510(k) Data Aggregation
(189 days)
APLIO DIAGNOSTIC ULTRASOUND SYSTEM; APLIO DIAGNOSTIC ULTRASOUND SYSTEM; APLIO DIAGNOSTIC ULTRASOUND SYSTEM
The Aplio TUS-A500/A400/A300 (v2.0) Diagnostic Ultrasound System is indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, pediatric, small organs, trans-vaginal, neonatal cephalic, adult cephalic, cardiac, peripheral vascular, transesophageal, and musculo-skeletal (both conventional and superficial).
The Aplio 500/400/300 system is a mobile ultrasound system. It is a Track 3 device that employs a wide array of probes that include flat linear array, convex linear array, and sector array with a frequency range of approximately 2 MHz to 12 MHz. The Aplio 500/400/300 is designed to support a wide rauge of applications depending on which software is installed. The system can be a dedicated system or a eneral purpose system.
The provided document is a 510(k) Premarket Notification for the Toshiba Aplio™ TUS-A300/400/500 (v2.0) Diagnostic Ultrasound System. This document focuses on demonstrating substantial equivalence to predicate devices based on technical characteristics and intended use, rather than on detailed performance criteria from a clinical study for a specific disease or condition. Therefore, directly addressing some of your questions, particularly those related to clinical performance metrics and ground truth establishment in a study for AI device detection, is not possible with this type of document because it describes a general-purpose diagnostic ultrasound system, not an AI-powered detection/diagnosis device.
However, I can extract information relevant to the device's functional performance and general safety/effectiveness criteria as described in this 510(k) submission:
The Aplio™ TUS-A300/400/500 (v2.0) is a diagnostic ultrasound system. Its acceptance criteria and proof of meeting these criteria are established through adherence to recognized medical device standards and demonstrating substantial equivalence to legally marketed predicate devices.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not provide a table of explicit acceptance criteria with numerical performance targets (e.g., sensitivity, specificity for a particular disease) in the way one might expect for an AI-based detection device. Instead, the "acceptance criteria" for a diagnostic ultrasound system in a 510(k) context are primarily related to:
- Compliance with recognized standards: Ensuring the device meets international and national standards for safety and performance.
- Substantial equivalence: Demonstrating that the new device is as safe and effective as existing legally marketed predicate devices.
- Intended Use: The device performs its intended functions for specified clinical applications.
The 'reported device performance' is implicitly the successful demonstration of compliance with these standards and equivalence in the 510(k) submission.
Acceptance Criterion Type | Specific Criteria Described in Document | Reported Device Performance |
---|---|---|
Safety & Essential Performance Standards | - IEC 60601-1 (medical electrical equipment general requirements) |
- IEC 60601-1-1 (safety requirements medical electrical systems)
- IEC 60601-1-2 (electromagnetic compatibility)
- IEC 60601-1-4 (programmable electrical medical systems)
- IEC 60601-2-37 (specific requirements for ultrasonic medical diagnostic and monitoring equipment)
- IEC 62304 (medical device software life cycle processes)
- AIUM-NEMA UD2 Output Measurement Standard (for Track 3 Ultrasound systems)
- AIUM-NEMA UD3 Output Display Standard (for Track 3 Ultrasound systems)
- Quality System Regulation (QSR) | Device is designed and manufactured in conjunction with these standards. Testing has been conducted as per these standards. (Implicitly, compliance is achieved for clearance). |
| Intended Use Equivalence | Performance for specified clinical applications (fetal, abdominal, intraoperative, pediatric, small organs, neonatal cephalic, adult cephalic, cardiac, transrectal, transvaginal, transesophageal, peripheral vascular, musculo-skeletal) across various transducers and operating modes (B, M, PWD, CWD, Color Doppler, Combined, THI, Dynamic Flow, Power, CHI 2D, 4D, etc.). | Determined "substantially equivalent" to predicate devices for the stated indications for use and associated transducers. (K103645, K090158, K092271) |
| Predicate Device Comparison | Substantial equivalence to: - K103645 Toshiba Diagnostic System Aplio XG SSA-790A v5.2
- K090158 Toshiba Diagnostic Ultrasound system Aplio Artida SSH-880A v2.0
- K092271 General Electric LOGIC E.9 Ultrasound System | FDA has determined the device is substantially equivalent to legally marketed predicate devices. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not detail specific "test set" sample sizes in terms of patient data or images for a comparative clinical performance study comparing the new device's diagnostic accuracy to a gold standard. For general-purpose ultrasound systems submitted under 510(k), real-world clinical performance data in terms of diagnostic accuracy (like sensitivity/specificity) is often not required if technical and safety specifications are met and substantial equivalence to existing devices can be proven. The "testing" mentioned refers to engineering and bench testing to demonstrate compliance with standards and a safety profile similar to predicate devices.
