510(k) Data Aggregation
(241 days)
ANNE One is a wireless monitoring platform indicated for the measurement of electrocardiography (ECG) waveforms. heart rate, respiratory rate, function of arterial hemoglobin (SpO2), pulse rate, activity, body position, fall detection, skin temperature by qualified healthcare professionals in home and healthcare settings. ANNE One is compatible with third-party, FDA-cleared devices for noninvasive blood pressure, SpO2, pulse rate, and body temperature measurements. The device is indicated for monitoring ECG waveforms and heart rate on ambulatory patients. The device is not intended to monitor or measure respiratory rate. SpO2, pulse rate, or noninvasive blood pressure while the patient undergoes significant motion or is active.
ANNE One continuously monitors the orients to aid in the prevention of pressure ulcers for at-risk patients. The system provides visual notification when the pation has not changed from a preset threshold of time.
The device is intended for use on general care patients who are 12 years of age or older as a general patient monitor to provide continuous physiological information as an aid to diagnosis and treatment. The data from ANNE One are transmitted wirelessly for display, storage, and analysis. The device is not intended for use on critical care patients.
ANNE One is a wireless monitoring platform that streams and stores real-time biosignals including electrocardiography (ECG), photoplethysmography (PPG), 3-axis accelerometry, and temperature to measure vital signs such as heart rate, respiratory rate, SpO2, pulse rate, skin temperature, and body temperature. The ECG signal is not intended for automated arrhythmia detection or classification; rather it is intended for manual interpretation, and the automated computation of heart rate through QRS identification using the well-known Pan-Tompkins beat detection algorithm. The displayed waveform is only intended for display as a check for normal ECG rhythm. The waveform is not intended for manual discrimination of any arrhythmias or cardiac conditions. The system features two skin-mounted, bio-integrated sensors that pair with the ANNE View software application for the continuous display, storage, and analysis of vital sign measurements and physiological waveforms. The system is also compatible with optional FDA-cleared third-party devices for SpO2, non-invasive blood pressure, and body temperature measurements.
Here's an analysis of the acceptance criteria and the studies that prove the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Parameter | Acceptance Criteria | Reported Device Performance | Study Type / Details |
|---|---|---|---|
| Heart Rate | Range: 30-300 bpm | 30-300 bpm | Safety and performance testing of ECG per IEC 60601-2-27:2011 and IEC 60601-2-47:2012. |
| Accuracy: The greater of ±10% or ± 5 bpm | The greater of ±10% or ± 5 bpm | Safety and performance testing of ECG per IEC 60601-2-27:2011 and IEC 60601-2-47:2012. | |
| Respiratory Rate | Range: 8-30 bpm | 8-30 bpm | Clinical Study: Comparison to etCO2 |
| Accuracy: ±3 bpm (Mean Absolute Error - MAE) | Mean absolute error within ±3 bpm | Clinical Study: Comparison to etCO2 | |
| Skin Temperature | Range: 73.4°F - 109.4°F (23°C - 43°C) | 73.4°F - 109.4°F (23°C - 43°C) | Performance testing (bench testing implied) |
| Accuracy: ±0.54°F (±0.3°C) | ±0.54°F (±0.3°C) | Performance testing (bench testing implied) | |
| SpO2 | Range: 70-100% | 70-100% | Clinical Study: Comparison to blood gas analysis |
| Accuracy: ARMS ≤ 3% | ARMS = 2.31% | Clinical Study: Comparison to blood gas analysis | |
| Pulse Rate | Range: 30-300 bpm | 30-300 bpm | Safety and performance testing of pulse oximeter per ISO 80601-2-61:2017. |
| Accuracy: The greater of ±10% or ± 5 bpm | The greater of ±10% or ± 5 bpm | Safety and performance testing of pulse oximeter per ISO 80601-2-61:2017. | |
| Body Position/Fall Detection | Continuous monitoring, visual notification for unchanged position | Continuous monitoring, visual notification | Performance testing (bench testing implied); supported by reference devices. |
| Activity | Measurement via Accelerometer | Accelerometer-based | Performance testing (bench testing implied); supported by reference devices. |
| ECG Waveform Display | Compliant to IEC 60601-2-27 and IEC 60601-2-47 | Compliant to IEC 60601-2-27 and IEC 60601-2-47 | Safety and performance testing of ECG per IEC 60601-2-27:2011 and IEC 60601-2-47:2012. |
Note: The document primarily focuses on demonstrating substantial equivalence to predicate and reference devices, with specific performance values provided for SpO2 and Respiratory Rate from clinical studies. The other parameters are stated to meet relevant standards or are equivalent to predicate/reference device performance.
2. Sample Size Used for the Test Set and Data Provenance
-
SpO2 Accuracy Study:
- Sample Size: n=12 healthy subjects.
- Data Provenance: Not explicitly stated (e.g., country of origin), but implies a controlled clinical setting. The study is described as "Sibel validated the accuracy...".
