Search Results
Found 1 results
510(k) Data Aggregation
(55 days)
ARx Illiac Spinal Screw System
The ARx Illiac Spinal Screw System, is intended for posterior pedicle screw fixation of the non-cervical posterior spine (T1 to S2/ilium) in skeletally mature patients and for pediatric patients to treat adolescent idiopathic scoliosis . It provides stabilization and immobilization of spinal segments as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the posterior thoracic, lumbar, and sacral spine. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
When used as a posterior spine thoracic/lumbar system, the ARx Illiac Spinal Screw System is indicated for one or more of the following: (1) degenerative disc disease (is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), (2) trauma (i.e. fracture or dislocation), (3) curvatures and spinal deformity (scoliosis, kyphosis, and/or lordosis), (4) spinal tumor, (5) failed previous fusion (pseudarthrosis), (6) spinal stenosis, (7) spondylolisthesis.
In order to achieve additional levels of fixation in skeletally mature patients, the ARx Illiac Spinal Screw System 5.5/6.0 rod system may be connected to the Solstice OccipitoCervicoThoracic Fixation System's 3.5mm rod.
The ARx Illiac Spinal Screw System consists of screws and longitudinal rods intended to provide temporary stabilization and immobilization following surgery to fuse a portion of the thoracic, lumbar, and/or sacral spine. The ARx Illiac Spinal Screw System consists of an assortment of rods and screws. The bone screw, head, and taper lock are assembled together during manufacturing to create the ARX Illiac Spinal Screw System screw assembly component. The ARx Illiac Spinal Screw System implant components are made from titanium alloy (Ti-6Al-4V ELI) as described by ASTM F136 and Cobalt Chrome (Co-28Cr-6Mo) as described by ASTM F1537. Do not use any of the ARx Illiac Spinal Screw System components with the components from any other system or manufacturer.
This document is a 510(k) premarket notification for the "ARx Illiac Spinal Screw System," a medical device used for spinal fixation. It focuses on demonstrating substantial equivalence to previously cleared predicate devices, primarily through mechanical performance testing.
Here's the breakdown of the requested information, which can be extracted directly from the provided text:
1. A table of acceptance criteria and the reported device performance
The document states that the testing was performed "to demonstrate substantially equivalent mechanical performance as compared to the Life Spine Arx (K200070)." This implies that the acceptance criteria for the ARx Illiac Spinal Screw System were to meet or exceed the performance of the predicate device (Life Spine Arx, K200070) for each tested parameter. Specific quantitative acceptance criteria are not provided in this summary, but the type of tests performed are mentioned:
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Meet or exceed predicate device performance for: | ARx Illiac Spinal Screw System's performance was |
- Static Axial Grip (per ASTM F1798) | found to be substantially equivalent to Life Spine Arx |
- Static Torsional Grip (per ASTM F1798) | (K200070) in the following tests: |
- Static Flexion-Extension Moment Testing (per ASTM F1798) | - Static Axial Grip |
- Static Compression Bending (per ASTM F1717) | - Static Torsional Grip |
- Static Torsion (per ASTM F1717) | - Static Flexion-Extension Moment Testing |
- Dynamic Compression Bending Testing (per ASTM F1717) | - Static Compression Bending |
- Static Torsion | |
- Dynamic Compression Bending Testing |
2. Sample sized used for the test set and the data provenance
The document does not specify the exact sample size used for the mechanical testing (e.g., number of screws, rods, or test constructs). It also does not explicitly state the provenance of the data (country of origin, retrospective or prospective). However, as this is a medical device approval in the U.S., it can be inferred that the testing was conducted to U.S. regulatory standards and likely performed by the manufacturer or a contracted lab. The data is generated from mechanical testing, not patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable to this type of device and study. The "ground truth" for mechanical testing is established by standardized test methods (ASTM F1717 & F1798) and objective measurements using calibrated equipment, not by human experts interpreting results.
4. Adjudication method for the test set
This information is not applicable. Adjudication methods are typically associated with human interpretation tasks (e.g., medical image reading) where there might be subjectivity or disagreement. Mechanical testing is objective and quantitative.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC comparative effectiveness study was not done. This study focuses on the mechanical performance of a spinal implant, not on the effectiveness of an AI-assisted diagnostic or interpretative tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
A standalone performance study was not done. This is not an AI algorithm.
7. The type of ground truth used
The ground truth used for this study is mechanical test standards and objective measurements. Specifically, the performance of the ARx Illiac Spinal Screw System was compared directly against the mechanical performance of the predicate device (Life Spine Arx, K200070) as measured by established ASTM standards (ASTM F1717 & F1798).
8. The sample size for the training set
This information is not applicable. There is no "training set" in the context of mechanical performance testing for spinal implants. This is not a machine learning or AI device.
9. How the ground truth for the training set was established
This information is not applicable, as there is no training set for this type of device and study.
Ask a specific question about this device
Page 1 of 1