Therefore, this document does not provide information on:
- Sample size used for a test set (in terms of patient data/images).
- Data provenance (country of origin, retrospective/prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable and not provided in this 510(k) summary. Since it's a general diagnostic ultrasound system and not an AI-powered analysis tool, there would not be a "ground truth" to establish for its diagnostic output in the way an AI diagnostic device needs. Its output is raw image data and measurements which are then interpreted by a human expert.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable and not provided for the same reasons as in point 3.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-assisted device; it is a diagnostic ultrasound system that produces images and measurements. Therefore, no MRMC study comparing human readers with and without AI assistance would have been performed or reported in this type of 510(k) submission.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This device is not an algorithm for standalone performance; it is a medical imaging hardware system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not applicable. As explained above, for a general-purpose ultrasound system, the "ground truth" concept for diagnostic accuracy is not typically applied to the device itself but rather to the interpretation of its images by a clinician.
8. The sample size for the training set
Not applicable. This device is a traditional diagnostic ultrasound system, not an AI/machine learning model that requires a training set of data.
9. How the ground truth for the training set was established
Not applicable for the same reasons as in point 8.
Ask a specific question about this device
(13 days)
SSA-770A - VERSION 4, APLIO DIAGNOSTIC ULTRASOUND SYSTEM
The APLIO is intended to be used for the following type of studies; fetal, abdominal, intraoperative, pediatric, small organs, neonatal cephalic, adult cephalic, cardiac, transrectal, transvaginal, transesophageal, peripheral vascular, musculo-skeletal (both conventional and superficial) and laparoscopic.
The APLIO Ultrasound System is a mobile system. This system is a Track 3 device that employs a wide array of probes that include flat linear array, convex linear array, and sector array with a frequency range of approximately 2 MHz to 12 MHz. This submission is to clear cardiac packages already in place on predicate devices.
The provided text is a 510(k) Summary of Safety and Effectiveness for the Toshiba SSA-770A, APLIO Version 4 Diagnostic Ultrasound System. It details the device's technical specifications, intended uses, and its substantial equivalence to previously cleared predicate devices.
However, the document does not contain information regarding acceptance criteria or a specific study that proves the device meets such criteria. The document focuses on regulatory clearance by demonstrating substantial equivalence to existing devices, and lists various transducers with their cleared modes of operation and clinical applications. There are no performance metrics, accuracy statistics, or study designs mentioned that would typically be associated with proving a device meets specific acceptance criteria based on quantifiable performance.
Therefore, I cannot provide a table of acceptance criteria and reported device performance or details about a study evaluating its performance against such criteria. The document primarily functions as a regulatory submission detailing the device's characteristics and intended use, rather than a performance evaluation report.
Ask a specific question about this device
(14 days)
SSA-700A, APLIO DIAGNOSTIC ULTRASOUND SYSTEM
The APLIO is intended to be used for the following type of studies; fetal, abdominal, intraoperative, pediatric, small organs, neonatal cephalic, adult cephalic, cardiac, transrectal, transvaginal, transesophageal, peripheral vascular, musculo-skeletal (both conventional and superficial) and laparoscopic.