- Retrospective/Prospective: Implied prospective as it's a validation study conducted by Sibel.
-
Respiratory Rate Accuracy Study:
- Sample Size: n=40 healthy adult subjects.
- Data Provenance: Not explicitly stated (e.g., country of origin), but implies a controlled clinical setting. The study is described as "Sibel validated the accuracy...".
- Retrospective/Prospective: Implied prospective as it's a validation study conducted by Sibel.
-
Other Parameters: For other parameters like Heart Rate, ECG waveform display, Activity, etc., "performance testing" and adherence to "consensus standards" are mentioned, but specific sample sizes for these test sets are not provided in this summary. The comparison table also mentions "ambulatory databases" for ECG during motion, but no sample size is given.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
- SpO2 Accuracy Study: Ground truth was established by "blood gas analysis." This typically involves laboratory analysis, not expert consensus on visual assessment.
- Respiratory Rate Accuracy Study: Ground truth was established by "etCO2" (end-tidal CO2). This is a physiological measurement, not an expert consensus.
- Other Parameters: No information is provided regarding experts or their qualifications for establishing ground truth for other parameters. Ground truth for these values would likely derive from established measurement techniques compliant with the referenced standards.
4. Adjudication Method for the Test Set
- Given that the ground truth for SpO2 and Respiratory Rate relied on objective physiological measurements (blood gas analysis and etCO2, respectively), there was likely no "adjudication method" in the sense of reconciling disagreements between multiple graders or clinicians. The measurements themselves serve as the ground truth.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC comparative effectiveness study involving human readers and AI assistance is mentioned in the provided text. The device is a "wireless monitoring platform" for physiological measurements. Its primary function is to collect and display vital sign data, not to interpret complex medical images or data that typically require a human reader for adjudication or enhancement by AI.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Yes, the performance data provided for SpO2 and Respiratory Rate are standalone algorithm capabilities. The device measures these parameters and the reported accuracy is the algorithm's performance against a reference standard, without human intervention in the measurement or calculation process. The description of "automated computation of heart rate through QRS identification using the well-known Pan-Tompkins beat detection algorithm" also indicates a standalone algorithmic function.
7. The Type of Ground Truth Used
- SpO2: Blood gas analysis (objective physiological measurement).
- Respiratory Rate: EtCO2 (objective physiological measurement).
- Heart Rate: Implied to be derived from ECG signals, with validation against established standards (e.g., IEC 60601-2-27, IEC 60601-2-47), which would use a recognized reference for HR. The Pan-Tompkins algorithm is for beat detection, which is then used to compute HR.
- Other Parameters (Skin Temperature, Body Position, Activity, ECG Waveform Display, Pulse Rate): Ground truth is likely established through a combination of physical reference measurements and adherence to recognized consensus standards (e.g., ISO, IEC). Bench testing is mentioned for several parameters.
8. The Sample Size for the Training Set
- The document does not explicitly state the sample size for any training set for machine learning models. The device description mentions the use of "the well-known Pan-Tompkins beat detection algorithm" for heart rate, which is a classical signal processing algorithm and may not require a 'training set' in the modern machine learning sense. While algorithms are likely involved in respiratory rate and SpO2 calculations, the summary focuses on validation, not the development or training phase.
9. How the Ground Truth for the Training Set Was Established
- As training set information is not provided, the method for establishing its ground truth is also not detailed.
Ask a specific question about this device
(138 days)
ANNE One is a wireless vital signs and physiological data monitoring platform indicated for the measurement of heart rate, respiratory rate, step count, fall count, skin temperature, and body temperature by qualified healthcare professionals in healthcare settings. The device is intended for use on general care patients who are 18 years of age or older as a general patient monitor to provide continuous physiological information as an aid to diagnosis and treatment. The device is not intended for use on critical care patients. The device is not intended to monitor or measure respiratory rate or heart rate on ambulatory patients.
ANNE One is a wireless multi-parameter vital signs monitoring system that consists of the following components, and subsystems:
A. ANNE Tablet (Samsung Galaxy Tablet) - manufactured by Samsung
B. ANNE View Mobile Application (software) manufactured by Sibel Inc.
C. ANNE Chest Sensor manufactured by Sibel Inc.
D. ANNE Limb Sensor manufactured by Sibel Inc.
E. ANNE Wireless Charger manufactured by Sibel Inc.
The ANNE Chest Sensor is a battery-operated, skin-mounted sensor with wireless transceiver capabilities worn on the upper body to measure heart rate, respiratory rate, step count, fall count, and skin temperature.
The ANNE Limb Sensor is an additional battery-operated skin-mounted sensor with wireless transceiver capabilities worn on the finger to measure skin temperature. When placed underneath the axillae, the ANNE Limb Sensor can measure body temperature.
Both the ANNE Chest and ANNE Limb Sensors continuously gather vital signs data from the person being monitored and then transmit the encrypted data to the ANNE Tablet operating the ANNE View Mobile Application-a mobile device software application.
