Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K122275
    Manufacturer
    Date Cleared
    2012-08-29

    (30 days)

    Product Code
    Regulation Number
    870.1280
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ARTISAN EXTEND CONTROL CATHETER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hansen Medical Artisan Extend Control Catheter is intended to facilitate manipulation, positioning and control, for collecting electrophysiological data within the heart atria with electro-anatomic mapping and recording systems, using the following percutaneous mapping catheters: the Polaris-Dx ™ Steerable Diagnostic catheters made by Boston Scientific Corporation and the Livewire™ Electrophysiology catheters made by St. Jude Medical.

    The Artisan Extend Control Catheter is intended to be used with the Hansen Medical Sensei® X Robotic Catheter System.

    Device Description

    The Hansen Medical Artisan Extend Control Catheter is designed to facilitate manipulation, positioning and control of mapping percutaneous catheters within the atria of the heart. The Artisan Extend Control Catheter consists of a Steerable Guide Catheter (SGC) and Sheath. The Control Catheter is designed to be used with the Hansen Medical Sensei® X Robotic Catheter System.

    AI/ML Overview

    The Hansen Medical Artisan Extend Control Catheter is a Class II steerable guide catheter, a modification of an existing device (K073225). The device is intended to facilitate manipulation, positioning and control of mapping percutaneous catheters within the atria of the heart, specifically with the Hansen Medical Sensei® X Robotic Catheter System and certain Boston Scientific and St. Jude Medical catheters.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission states, "All of the pre-determined acceptance criteria were met with passing results." However, it does not explicitly list the quantitative acceptance criteria for each test. Instead, it lists the types of tests performed.

    Test CategoryAcceptance Criteria (Implied)Reported Device Performance
    Functional TestsDevice performs as intended, no leaks, proper flush, adequate tensile strength, compatibility with Sensei X.Met all pre-determined criteria and passed.
    Pressure Leak Test(Presumably no leaks or within acceptable leak rate)Passed
    Vacuum Test(Presumably no failure under vacuum)Passed
    Tensile Strength Test(Presumably maintains structural integrity under specified load)Passed
    Flush Test(Presumably allows proper flushing)Passed
    Compatibility with Sensei X Test(Presumably proper integration and function with the robotic system)Passed
    Biocompatibility TestsNo adverse biological reactions beyond accepted levelsMet all pre-determined criteria and passed.
    ISO MEM Elution Cytotoxicity Test(Presumably non-cytotoxic)Passed
    ISO Kligman Maximization Test(Presumably non-sensitizing)Passed
    ISO Intracutaneous Reactivity Study(Presumably non-irritating)Passed
    ISO Acute Systemic Toxicity Study(Presumably non-toxic systemically)Passed
    ASTM Hemolysis(Presumably acceptable levels of hemolysis)Passed
    ISO Thrombogenicity Study(Presumably low thrombogenic potential)Passed
    ISO Complement Activation Assay(Presumably low complement activation)Passed

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for each of the performance and biocompatibility tests. It only states that design verification testing was performed.

    Regarding data provenance: The tests were conducted by the manufacturer, Hansen Medical, Inc., as part of a 510(k) submission to the FDA. This indicates the data is from an internal testing environment rather than a clinical setting. Given the nature of these engineering and biocompatibility tests, they are inherently prospective for the device being evaluated. There is no information regarding the country of origin of the data beyond the manufacturer's location in Mountain View, CA, USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to the type of testing performed. The "ground truth" for these engineering and biocompatibility tests is established by industry standards (e.g., ISO, ASTM) and internal design specifications, not by expert consensus on clinical findings. For example, a leak test has an objective pass/fail criterion based on measurable parameters, not on an expert's interpretation.

    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods (like 2+1, 3+1) are used in studies involving subjective human interpretation of data, typically in clinical evaluations or image analysis. The tests described are objective, pass/fail engineering and laboratory tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical evaluation was not required for this device." This device is a modification of a previously cleared predicate device, and the changes (simplified proximal flush assembly) were determined not to affect the intended use or fundamental scientific technology, precluding the need for clinical studies.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This concept is not applicable here. This is a medical device (a catheter), not a software algorithm or AI model. The performance is assessed through physical and biological testing, not through standalone algorithm evaluation.

    7. The Type of Ground Truth Used

    The "ground truth" for the tests performed is based on:

    • Industry Standards: Many tests refer to ISO (International Organization for Standardization) and ASTM (American Society for Testing and Materials) standards, which define methodologies and acceptable ranges for performance and biocompatibility.
    • Manufacturer Design Specifications: The "pre-determined acceptance criteria" would have been set by Hansen Medical based on the functional requirements of the device and comparison to the predicate device.
    • Objective Measurement: For tests like Pressure Leak, Vacuum, and Tensile Strength, the ground truth is based on measurable physical properties and compliance with a defined range or threshold.

    8. The Sample Size for the Training Set

    This is not applicable. This device is not an AI/ML algorithm that requires a "training set." It is a physical medical device.

    9. How the Ground Truth for the Training Set was Established

    This is not applicable as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1