Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K030316
    Date Cleared
    2003-08-07

    (189 days)

    Product Code
    Regulation Number
    870.2360
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    AUDICOR COR SENSOR (AND ACCESSORY ADAPTOR)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Audicor™ Sensor is a single-use, disposable medical device designed for short duration contact (up to 1 hour) for use at the V4 ECG electrode position to acquire both heart sounds and ECG electrical signals when attached to the body surface of a resting, adult (18+ years of age) patient, and to transmit these signals to compatible electrocardiographs and physiological recorders/monitors. Such equipment is commonly located in hospitals, doctor's offices and emergency vehicles.

    Device Description

    The sensor is a single use, non-sterile, disposable medical device. When used with the Audicor Adaptor the device is designed to perform the functions of both a standard resting ECG Electrode by acquiring ECG signals and an Electronic Stethoscope by acquiring heart sounds from the surface of a patients skin. The device is intended for use on the surface of an adult patient's skin in the V4 electrode position on the chest.

    AI/ML Overview

    The provided document describes the Inovise Medical, Inc. Audicor™ Sensor, an ECG Electrode & Heart Sound Transducer. The study conducted for this device is focused on showing its substantial equivalence to predicate devices and its compliance with specified performance standards and biocompatibility requirements.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria for the Audicor™ Sensor seem to be its compliance with specific performance standards and biocompatibility requirements, as well as its functional similarity to predicate devices. The reported device performance is its successful compliance with these.

    Acceptance Criteria & Performance StandardReported Device Performance
    AAMI EC12:2000 (Disposable ECG Electrodes)The sensor was tested to the applicable requirements... and shown to comply. Laboratory and bench testing indicates compliance to the standard.
    AAMI EC53:1995 (ECG Cables and Lead Wires)The sensor was tested to the applicable requirements... and shown to comply. Laboratory and bench testing indicates compliance to the standard.
    Biocompatibility (Hydrogel - primary patient contact material)Evaluated for cytotoxicity, primary skin irritation, and delayed dermal contact sensitization. Test results demonstrate that the material was found acceptable.
    Functional Equivalence (ECG Electrode Functionality)Performs the functions of a standard resting ECG Electrode. Compared to Kendall Disposable ECG Electrode (K953649) and found to have "ECG Electrode Functionality: Yes".
    Functional Equivalence (Heart Sound Acquisition)Performs the functions of an Electronic Stethoscope by acquiring heart sounds. Compared to E-Scope Electronic Stethoscope (K961301) and found to "Acquires Heart Sounds: Yes" and "Uses Microphone: Yes".
    General Performance and Safety"Based on the results of the engineering/design level and biocompatibility tests, it is concluded that the sensor performs as expected and compares well, in terms of overall performance to the selected predicate devices and raises no new questions with regard to safety and efficacy."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample size for any test set or the data provenance (country of origin, retrospective/prospective). The study appears to be primarily focused on laboratory and bench testing of the device itself and biocompatibility testing of its hydrogel.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    The document does not mention the use of experts to establish ground truth for a test set in the context of diagnostic performance. The evaluation is against compliance with technical standards and biocompatibility, not clinical diagnostic accuracy.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. The study is about compliance with engineering standards and biocompatibility, not an assessment of diagnostic performance that would require adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was mentioned. The device is a sensor, not an AI-powered diagnostic tool, and the study focuses on its physical and electrical performance and biocompatibility.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is a sensor for acquiring physiological signals, not an algorithm. The "standalone" performance refers to the device's ability to meet the specified engineering and biocompatibility standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context refers to the defined parameters within the AAMI EC12:2000 and AAMI EC53:1995 standards for ECG electrodes and cables, respectively, and established methodologies for biocompatibility testing (cytotoxicity, primary skin irritation, and delayed dermal contact sensitization). It's essentially compliance with established industry and regulatory standards.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a training set. The study is an evaluation of hardware and material properties.

    9. How the ground truth for the training set was established

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1