Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K131299
    Date Cleared
    2013-05-22

    (15 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Anythink is a medical image processing system, which offers comprehensive solutions to viewing, manipulation, communication and storage of multi-modality DICOM images and data on exchange media.

    The Anythink® is a universal imaging platform, and supports different modalities, but it is not intended for the displaying of digital mammography images for diagnosis in the U.S.

    Device Description

    The Anythink® is based on Windows XP, providing a set of software solutions with flexible configurations in accordance with different medical imaging missions and demands. The system accepts multi-modality DICOM images and allows for view, post-processing, and communication. This product is not intended for use with or for the primary diagnostic interpretation of mammography images in the U.S.

    Due to specific customer requirements and the clinical focus, the Anythink® consists of two parts: Basic Functions and Optional Functions.

    Basic functions are mandatory, allowing view, easy manipulation, storage and communication of DICOM formatted images, except in the case of mammography images. Management of patient information is also included.

    Optional Functions are a set of professional image processing functions, designed for specific modalities. Depending on the precise requirement, customers can select the appropriate function(s) from the optional functions, under the precondition of installing Basic Functions.

    The clinician retains the ultimate responsibility for making the pertinent diagnosis based on their standard practices. The Anythink® is intended to assist the physician in diagnosis or treatment planning.

    AI/ML Overview

    The provided FDA 510(k) summary for the Anythink® / Anythink PACS Workstation (K131299) establishes substantial equivalence to a predicate device (Siemens Syngo Multimodality Workplace, K072728). However, this summary does not include a study that explicitly proves the device meets specific acceptance criteria based on quantitative performance metrics, such as sensitivity, specificity, or reader performance.

    Instead, the submission relies on a comparison of technological characteristics and intended use to demonstrate substantial equivalence. For medical image processing systems like PACS workstations, the FDA often accepts this approach without requiring a full-scale clinical performance study if the new device shares fundamental technological characteristics and intended use with a legally marketed predicate.

    Therefore, many of the requested details about acceptance criteria, study design, sample size, ground truth, and reader studies are not present in this 510(k) summary. This is typical for Class II medical devices considered substantially equivalent, where the focus is on demonstrating that the new device does not raise new questions of safety and effectiveness compared to an existing one.

    However, I can extract information related to the device's technical specifications and general safety and effectiveness claims.

    Here's a breakdown of what can be inferred or explicitly stated based on the provided text, and what information is missing:

    1. Table of Acceptance Criteria and Reported Device Performance

    This document does not present a table of quantitative acceptance criteria (e.g., minimum accuracy, sensitivity, or specificity) or numerical performance metrics for the Anythink® PACS Workstation. The "acceptance criteria" for a 510(k) submission are fundamentally about demonstrating substantial equivalence, meaning the device is as safe and effective as a predicate device. The performance is largely implied to be equivalent to the predicate through the comparative analysis of features.

    The document mainly focuses on comparing the functions of the Anythink® with the Siemens Syngo-Multimodality Workplace:

