Found 2 results

510(k) Data Aggregation

    K Number
    K191867
    Manufacturer
    Date Cleared
    2020-04-09

    (272 days)

    Product Code
    Regulation Number
    888.3030
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ATLAS Plating System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ATLAS™ Plating System is intended to stabilize and aid in the repair of fractures, fusions, and osteotomies of small bones and bone fragments in adult and pediatric patients.

    Device Description

    The ATLAS™ Plating System is a metallic plate and screw system for treatment and fixation of fractures, fusions and osteotomies of bones in the Foot, Ankle, Hand, and Wrist of children (2-12 years), adolescents (12-21 years), and adults. The screws are available in different diameters from Ø2.7mm to Ø5.5mm, non-locking configurations with lengths ranging from 8mm to 60mm. All screws are manufactured from Titanium alloy (Ti-6Al-4V ELI per ASTM F136-13). The plates are available in different geometries to accommodate different patient and bone anatomies. The plates are manufactured from MoRe® alloy (Molybdenum-47.5Rhenium Alloy per ASTM F3273-17).

    AI/ML Overview

    The provided text is a 510(k) summary for the ATLAS™ Plating System. This document focuses on demonstrating substantial equivalence to predicate devices for regulatory clearance, primarily through mechanical testing of the device itself rather than studies involving human or animal subjects or complex algorithms. Therefore, much of the requested information regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of AI/ML or diagnostic device performance is not present in this type of submission.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not specify quantified acceptance criteria (e.g., specific tensile strength values, fatigue limits with minimum requirement numbers) or explicit reported device performance in a table format. Instead, it states that "Mechanical testing for screws (ASTM F543-17)" and "Mechanical testing for plates (ASTM F382-17)" were performed. The conclusion is that "Performance data demonstrate that the ATLAS™ Plating System is substantially equivalent legally marketed predicate devices." This implies that the device met the performance characteristics demonstrated by the predicate devices through these standard tests, which is the basis for substantial equivalence.

    If this were an AI/ML device, this section would typically include metrics like Sensitivity, Specificity, AUC, F1-score, accuracy, etc., with associated thresholds.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable to this type of device submission. The "test set" in this context refers to the physical devices (screws and plates) subjected to mechanical testing, not a dataset of patient information. No patient data provenance is relevant here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable. The "ground truth" for mechanical testing is established by engineering standards (ASTM F543-17 and ASTM F382-17) and the physical properties of the materials and device design, not by expert human interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable to mechanical testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This is a medical device (plating system) submission, not an AI/ML diagnostic or assistive device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this submission is implicitly the established performance characteristics dictated by the referenced ASTM standards (ASTM F543-17 for screws and ASTM F382-17 for plates), which define the methods and expected results for mechanical properties. The device's performance is compared against the performance of predicate devices tested under similar standards to establish substantial equivalence.

    8. The sample size for the training set

    This is not applicable. There is no "training set" in the context of mechanical device testing for a 510(k) submission of this nature.

    9. How the ground truth for the training set was established

    This is not applicable.


    Summary regarding the provided document:

    The provided 510(k) summary for the ATLAS™ Plating System represents a traditional medical device submission focused on mechanical performance and material compatibility. The "study" proving acceptance criteria is a series of mechanical tests conducted according to recognized ASTM standards (ASTM F543-17 and ASTM F382-17). The "acceptance criteria" are not explicitly quantified in the document but are understood to be the successful demonstration of mechanical properties comparable to predicate devices, thus establishing substantial equivalence.

    This type of submission does not involve clinical studies with patient data, AI/ML algorithms, or human reader performance evaluations, which are typically found in submissions for diagnostic software or AI-enabled devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190415
    Manufacturer
    Date Cleared
    2019-04-19

    (57 days)

    Product Code
    Regulation Number
    888.3030
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ATLAS Plating System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ATLAS™ Plating System is intended to stabilize and aid in the repair of fractures, fusions, and osteotomies of small bone fragments in adult and pediatric patients.

    Device Description

    The ATLAS™ Plating System is a metallic plate and screw system for treatment and fixation of fractures, fusions and osteotomies of bones in the Foot, Ankle, Hand, and Wrist.

    The plates are available in different geometries to accommodate different patient and bone anatomies. The plates are made from Titanium alloy (Ti-15Mo Allov per ASTM F2066-18).

    The screws are available in different diameters from Ø2.7mm to Ø5.5mm, non-locking or locking configurations with lengths ranging from 8mm to 60mm. All screws are manufactured from Titanium alloy (Ti-6Al-4V ELI per ASTM F136-13).

    Implants are single use and the system is provided non-sterile.

    AI/ML Overview

    The provided text describes a medical device, the ATLAS™ Plating System, and its regulatory clearance. However, it does not contain information about acceptance criteria or a study proving that the device meets specific performance criteria in the context of an AI/ML algorithm.

    The document is a 510(k) summary for a metallic bone fixation system, which is a physical implant, not a software or AI/ML device. The "performance data" section (Section IX) refers to mechanical testing of the screws and plates (ASTM F543 and F382) to demonstrate the physical integrity and strength of the hardware. These are standard engineering tests for orthopedic implants.

    Therefore, I cannot fulfill the request for information related to AI/ML algorithm acceptance criteria and study details because the provided document is not about an AI/ML device.

    Here's why the requested information isn't present:

    • No AI/ML Component: The device, "ATLAS™ Plating System," is described as a "metallic plate and screw system." There is no mention of software, artificial intelligence, machine learning, image analysis, or any digital diagnostic component.
    • Mechanical Performance Data: The "performance data" cited are ASTM standards for mechanical testing of screws and plates, which are physical tests, not evaluations of an algorithm's accuracy or efficacy.
    • Regulatory Pathway: A 510(k) for a traditional hardware device like this focuses on substantial equivalence to existing predicate devices based on materials, design, and physical performance. It does not involve the types of clinical performance studies and ground truth establishment required for AI/ML devices.

    If you have a document pertaining to an AI/ML medical device, please provide that, and I would be happy to analyze it against your criteria.