Search Results
Found 1 results
510(k) Data Aggregation
(57 days)
AP50/30 Insufflator with Insuflow Port
The AP50/30 Insufflator is a CO2 insufflator intended for use during diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with gas. The Standard/High Flow, Pediatric operating modes of the device are indicated to fill and distend a peritoneal cavity with gas during a laparoscopic procedure. The Pediatric operating mode is specifically indicated for pediatric laparoscopic procedures. The Vessel Harvesting mode is indicated for use during endoscopic vessel harvesting procedures to create a cavity along the saphenous vein or radial artery. The TAMIS operating mode is indicated to fill and distend the rectum and colon using CO2 gas during trans and minimal invasive surgery.
The Insuflow® Port (5 mm, 8 mm, 10mm and 12mm) devices have applications in thoracic, abdominal and gynecologic minimally invasive endoscopic surgical procedures to establish a path of entry for endoscopic instruments and to heat, humidify, filter and introduce a CO2 gas stream for insufflation of the surgical cavity.
The Lexion AP50/30 Insufflator is a microprocessor-controlled CO2 (carbon dioxide) insufflator with multiple operating modes. The insufflator instrumentation is intended for hospital use for Standard/High Flow, Pediatric, and Bariatric laparoscopic procedures used in conjunction with a laparoscope to fill and distend a peritoneal cavity with gas; for trans anal minimal invasive surgery (TAMIS) to fill and distend the rectum and colon using CO2 gas; and for Vessel Harvesting procedures used to create a cavity along the saphenous vein and/or the radial artery during an endoscopic vessel harvesting procedure.
The device incorporates the following major components and features: a metal housing, a world power supply, pressure reducers, a venting system, a gas heater control and a touch screen user interface with various settings and display elements. The software operation includes system checks, user interface, setting adjustments, warning/error messages and service info. The device is equipped with a continuous pressure measurement mode that controls the conformity of the actual pressure in the peritoneal or extraperitoneal cavity with the pre-set nominal pressure. The AP50/30 Insufflator is designed with several alarms/warnings to inform the operator in case of an overpressure or other malfunctions. The device is to be used with specially designed singleuse tubing sets, the Insuflow® Port, in order to utilize the full capabilities of continuous pressure measurement and gas heating and humidification. (The insufflator can also accept other tubing sets for gas deliver only.)
The Insuflow® Port (5, 8, 10, 12 mm) devices are gas conditioning/access port devices that attach to the outlet of the AP50/30 Insufflator and are designed to warm and humidify the CO2 gas stream prior to insufflation via an integral path of entry device during minimally invasive surgery. The Insuflow® Port consists of an ethylene oxide sterilized, disposable single use tubing set and a path of entry access port device which contains the pressure sensors, a filter, and gas heater/humidifier. The access port device materials are intended for patient contact of less than 24 hours. The Insuflow® Port is connected to the AP50/30 Insufflator via a plug connector cable, which controls the pressure sensing, gas heating and safety circuits for the system. Regulated CO2 gas from the AP50/30 Insufflator flows into the Insuflow® Port, through the in-line filter, continues along the tubing to enter the path of entry access device that contains the heating element and humidification media, and through the path of entry access device lumen for delivery into the patient's surgical cavity.
The provided text is a 510(k) Premarket Notification for a medical device, the AP50/30 Insufflator with Insuflow® Port. It focuses on demonstrating substantial equivalence to predicate devices, particularly for a newly added operating mode (TAMIS).
Important Note: This document describes the acceptance criteria and study for a hardware medical device (insufflator), not an AI/ML-based device. Therefore, many of the requested fields related to AI/ML (e.g., ground truth establishment for training set, MRMC study, sample size for training set, number of experts for ground truth) are not applicable to this type of medical device submission. The study described is an engineering performance verification and validation.
Here's a breakdown based on the provided text, addressing the applicable criteria for this device:
Acceptance Criteria and Device Performance for AP50/30 Insufflator with Insuflow® Port (TAMIS Mode)
This submission focuses on demonstrating substantial equivalence of the modified AP50/30 Insufflator (with the addition of a TAMIS operating mode) to its predicate devices. The acceptance criteria are implicit in matching the performance specifications of the predicate devices for existing modes and ensuring the new TAMIS mode meets appropriate performance targets comparable to a predicate device.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally established by matching or demonstrating comparable performance to the predicate devices, particularly the PNEUMOCLEAR Insufflator (K170784) for the TAMIS mode. The tables provided in the document serve as the performance data compare to historical predicate performance.
