running the tests according to protocol.

    7. The Type of Ground Truth Used

    • Method Comparison: The predicate device (Access 25(OH) Vitamin D Total assay run on the Access 2 Immunoassay System, K142373) serves as the comparator or "reference" for this study. While not an absolute ground truth, it's the established method against which the new device on the new instrument is compared for substantial equivalence.
    • Precision, Linearity, LoB, LoD, LoQ: These studies rely on the inherent properties of the assay and statistical methodologies (e.g., CLSI guidelines) to characterize the device's analytical performance. The reference standard for traceability for the analyte itself is NIST-Ghent ID-LC-MS/MS. Low-level and zero-analyte samples are used to establish limits.

    8. The Sample Size for the Training Set

    This document describes the validation studies for a new IVD assay and instrument combination. It does not mention a "training set" in the context of machine learning, as this is a chemical immunoassay, not an AI/ML-based device. The term "training set" is not relevant here.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    Access 25(OH) Vitamin D Total for Use on the Access 2 Immuno.
    Syst., Access 25(OH) Vitamin D Total Calibrators for Use on the Access 2 Immuno. Syst

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access 25(OH) Vitamin D Total assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of total 25-hydroxyvitamin D [25(OH) vitamin D] levels in human serum and plasma using the Access 2 Immunoassay Systems. Results are to be used as an aid in the assessment of vitamin D sufficiency.

    The Access 25(OH) Vitamin D Total Calibrators are intended to calibrate the Access 25(OH) Vitamin D Total assay for the quantitative determination of total 25-hydroxyvitamin D [25(OH) vitamin D] levels in human serum and plasma using the Access 2 Immunoassay Systems.

    Device Description

    The Access 25(OH) Vitamin D Total for use on the Access 2 Immunoassay System, Access 25(OH) Vitamin D Total Calibrators for use on the Access 2 Immunoassay System, and the Access 2 Immunoassay analyzer comprise the Access Immunoassay System for the quantitative determination of total 25-hydroxyvitamin D[25(OH) vitamin D] levels in human serum and plasma.

    The Access 25(OH) Vitamin D Total assay is a two-step sequential competitive binding immunoenzymatic assay. In the initial incubation, sample is added to a reaction vessel with a vitamin D binding protein (DBP) releasing agent and paramagnetic particles coated with sheep monoclonal anti-25(OH) vitamin D antibody. 25(OH) vitamin D is released from DBP and binds to the immobilized monoclonal anti-25(OH) vitamin D on the solid phase. Subsequently, a 25(OH) vitamin D analogue-alkaline phosphatase conjugate is added which competes for binding to the immobilized monoclonal anti-25(OH) vitamin D. After a second incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate Lumi-Phos 530 is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is inversely proportional to the concentration of 25(OH) vitamin D in the sample. The amount of analyte in the sample is determined from a stored, multi-point calibration curve.

    AI/ML Overview

    The document describes the Access 25(OH) Vitamin D Total assay for use on the Access 2 Immunoassay System. This device is intended for the quantitative determination of total 25-hydroxyvitamin D levels in human serum and plasma to aid in the assessment of vitamin D sufficiency.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state "acceptance criteria" as a set of pass/fail thresholds for each metric. Instead, it presents performance characteristics from various studies, comparing them to the predicate device and in some cases, a reference method. I will infer "acceptance criteria" from typical requirements for such assays and the context of the reported results. The reported device performance is directly extracted from the "Summary of Studies" section.

    Performance CharacteristicInferred Acceptance Criteria (Typical for IVD)Reported Device Performance (Access 25(OH) Vitamin D Total)
    Method Comparison (vs. Predicate)High correlation (e.g., r > 0.90)Correlation coefficient: 0.89 (against 279 samples, 7.48-118.83 ng/mL)
    Method Comparison (vs. Reference Method)High correlation and close to 1.0 slopeSlope: 1.01, Correlation coefficient: 0.95 (against 110 samples, 8.0-98.6 ng/mL)
    Imprecision (Total CV > 15.0 ng/mL)≤ 10.0%≤ 10.0% (across studies: 5.5% to 7.5%)
    Imprecision (Total SD ≤ 15.0 ng/mL)≤ 1.5 ng/mL≤ 1.5 ng/mL (across studies: 1.0 to 1.1 ng/mL)
    Linearity (Measuring Range)Linear response across the claimed rangeLinear across 7.0 to 120 ng/mL
    Limit of Blank (LoB)Expected to be very low, below clinical significance1.50 ng/mL (Highest observed with no analyte: 0.55 ng/mL)
    Limit of Detection (LoD)Expected to be low to detect low concentrations2.00 ng/mL (Lowest detected with 95% probability: 1.0 ng/mL)
    Limit of Quantitation (LoQ)Low enough for clinical utility (e.g., %CV
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    Access 25(OH) Vitamin D Total for Use on the UniCel Dxl Immuno. Syst.
    Access 25(OH) Vitamin D Total Calibrators for Use on the UniCel Dxl Immuno.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Access 25(OH) Vitamin D Total assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of total 25-hydroxyvitamin D [25(OH) vitamin D] levels in human serum and plasma using the UniCel DxI Immunoassay Systems. Results are to be used as an aid in the assessment of vitamin D sufficiency.