The provided text describes the ANNE One device, a wireless vital signs and physiological data monitoring platform, and its FDA 510(k) submission (K211305). While the document details various performance tests and validation methods, it does not provide specific acceptance criteria in a numerical format for each vital sign, nor does it present the reported device performance against such criteria in a table. It mostly states what was tested and how it was tested (e.g., according to specific ISO/IEC standards or against FDA-cleared reference devices).
Therefore, I cannot construct a table with acceptance criteria and reported device performance as requested, because that data is not explicitly present in the provided document. The document lists the types of measurements the device takes (heart rate, respiratory rate, step count, fall count, skin temperature, body temperature) and broadly states that bench and simulated use testing was conducted to demonstrate effectiveness and conformity to specifications.
However, I can extract information about the studies performed to demonstrate the device's capabilities:
Summary of Device Performance Studies for ANNE One (K211305)
The provided document describes various performance tests conducted to support the substantial equivalence determination for the ANNE One device. It does not explicitly state numerical "acceptance criteria" and "reported device performance" in a table format for each physiological parameter. Instead, it describes compliance with relevant standards and comparative studies against reference devices.
1. Table of Acceptance Criteria and Reported Device Performance
As mentioned, specific numerical acceptance criteria and reported performance values are not provided in the given text. The document broadly states that tests were performed to verify device specifications and ensure mechanical and electrical requirements were met, and that effectiveness was validated according to standards or against reference devices.
For example, for temperature measurement, it states: "Laboratory accuracy testing according to ISO 80601-2-56:2017 Section 201.101.2 validated the effectiveness of the ANNE Chest and Limb sensor for skin temperature measurements and the Limb Sensor for body temperature measurements when placed under the axillae." This indicates compliance with the standard but doesn't give a numerical accuracy. Similar statements are made for heart rate and respiratory rate.
2. Sample Sizes Used for the Test Set and Data Provenance
- Heart Rate:
- Simulated Use Study: 35 healthy subjects.
- Data Provenance: Not explicitly stated, but clinical studies are generally prospective. The phrase "healthy subjects" implies volunteers, likely from a single or a few sites.
- Respiratory Rate:
- Study: 50 healthy subjects.
- Data Provenance: Not explicitly stated, but generally prospective for such studies.
- Skin Temperature, Step Count, Fall Detection:
- Simulated Use Testing: n=10 subjects.
- Data Provenance: Not explicitly stated, but generally prospective.
- Biocompatibility Testing: All patient-contacting materials were tested.
- Electrical Safety & EMC Testing: Conducted on the device.
- Software Verification & Validation Testing: Conducted on the software.
- Animal Studies: Not required or performed.
- Clinical Studies: Not required or performed for safety and effectiveness. (Instead, substantial equivalence was supported by Bench and Simulated Use testing).
- Usability Testing: Conducted, but sample size not specified.
3. Number of Experts Used to Establish Ground Truth and Qualifications
The document does not mention the use of experts to establish a "ground truth" in the context of human reader studies (e.g., for image interpretation). Instead, the ground truth for physiological measurements was established using FDA-cleared reference devices or manual observation.
4. Adjudication Method for the Test Set
Not applicable. The studies described are not "human-in-the-loop" studies requiring expert adjudication of reader performance. Measurements were compared against reference devices or manual counts.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not performed, nor was it relevant to this device's function. The ANNE One is a vital signs monitoring device, not an imaging analysis AI intended to assist human readers in diagnosis. Therefore, there is no information on how human readers might improve with AI assistance. The performance studies focused on the accuracy of the device's measurements compared to established reference methods.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the performance studies described (accuracy of heart rate, respiratory rate, temperature, step count, fall count) essentially describe the standalone performance of the ANNE One system in measuring these vital signs. The device generates the data, and its accuracy is validated against gold standards (reference devices or manual counts). There isn't a human processing the algorithm's output for further interpretation in the way one might with an AI for medical image analysis.
7. The Type of Ground Truth Used
The ground truth for the vital signs measurements was primarily established using:
- FDA-cleared reference devices: For heart rate (against an FDA-cleared reference device) and respiratory rate (against an FDA-cleared End Tidal Carbon Dioxide (EtCO2) reference device).
- Reference standard/protocol conformance: For temperature (Laboratory accuracy testing according to ISO 80601-2-56:2017).
- Manual counts: For step and fall counts, demonstrated to conform to design specifications through comparison to manual counts.
8. The Sample Size for the Training Set
The document does not specify any training set size. This is likely because the device is a hardware sensor system with associated software for data collection and display, rather than a machine learning model that requires a large separate training dataset in the typical sense of AI/ML-driven diagnostics. The "software verification and validation testing" refers to the traditional software development lifecycle testing, not machine learning model training.
9. How the Ground Truth for the Training Set Was Established
As no training set is mentioned in the context of a machine learning model, this question is not applicable. The device's functionality is based on established physiological measurement principles and signal processing, not a trained AI model in the common sense.
Ask a specific question about this device