Key Performance Specs | AP50/30 Insufflator with TAMIS (Reported Performance) | AP50/30 Insufflator K170799 (Primary Predicate) | PNEUMOCLEAR K170784 (Secondary Predicate - for TAMIS) | Acceptance Criteria (Implicit from Predicate) |
---|---|---|---|---|
Insufflator Indications for Use | Includes TAMIS mode to fill and distend rectum and colon using CO2 gas during trans anal minimal invasive surgery, in addition to existing modes. | Standard/High Flow, Pediatric, Bariatric, Vessel Harvesting. | Standard, High Flow/Bariatric, Pediatric, Advanced Flow, Vessel Harvest, and TAMIS mode for rectum and colon distention. | Device's TAMIS indication should match the secondary predicate. Other indications should match the primary predicate. |
Operating Modes (New) | Standard/High Flow, Pediatric, Bariatric, Vessel Harvesting, and TAMIS | Standard/High Flow, Pediatric, Bariatric and Vessel Harvesting | Standard, High Flow/Bariatric, Pediatric, Advanced Flow, Vessel Harvesting, and TAMIS | New TAMIS mode should be consistent with the secondary predicate. |
Max Gas Flow (TAMIS) | 40 lpm (TAMIS) | N/A (No TAMIS mode) | 40 lpm (TAMIS) | Max gas flow for TAMIS should be 40 lpm, matching secondary predicate. |
Pressure Range (TAMIS) | 1-20 mmHg (TAMIS) | N/A (No TAMIS mode) | 1-20 mmHg (TAMIS) | Pressure range for TAMIS should be 1-20 mmHg, matching secondary predicate. |
Intra-abdominal Max Pressure (TAMIS) | 20 mmHg (TAMIS) | N/A (No TAMIS mode) | 20 mmHg (TAMIS) | Max pressure for TAMIS should be 20 mmHg, matching secondary predicate. |
Maximum Supply Pressure (TAMIS) | 65 mmHg in TAMIS mode | N/A (No TAMIS mode) | 65 mmHg in TAMIS mode | Max supply pressure for TAMIS should be 65 mmHg, matching secondary predicate. |
Other Existing Modes (Flow, Pressure, Max Pressure, Supply Pressure, Adjustable Values, Pneumo Pressure Control, User Interface, Power, Error Warnings, Gas Conditioning, Tubing Sets) | Match previous AP50/30 performance. | Specified values for each mode. | Specified values for each mode. | Must remain consistent with the primary predicate. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document refers to "comparative bench testing" and "design verification testing." It does not specify a numerical sample size in terms of "units tested" or "number of unique test scenarios." This type of engineering testing typically involves a set number of test runs or units to demonstrate specifications are met across operational ranges and conditions.
- Data Provenance: The testing was "performed to demonstrate that the performance of the proposed AP50/30 Insufflator is substantially equivalent to that of the predicate devices." This implies that the data is prospective data generated from testing of the new device intended for submission. The location of testing is not specified, but it would have been conducted by the manufacturer (Lexion Medical LLC) or a contracted test facility.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- N/A. For a hardware medical device like an insufflator, "ground truth" as it relates to expert consensus on images or clinical outcomes is not applicable. The device's performance is objectively measured against engineering specifications (e.g., flow rate, pressure accuracy, alarm function). Verification and validation testing is conducted by engineers and technicians against established specifications.
4. Adjudication Method for the Test Set
- N/A. Adjudication methods (like 2+1, 3+1) are relevant for subjective assessments, typically in image interpretation or clinical trial endpoints that require expert consensus. For an insufflator, performance is measured against objective, quantifiable engineering specifications.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- N/A. MRMC studies are specific to evaluating the diagnostic performance of imaging devices or algorithms, usually involving human readers interpreting images. This is a hardware device for surgical procedures (insufflation), not an imaging or diagnostic AI/ML device.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- N/A. This is not an AI/ML algorithm. The device has software that controls its functions, which was validated ("The software used in the AP50/30 Insufflator was determined to have a major level of concern, and was developed and successfully validated in accordance with the FDA guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," issued May 11, 2005."). This software validation assesses the correct functioning of the control algorithms, which operates autonomously to maintain pressure and flow.
7. The Type of Ground Truth Used
- Engineering Specifications / Predicate Device Performance: The "ground truth" for this device's performance is its ability to meet established engineering specifications (e.g., precise control of gas pressure and flow, accurate alarm triggers) and to perform comparably to its legally marketed predicate devices, particularly for the added TAMIS mode. The performance is quantified through direct physical measurements during bench testing.
8. The Sample Size for the Training Set
- N/A. This is not an AI/ML device; there is no "training set." The software is designed and programmed based on engineering principles and requirements, not trained on data.
9. How the Ground Truth for the Training Set was Established
- N/A. As there is no training set for an AI/ML model, this question is not applicable. The software ground truth is derived from design requirements and engineering specifications documented during the device development process.
In summary, the provided document details the regulatory submission for a hardware medical device with specific performance characteristics. The "acceptance criteria" revolve around demonstrating that the modified device (with a new operating mode) performs comparably to its predicates and meets all relevant safety and performance standards. The "study" proving this involves comprehensive bench testing and software validation against objective engineering specifications. The concepts of AI/ML-specific evaluations (like training sets, data provenance for AI models, MRMC studies, or multi-expert adjudication for diagnostic ground truth) do not apply to this type of device submission.
Ask a specific question about this device
Page 1 of 1