The APLIO Ultrasound System is a mobile system. This system is a Track 3 device that employs a wide array of probes that include flat linear array, convex linear array, and sector array with a frequency range of approximately 2 MHz to 12 MHz.
This 510(k) submission for the Toshiba SSA-700A, APLIO Diagnostic Ultrasound System and its associated transducers focuses on substantial equivalence to predicate devices rather than establishing new performance criteria through clinical studies. Therefore, much of the requested information about acceptance criteria and clinical study details (like sample sizes, ground truth establishment, expert qualifications, and MRMC studies) is not present.
However, based on the provided document, here's what can be inferred:
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not specify quantitative acceptance criteria for device performance (e.g., sensitivity, specificity, accuracy) from a clinical study. Instead, the acceptance criterion for this 510(k) submission is Substantial Equivalence to legally marketed predicate devices. The "reported device performance" is implicitly that the device performs equivalently to these predicates for the indicated uses.
Acceptance Criteria (Implicit) | Reported Device Performance (Implicit) |
---|---|
Substantially equivalent in safety and effectiveness to predicate ultrasound systems (UIDM-400A (K992886) and SSA-770A APLIO (K013633)). | The APLIO Ultrasound System is considered substantially equivalent to the UIDM-400A (K992886) and SSA-770A APLIO (K013633) for its specified clinical applications and modes of operation. Engineering assessments identified no unmitigated issues of risk or safety compared to the Toshiba SSA-770A Aplio. |
Compliance with relevant safety and performance standards (e.g., Quality System Regulation, IEC 60601, AIUM-NEMA UD2, AIUM-NEMA UD3). | The device is designed and manufactured in conjunction with the Quality System Regulation, IEC 60601 (applicable portions), IEC60601-2-37 (applicable portions), the AIUM-NEMA UD2 Output Measurement Standard as applied to Track 3 Ultrasound systems, and the AIUM-NEMA UD3 Output Display Standard. |
Ability to perform specified clinical applications with various modes (B, M, PWD, CWD, Color Doppler, Amplitude Doppler, Color Velocity Imaging, Combined, Harmonic Imaging). | The device, with its various transducers, supports a wide range of clinical applications including fetal, abdominal, intraoperative, pediatric, small organs, neonatal cephalic, adult cephalic, cardiac, transrectal, transvaginal, transesophageal, peripheral vascular, musculo-skeletal (conventional and superficial), and laparoscopic, across various imaging modes (as detailed in the "Indications for Use Form" for each transducer). |
2. Sample Size Used for the Test Set and Data Provenance:
The document does not describe a clinical "test set" in the context of a performance study with patient data. The basis for substantial equivalence is primarily through engineering assessments and comparison to existing, legally marketed devices. Therefore, information regarding sample size, country of origin, or retrospective/prospective nature of a clinical test set is not provided.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
Since no clinical test set requiring ground truth establishment is described, this information is not available in the document.
4. Adjudication Method for the Test Set:
Not applicable, as no clinical test set requiring adjudication is described.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No MRMC study is mentioned in the document. The submission relies on substantial equivalence to predicate devices rather than demonstrating comparative effectiveness with human readers.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:
This device is a diagnostic ultrasound system intended for use by a human operator, not a standalone AI algorithm. Therefore, no standalone algorithm performance study is described.
7. Type of Ground Truth Used:
Not applicable as there is no mention of a primary clinical study requiring ground truth. The "ground truth" for this submission is implicitly the established safety and effectiveness of the identified predicate devices, as demonstrated through their prior clearance and market history.
8. Sample Size for the Training Set:
The document does not describe the use of a "training set" in the context of an algorithm or AI development. This is a traditional medical device submission for an ultrasound system, not an AI/ML product.
9. How the Ground Truth for the Training Set Was Established:
Not applicable, as no training set is described.
Ask a specific question about this device