510(k) Data Aggregation

    K Number
    K073396
    Date Cleared
    2008-08-07

    (248 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The B. Braun Antibacterial Ultrasite Valve (Ultrasite Ag) is intended for aspiration, injection or gravity/pump flow of IV fluids upon insertion of a male luer fitting. Not intended for use with power injectors.

    The Antibacterial Ultrasite Valve contains a silver based compound which may help to reduce bacterial growth on the surface and within the fluid path of the device.

    Testing with Ultrasite Ag Valve demonstrated antibacterial effectiveness against Pseudomonas aeruginosa, Klebsiella pneumoniae, and Enterobacter aerogenes over 96 hours.

    The Antibacterial Ultrasite Valve is not intended for use in the treatment of existing infections.

    Device Description

    B. Braun's Antibacterial Ultrasite® Valve is an individually packaged, sterile, non-pyrogenic, single use, disposable device. The valve is used as an injection site and requires swabbing to disinfect prior to use. The Antibacterial Ultrasite Valve is a needle free reflux valve or needleless access device intended for IV therapy. The valve is accessed by attaching a male luer fitting.
    The Antibacterial Ultrasite Valve contains a silver based compound that is impregnated into the polymeric resin of the device components. The antimicrobial compound is intended to reduce bacterial growth.

    AI/ML Overview

    The provided document for K073396 describes the Antibacterial Ultrasite® Valve, Ultrasite Ag. This device is a needle-free reflux valve intended for IV therapy, incorporating a silver-based compound to reduce bacterial growth.

    Here's an analysis of the acceptance criteria and the study information based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criterion (Type of Test)Reported Device Performance
    BiocompatibilityTests conducted (details not provided in excerpt)
    ToxicologyTests conducted (details not provided in excerpt)
    Functional PerformanceTests conducted (details not provided in excerpt)
    Package IntegrityNone mentioned beyond "tests conducted"
    ShippingTests conducted (details not provided in excerpt)
    Microbial Ingress ChallengeTests conducted (details not provided in excerpt)
    Antimicrobial EffectivenessDemonstrated effectiveness against Pseudomonas aeruginosa, Klebsiella pneumoniae, and Enterobacter aerogenes over 96 hours.
    NeutralizationTests conducted (details not provided in excerpt)
    ElutionTests conducted (details not provided in excerpt)

    Important Note: The provided document is a 510(k) summary, which often summarizes testing without providing granular details of the acceptance criteria (e.g., specific quantitative thresholds for "functional performance" or "biocompatibility") or detailed results beyond a general statement of "demonstrated effectiveness" for the antimicrobial aspect.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for any of the tests aformentioned for the test set or the provenance of the data (e.g., country of origin, retrospective or prospective). It simply states that "The Antibacterial Ultrasite Valve was subjected to a variety of tests."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable and not provided in the document. This device is a medical device (intravascular administration set) that undergoes laboratory and performance testing, not human-based interpretive testing that would require expert consensus for ground truth establishment.

    4. Adjudication Method for the Test Set

    This information is not applicable and not provided in the document, as it pertains to human interpretation tasks.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC study was not conducted or mentioned. This type of study is more relevant for diagnostic imaging or interpretation devices where human readers' performance with and without AI assistance is evaluated.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in essence, the "Antimicrobial Effectiveness" testing can be considered analogous to "standalone" performance for this device's specific claim. The device itself (the valve with the silver compound) was directly tested for its ability to reduce bacterial growth. This testing was performed independent of human interaction or interpretation beyond laboratory procedures.

    7. The Type of Ground Truth Used

    For the antimicrobial effectiveness claim, the ground truth was based on microbiological testing results. This typically involves:

    • Inoculating the device (or control) with specific bacteria.
    • Incubating for a defined period.
    • Measuring bacterial viability/reduction compared to control conditions.
    • The reduction in bacterial count serves as the "ground truth" for the device's antibacterial efficacy.

    For other tests (biocompatibility, functional performance, etc.), the ground truth would be established by validated laboratory methods and predetermined specifications/standards for those types of tests.

    8. The Sample Size for the Training Set

    This information is not applicable and not provided. This device is a physical medical device, not an AI or machine learning algorithm that requires a "training set" in the computational sense.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as there is no "training set" for this type of device.