Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K132043
    Manufacturer
    Date Cleared
    2013-07-31

    (29 days)

    Product Code
    Regulation Number
    878.5000
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ARTHREX SPEEDCINCH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arthrex SpeedCinch is an implantable suture retention device which facilitates percutaneous or endoscopic soft tissue repairs, including the repair of meniscal tears.

    Device Description

    The Arthrex SpeedCinch is an implantable suture retention device which consists of implants and an implant delivery mechanism. The implants are composed of PEEK Optima® (polyetheretherketone) anchors and FiberWire (Polyethylene/Polyester) suture offered in #2-0 size. The implant delivery mechanism is a handheld manual surgical instrument with trigger for implant delivery.

    AI/ML Overview

    The Arthrex SpeedCinch is intended as a suture retention device to facilitate percutaneous or endoscopic soft tissue repairs, specifically meniscal tears. The device's performance was evaluated through mechanical testing to demonstrate substantial equivalence to the predicate device (Arthrex Meniscal Cinch, K073149).

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal acceptance criteria values that the device needed to meet. Instead, it describes a comparative study showing the device's performance relative to a predicate device. The primary performance metrics were tensile strength (pull-out strength) and cyclic displacement.

    Acceptance Criteria (Implied)Reported Device Performance
    Tensile strength (pull-out strength) substantially equivalent to or better than predicate deviceTensile strength (pull-out strength) substantially equivalent to or better than the predicate device.
    Cyclic displacement substantially equivalent to or better than predicate deviceCyclic displacement substantially equivalent to or better than the predicate device.
    Implant materials and general function comparable to predicate deviceImplant materials and general function the same as the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "Mechanical testing data" but does not specify the sample size used for the mechanical testing of either the Arthrex SpeedCinch or the predicate device.

    The data provenance is not explicitly stated regarding country of origin or whether it was retrospective or prospective. Given it's mechanical testing of a physical device, it's typically prospective testing conducted in a laboratory setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This question is not applicable as the study involved mechanical testing of a medical device, not a diagnostic or AI-driven system requiring expert ground truth for interpretation. Performance was measured objectively through physical tests.

    4. Adjudication Method for the Test Set

    This question is not applicable for mechanical testing.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The study was mechanical testing comparing device performance to a predicate device. This is not an AI-assisted diagnostic or interpretation study.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, this is not applicable. The device is a physical implantable suture retention device, not an algorithm.

    7. The Type of Ground Truth Used

    For the mechanical testing, the "ground truth" (or the reference for comparison) was the objective mechanical measurements of tensile strength and cyclic displacement, compared against the same measurements from the predicate device. There was no expert consensus, pathology, or outcomes data used in this specific submission for substantial equivalence.

    8. The Sample Size for the Training Set

    This question is not applicable. There is no software algorithm or AI model mentioned that would require a training set. The device is a physical medical device.

    9. How the Ground Truth for the Training Set was Established

    This question is not applicable as there is no training set mentioned or implied for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1