    FunctionAnythink DescriptionPredicate (Siemens Syngo) Description
    Hardware requirementCPU: Dual-core 2.8G, Hard disk: 250G, Memory: 4 GB, Monitor: 19 inch LCD (1280x1024), Video card: Nvidia serial (1G), DVD-RW, Keyboard/Mouse, 100M/1000M network cardType: FSC Celsius R640, Processor: 2 x 3.0 GHz Intel Xeon Dual core, RAM: 4 GB (upgradable to 12 GB), System disk: 73 GB, Hard disk for image data: 147 GB, DVD writer, DVD reader, Floppy disk, Graphics card: OpenGL, Network: 2 x Gigabit Ethernet LAN, Monitors: Flat-screen color/monochrome
    Operating systemMicrosoft Windows XPWindows XP
    Patient information managementCreating new, editing, deleting patient info; inquires and ranks patient info; disk backup; burns data to DICOM, AVI, MPEG.Offers consistent access to patient and exam data from all applications. DICOM Media Storage for data exchange on CD/DVD.
    Image BrowsingAdjusting windows width/level by hotkeys, presetting tools, or manually; single/multiple window display, switching images within/among series, between patients.Comprehensive functions for 2D processing and image evaluation; Multimodality display with intuitive tools; Display and arrange images to suit diagnostic task.
    Image ManipulationApplies to DICOM compliant images (except digital mammography); adjusts for different human tissue/body positions; positive-negative conversion, local/overall zoom, dying scheme, flip, rotate, edge enhancement/smoothing, measurement, annotation.Review, process, evaluate results, prepare for supporting physician's diagnosis. Send to syngo Filming, store, or send to other hospital locations.
    Image TransmitApplies to DICOM compliant images (except digital mammography); supports image transformation between system and other DICOM devices; input images from disk to system.DICOM functions for receiving/sending digital examinations and local data exchange; connecting to radiological network.
    Image StorageSix image/series storage functions; saving current image to temporary area/clipboard/file/database; saving current sequence as video file or to database. Applicable for DICOM compliant images (except digital mammography).DICOM Storage for data transfer/archiving; DICOM Storage Commitment; DICOM Query & Retrieve; DICOM Print.
    3D reconstruction (3D View)MPR, Max/Min Intensity projection (MIP), Shaded surface Display, Volume Rendering, Virtual Endoscopy for CT/MRI images; 3D display of various tissues, bone, angiosteosis, etc.Processes volume datasets (MIP, MPR, SSD); Displays CT, MR, NM, XA volume with editing (e.g., bone removal); Volume Rendering Technique.
    XA heart and coronary artery analysis (QCA, LVA)QCA: Applicable for XA coronary artery projection, vessel stenosis analysis (Live-wire algorithm). LVA: For XA heart projection, analysis of cardiac ejection functional and wall motion (radial/center line method).syngo Angio (DSA): Shifts DSA image processing. syngo QCA: Quantitative coronary vessel analysis. syngo LVA: Left ventricle analysis (ejection fraction, wall motion). syngo LVA biplane: For simultaneous biplane acquisitions. syngo QVA: Quantitative vessel analysis for abdominal/peripheral vessels.
    CT Coronary Artery AnalysisExtracting/adjusting coronary artery tree in CTA, single coronary artery vessel and analysis in 3D (QCA method).syngo Circulation: Comprehensive cardiac/chest pain evaluation with reporting. syngo Circulation QCA with Plaque Analysis: Fast coronary tree segmentation, stenosis quantification, stent planning, plaque analysis. syngo Circulation LVA: Complete ventricular function evaluation in multiphase cardiac datasets.
    Virtual ColonoscopyAssistant tool for detecting colonic lesions from CT images (extracting colon image, endoluminal display, marking suspected lesions). Applies VR+SSD.syngo Colonography: Locates/evaluates colon polyps using non-invasive, real-time virtual 3D endoluminal viewing. syngo Colonography Polyp Enhanced Viewing (PEV): Automated second reader tool. syngo Colonography Unfolding: Allows unfolding colon for easier polyp visualization.
    3D Angiostenosis analysis (syngo IC3D)Creates 3D models of coronary vessel segments from two projection images for image analysis and stent indications; quantitative analysis of vessel.Creates 3D models of coronary vessel segments for highly accurate quantification of lesions - stent size and length quantification with as few as two projection images.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not specified. The submission does not detail a specific test set of medical cases used for performance evaluation in the traditional sense (e.g., a cohort of patients/studies).
    • Data Provenance: Not specified. Since no specific test set study is presented, the country of origin or whether the data was retrospective or prospective is not mentioned. The device's intended market is the U.S.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Number of Experts: Not specified. As there's no described performance study with ground truth establishment, this information is not provided.
    • Qualifications of Experts: Not specified.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. No study with an adjudication process is described in this summary.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • MRMC Study: No. This type of study (comparing human readers with and without AI assistance) is not mentioned. The device is a PACS workstation, not an AI-assisted diagnostic tool in the sense of providing specific lesion detection or characterization beyond standard viewing and manipulation functions. The comparative effectiveness is implicitly demonstrated by comparing the functionalities and technical characteristics to the predicate, not by a human-in-the-loop study.
    • Effect Size: Not applicable as no MRMC study was performed.

    6. If a Standalone (Algorithm Only) Performance Study Was Done

    • Standalone Study: No. The device is a PACS workstation, which is a tool for viewing and manipulating images, not a standalone diagnostic algorithm that would generate outputs independent of a human operator. Its performance is based on its functionality and ability to process and display images robustly, not on an "algorithm only" diagnostic accuracy.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: Not applicable. Since no performance study measuring diagnostic accuracy against a ground truth is described, this information is not provided. The "ground truth" for a PACS workstation primarily relates to its ability to accurately and reliably display and process DICOM images according to established standards.

    8. The Sample Size for the Training Set

    • Sample Size: Not applicable/Not specified. The device is a software solution with "Basic Functions" and "Optional Functions" designed for image processing. It's not described as a deep learning or machine learning system that would require a large training set in the typical sense for algorithms that learn patterns for diagnostic tasks. Its development likely involved standard software engineering validation and verification processes rather than data-driven model training.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth Establishment: Not applicable/Not specified. As there's no mention of a training set for a machine learning model, the method for establishing its "ground truth" is not relevant or provided in this document.

    In conclusion:

    The provided 510(k) summary for the Anythink® PACS Workstation demonstrates substantial equivalence primarily through a comparison of functional and technical characteristics with a legally marketed predicate device, the Siemens Syngo Multimodality Workplace. It does not present quantitative performance data from a clinical study, involve specific test sets of patient cases, or refer to machine learning model training/evaluation, which are common for AI/ML-based diagnostic devices today. The "acceptance criteria" here are met by demonstrating that the new device performs its intended functions (viewing, manipulation, communication, storage of DICOM images) reliably and does not raise new issues of safety or effectiveness compared to the predicate.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1