AROX 45-JBP; AROX 53-JBP; MEROX 45-JBP; MEROX 53-JBP
Arox and Merox bipolar, passive-fixation, endocardial pacing leads are intended to provide permanent pacing and sensing in the atrium and or ventricle when used with a compatible pulse generator.
The BIOTRONIK Arox and Merox leads are bipolar, passive-fixation, endocardial pacing leads available in straight and "J"-shaped configurations, for placement in the ventricle or atrium. The Arox and Merox xx-JBP models have a permanent bend proximal to both lead electrodes, resulting in the distal portion of the lead body having what is commonly referred to as a "U" or "J" shape. This lead shape facilitates placement in the right atrial appendage. The device modification only applies to the atrial Arox xx-JBP bipolar, passive-fixation, endocardial, pacing leads. The modification consists of an exact specification for the J-shape of the leads.
The presented document is a 510(k) premarket notification for a medical device (Arox and Merox Leads), not a study report or clinical trial. Therefore, it does not contain the detailed information typically found in a study that proves a device meets acceptance criteria. The submission is for a "Special 510(k)" due to a device modification – specifically, the precise specification of the J-shape for the atrial Arox xx-JBP leads.
Based on the provided text, here's an analysis of what information is available and what is missing:
1. Table of Acceptance Criteria and Reported Device Performance
This information is not present in the document. A 510(k) summary typically focuses on substantial equivalence to a predicate device rather than detailed acceptance criteria and performance data from a specific study for the modification. The modification is described as an "exact specification for the J-shape of the leads," implying a design control change, likely validated through engineering and bench testing, rather than a clinical performance study.
2. Sample Size Used for the Test Set and Data Provenance
This information is not present. No test set or study data is described in this regulatory submission.
3. Number of Experts Used to Establish Ground Truth and Qualifications
This information is not present. No ground truth establishment is described, as it's not a diagnostic AI device or a study requiring expert consensus.
4. Adjudication Method
This information is not present. No adjudication method is described.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
This information is not present. The device is a pacing lead, not an AI-assisted diagnostic tool. An MRMC study would be irrelevant.
6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance
This information is not present. The device is a physical medical device, not an algorithm.
7. Type of Ground Truth Used
This information is not present. No ground truth is mentioned. The modification is a design specification for a physical characteristic.
8. Sample Size for the Training Set
This information is not present. Since this is a physical device modification, not an AI algorithm, there would be no "training set."
9. How the Ground Truth for the Training Set Was Established
This information is not present. As above, there is no training set mentioned.
Summary of Device and Modification relevance to Acceptance Criteria:
The core of this 510(k) submission is a "Special 510(k)" for a modification to existing, previously approved pacemaker leads (Arox and Merox). The modification is:
- Device Modification: "The modification consists of an exact specification for the J-shape of the leads. Previously, BIOTRONIK did not specify the precise J-shape of the leads."
For such a modification, acceptance criteria would typically revolve around:
- Dimensional Accuracy: Does the new manufacturing process consistently produce the J-shape within the specified tolerances? This would be verified through manufacturing quality control and possibly bench testing.
- Mechanical Integrity/Performance: Does the precisely defined J-shape maintain the lead's mechanical properties (e.g., flexibility, durability, resistance to kinking, ability to be positioned as intended)? This would involve bench testing for fatigue, tensile strength, and simulated use.
- Biocompatibility: (Likely unchanged, as materials are not mentioned as being modified).
- Electrical Performance: (Likely unchanged if electrodes and insulation are not modified).
The current document only states that the FDA reviewed the submission and found the device to be "substantially equivalent" to predicate devices, subject to general controls. This implies that the modification was deemed not to raise new questions of safety or effectiveness. The specific evidence (e.g., engineering reports, bench test results) demonstrating that the modified J-shape meets the internal design and performance specifications (the "acceptance criteria" for this specific change) would have been part of the full 510(k) submission to the FDA, but it is not provided in this summary document.
Ask a specific question about this device
(14 days)
AROX 53-BP; AROX 60-BP; AROX 45-JBP; AROX 53-JBP
Arox bipolar, passive-fixation, endocardial pacing leads (Arox xx-BP and Arox xx-JBP) are intended to provide permanent pacing and sensing in the atrium and or ventricle when used with a compatible pulse generator.
The BIOTRONIK Arox lead is a bipolar, passive-fixation, endocardial pacing lead available in straight and " J"-shaped configurations, for placement in the ventricle or atrium. The designation Arox xx-BP refers to Arox straight leads, which are available in lengths (xx) of 53 or 60 cm; Arox xx-JBP refers to Arox "J"-shaped leads, which are available in lengths of 45 and 53 cm.
The Arox xx-JBP model has a permanent bend proximal to both lead electrodes, resulting in the distal portion of the lead body having what is commonly referred to as a "U" or "J" shape. This lead shape facilitates placement in the right atrial appendage.
Arox xx-BP and Arox xx-JBP leads feature four silicone tines for passive-fixation in the heat's trabeculae. The tip electrode is made of a platinum/iridium base material, with a fractal iridium surface and electrically active surface area of 3.5 mm². The ring electrode is made of a platinum/iridium base material, with a fractal iridium surface and electrically active surface area of 22.6 mm². The lead conductor consists of quadrifilar MP35N wire in a coaxial configuration, insulated with silicone rubber tubing. All Arox leads utilize a 3.2 mm IS-1 connector.
The provided text is a 510(k) summary for the BIOTRONIK Arox Passive-Fixation, Bipolar, Endocardial Pacing Lead. This document describes the device, its intended use, and its substantial equivalence to previously cleared predicate devices. It does not contain information about acceptance criteria, study methodologies, sample sizes, ground truth establishment, or expert involvement for a study proving device performance against acceptance criteria.
The 510(k) process is primarily a premarket notification for demonstrating substantial equivalence to a legally marketed predicate device, not necessarily a comprehensive clinical study to prove performance against specific acceptance criteria in the way a clinical trial would. Therefore, I cannot extract the requested information from this document.
The document mainly focuses on:
- Device Description: Physical characteristics, materials, and configurations.
- Device Modification: Differences from a previously marketed device (Merox leads) and similarities to another (Polyrox leads).
- Predicate Devices: Identification of existing devices to which the Arox leads are compared for substantial equivalence.
- Indications for Use: The medical conditions and manner in which the device is intended to be used.
To answer your questions, one would typically need access to the full 510(k) submission, including any testing reports, bench test results, or, if applicable, clinical study summaries, which are not provided in this specific excerpt.