    The Access 25(OH) Vitamin D Total Calibrators are intended to calibrate the Access 25(OH) Vitamin D Total assay for the quantitative determination of total 25-hydroxyvitamin D [25(OH) vitamin D] levels in human serum and plasma using the UniCel DxI Immunoassay Systems.

    Device Description

    The Access 25(OH) Vitamin D Total for use on the UniCel DxI Immunoassay System, The Access 25(OH) Vitamin D Total Calibrators for use on the UniCel DxI Immunoassay System, and the UniCel DxI Immunoassay analyzer comprise the Access Immunoassay System for the quantitative determination of total 25-hydroxyvitamin D [25(OH) vitamin D] levels in human serum and plasma.

    The Access 25(OH) Vitamin D Total assay is a two-step sequential competitive binding immunoenzymatic assay. In the initial incubation, sample is added to a reaction vessel with a vitamin D binding protein (DBP) releasing agent and paramagnetic particles coated with sheep monoclonal anti-25(OH) vitamin D antibody. 25(OH) vitamin D is released from DBP and binds to the immobilized monoclonal anti-25(OH) vitamin D on the solid phase. Subsequently, a 25(OH) vitamin D analogue-alkaline phosphatase conjugate is added which competes for binding to the immobilized monoclonal anti-25(OH) vitamin D. After a second incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate Lumi-Phos 530 is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is inversely proportional to the concentration of 25(OH) vitamin D in the sample. The amount of analyte in the sample is determined from a stored, multi-point calibration curve.

    AI/ML Overview

    The provided text describes the Beckman Coulter Access 25(OH) Vitamin D Total assay, an in vitro diagnostic immunoassay system cleared by the FDA through the 510(k) pathway. This pathway establishes substantial equivalence to a predicate device, rather than proving a new device meets specific acceptance criteria based on established ground truth in the same way a de novo or PMA would. Therefore, the information provided focuses on comparative performance against a predicate and a reference method, rather than pre-defined acceptance criteria against a clinical ground truth.

    Here's an interpretation of the available information regarding acceptance criteria and supporting studies, framed as closely as possible to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a 510(k) submission, the "acceptance criteria" are implied by the need to demonstrate substantial equivalence to the predicate device and acceptable analytical performance. The document doesn't explicitly state "acceptance criteria" numerical targets in the same way a clinical trial might, but rather describes the results of analytical studies showing performance characteristics.

    Performance MetricImplied "Acceptance Criteria" (Demonstrate equivalence / acceptable analytical performance)Reported Device Performance (Access 25(OH) Vitamin D Total)
    Method Comparison (vs. Predicate)Strong correlation with predicate deviceCorrelation coefficient of 0.88 (Passing-Bablok analysis)
    Method Comparison (vs. Reference Method)Strong correlation with Reference Measurement Procedure (RMP)Slope of 0.99 and correlation coefficient of 0.94 (vs. NIST-Ghent ID-LC-MS/MS RMP)
    Imprecision (Total CV)≤ 10.0% for concentrations > 15.0 ng/mLRanged from 4.2% to 9.3%
    Imprecision (Total SD)≤ 1.5 ng/mL for concentrations ≤ 15.0 ng/mLRanged from 0.7 to 1.3 ng/mL
    LinearityDemonstrate linearity across the measuring rangeDemonstrated linearity from 7.0 ng/mL to 120 ng/mL
    Limit of Blank (LoB)Established LoB for assay1.50 ng/mL (Highest observed with no analyte: 0.98 ng/mL)
    Limit of Detection (LoD)Established LoD for assay2.00 ng/mL (Lowest detectable with 95% probability: 1.47 ng/mL)
    Limit of Quantitation (LoQ)Established LoQ for assay (e.g., %CV of 20%)7.0 ng/mL (Lowest quantifiable with 20% CV: 4.4 ng/mL)
    Analytical Specificity (Interference)No significant interference from normal blood constituents/common medicationsNo interference from normal human blood constituents and common medications; Hemoglobin up to 0.5 g/L without interference
    Cross ReactivityLimited or acceptable cross-reactivityParicalcitol caused falsely elevated results; other cross-reactants did not cause interference

