Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K171857
    Date Cleared
    2018-03-30

    (282 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AMSure® Enteral Feeding Syringe with ENFit Tip is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syringes are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians in adults only.

    Device Description

    The AMSure® Enteral Feeding Syringes with ENFit Tip are standard piston style syringes consisting of a syringe barrel with ENFit syringe nozzle, syringe plunger, syringe stopper and protective cap. The devices are made with polypropylene and polyisoprene rubber. The plunger can be linearly pulled and pushed along the inside wall of the barrel, allowing the syringe to take in and expel fluid through the nozzle at the front end of the barrel. The integral syringe nozzle has a female ENFit connector designed to be compatible only with enteral access devices or accessories having ENFit compliant or compatible male connectors to form a dedicated system that prevents wrong-route administration of fluids. The Enteral Feeding Syringes with ENFit Tip have a capacity of 60mL for single-use. The devices are provided either sterilized by EtO or non-sterile.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device, specifically an "AMSure® Enteral Feeding Syringe with ENFit Tip." This type of document focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than detailing studies and acceptance criteria for a novel AI/software device.

    Therefore, the specific information requested, such as acceptance criteria table with reported device performance for an algorithm, sample sizes for test/training sets for an AI model, data provenance, number and qualifications of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and how ground truth was established for training sets, are not applicable to this type of medical device submission.

    The "studies" and "acceptance criteria" discussed in this document refer to traditional medical device testing, primarily non-clinical, to ensure the physical and biological safety and performance of the syringe itself.

    However, I can extract the information that is present concerning the non-clinical testing performed on the AMSure® Enteral Feeding Syringe with ENFit Tip, which serves a similar purpose to demonstrating a device meets its intended performance.

    Here's the relevant information based on the provided text, re-interpreting "acceptance criteria" and "study" in the context of a physical medical device submission:


    1. A table of acceptance criteria and the reported device performance

    The document lists various performance characteristics that were tested. For a 510(k) submission, "acceptance criteria" typically refers to established standards (e.g., ISO standards) or pre-defined thresholds that the device must meet to be considered safe and effective and substantially equivalent to its predicate. The document states a general conclusion that the device "passed all acceptance criteria" without specifying the exact numerical criteria or individual results for each test.

    Test CategorySpecific Test (where listed)Acceptance Criteria (General, based on passing relevant standards)Reported Device Performance
    BiocompatibilityCytotoxicityMeets ISO 10993-5 standardsPassed
    SensitizationMeets ISO 10993-10 standardsPassed
    IrritationMeets ISO 10993-10 standardsPassed
    Syringe Tip (ENFIT) per ISO 80369-3Dimensional analysisMeets ISO 80369-3 standardsPassed
    Enteral Connector Misconnection assessmentPrevents misconnection as per ISO 80369-3Passed
    Fluid leakageWithin acceptable limits as per ISO 80369-3Passed
    Stress CrackingMeets ISO 80369-3 standardsPassed
    Resistance to separation from axial loadMeets ISO 80369-3 standardsPassed
    Resistance to separation from unscrewingMeets ISO 80369-3 standardsPassed
    Resistance to overridingMeets ISO 80369-3 standardsPassed
    Disconnection by unscrewingMeets ISO 80369-3 standardsPassed
    Dose accuracyMeets ISO 80369-3 standardsPassed
    Syringe per ISO 7886-1Appearance performanceMeets ISO 7886-1 standardsPassed
    Capacity ToleranceMeets ISO 7886-1 standardsPassed
    Graduated ScaleMeets ISO 7886-1 standardsPassed
    Critical DimensionMeets ISO 7886-1 standardsPassed
    Freedom from air and liquid leakageMeets ISO 7886-1 standardsPassed
    Tolerance on graduated capacityMeets ISO 7886-1 standardsPassed
    Dead spaceMeets ISO 7886-1 standardsPassed
    Piston/plunger assembly (Piston sliding, Piston Fit in Barrel)Meets ISO 7886-1 standardsPassed
    Silicone oil quantityMeets ISO 7886-1 standardsPassed
    Chemical (Limits for acidity or alkalinity, Limits for extractable metals)Meets ISO 7886-1 standardsPassed
    Ink adhesionMeets ISO 7886-1 standardsPassed
    Additional TestingStability testingEnsures performance over timePassed
    Sterilization validationMeets sterilization standards (e.g., EtO efficacy)Passed
    Package testingEnsures package integrity and sterilityPassed
    Dose accuracy testing (additional)Confirms accurate fluid deliveryPassed
    Risk managementDemonstrates identified risks are mitigatedPassed
    Usability testing (for cleaning instructions)Confirms clear and effective cleaning instructionsPassed
    Cleaning validation for reprocessingValidates cleaning procedure per FDA guidance for 8 uses/washesPassed
    Performance testing following 8 uses and washes within 24 hoursMaintains performance after reprocessingPassed

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document for the non-clinical tests. Sample sizes for physical device testing typically refer to the number of physical devices subjected to each test. The provenance would be the manufacturing site (Amsino International Inc. in Pomona, CA). The tests are prospective in nature, as they are performed on new devices to demonstrate compliance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as the device is a physical enteral feeding syringe, not an AI/software device requiring expert ground truth for image interpretation or similar tasks.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable as the device is a physical enteral feeding syringe, not an AI/software device requiring expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable as the device is a physical enteral feeding syringe, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is not applicable as the device is a physical enteral feeding syringe, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For manufacturing quality and performance testing of a physical device, the "ground truth" is typically defined by established engineering specifications, recognized national/international standards (e.g., ISO, ASTM), and regulatory guidance documents (e.g., FDA guidance). Device performance is measured against these objective criteria.

    8. The sample size for the training set

    This is not applicable as the device is a physical enteral feeding syringe, not an AI/software device that requires a training set.

    9. How the ground truth for the training set was established

    This is not applicable as the device is a physical enteral feeding syringe, not an AI/software device that requires a training set and associated ground truth.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1