    2. Sample Size and Data Provenance

    • Test Set for Method Comparison (vs. Predicate): 278 samples
    • Test Set for Method Comparison (vs. Reference Method): 110 samples
    • Test Set for Imprecision: Specific sample numbers not explicitly stated, but "representative studies" were performed with multiple concentrations.
    • Test Set for LoB, LoD, LoQ, Linearity, Analytical Specificity, Cross Reactivity: Not explicitly stated as "test sets" with distinct sample counts, but rather as results from specific analytical experiments.
    • Population Observation Group (for typical concentration range): 367 apparently healthy male and female adults.
    • Data Provenance: The document does not explicitly state the country of origin for the samples. The studies are described as analytical validation studies performed by the manufacturer (Beckman Coulter). The reference method (NIST-Ghent ID-LC-MS/MS) suggests an international reference, but the origin of the samples used in comparison to it is not specified. The context of a US FDA submission suggests the intent for US regulatory compliance. The studies are prospective in nature, as they involve testing the device's performance characteristics.

    3. Number of Experts and their Qualifications for Ground Truth

    • For Method Comparison (vs. Predicate): The "ground truth" is effectively the results obtained from the predicate device (DiaSorin LIAISON 25 OH Vitamin D Total Assay). This predicate device's performance would have been established previously, but no expert review of its results is mentioned for this study.
    • For Method Comparison (vs. Reference Method): The "ground truth" is the NIST-Ghent ID-LC-MS/MS Reference Measurement Procedure (RMP). This is a highly accurate and standardized analytical method, a "gold standard" in laboratory measurement, and does not involve human expert interpretation in the diagnostic sense (like a radiologist reading an image). It's a laboratory-based, instrumental reference method.
    • For other analytical studies (Imprecision, Linearity, LoB, LoD, LoQ, Specificity, Cross-reactivity): The "ground truth" is defined by the experimental setup and the known concentrations or characteristics of the samples used (e.g., control samples with known concentrations, samples spiked with interferents). No human experts are used to establish ground truth in this context.

    4. Adjudication Method for the Test Set

    Not applicable. Diagnostic immunoassay studies for 510(k) clearance typically do not involve human adjudication in the way clinical studies for imaging or histology might. The "ground truth" is based on the performance of a reference method or predetermined analytical characteristics.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This type of study (MRMC, human-in-the-loop performance) is not applicable to an automated immunoassay device. The device provides a quantitative numerical result, not an interpretation that radiologists or other human readers would perform.

    6. Standalone (Algorithm Only) Performance

    Yes, the studies described are standalone performance evaluations of the Access 25(OH) Vitamin D Total assay. The device itself performs the measurement and produces the result without human intervention for interpretation beyond operating the instrument and collecting the sample.

    7. Type of Ground Truth Used

    • Predicate Device Performance: For the primary substantial equivalence comparison.
    • Reference Measurement Procedure (RMP): Specifically, the NIST-Ghent ID-LC-MS/MS for the 25(OH) Vitamin D ID LC-MS/MS (a highly accurate and standardized laboratory method). This is considered a highly reliable analytical "ground truth".
    • Known Sample Characteristics: For analytical characteristics like linearity, limits of detection/quantitation, specificity, and cross-reactivity, ground truth is established by using control materials, spiked samples, or samples with otherwise analytically verified characteristics.

    8. Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning or AI. This immunoassay system is a hardware-based diagnostic device with chemical reagents, and its "training" refers to its analytical calibration and development, not algorithm training.
    The closest analogy would be the samples used to develop and refine the assay's performance characteristics and define its calibration curve. However, specific numbers for this purpose are not provided. The development process would involve numerous samples, but these are distinct from a "training set" as understood in a machine learning context.

    9. How the Ground Truth for the Training Set was Established

    As discussed above, an explicit "training set" with ground truth in the AI sense is not applicable. The assay is based on established chemiluminescent immunoassay principles. The "calibration curve" for the device is established using calibrators with known concentrations of 25(OH) Vitamin D. These calibrator concentrations would be precisely determined through highly accurate reference methods, likely similar to the NIST-Ghent ID-LC-MS/MS RMP. The document states "The Access 25(OH) Vitamin D Total Calibrators are intended to calibrate the Access 25(OH) Vitamin D Total assay," and list 6 calibrator levels (0, 6, 17, 37, 87, and 210 ng/mL) containing human serum with 25(OH) vitamin D. The ground truth for these calibrators would be established through a rigorous metrological chain to a primary reference standard.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1