Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K173665
    Date Cleared
    2018-09-05

    (280 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    AMSafe Sure-Lok Needle-Free Connector

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AMSafe® Sure-Lok™ Needle-Free Connector is an accessory to an Intra-vascular Administration Set used as a secondary sterile injection site for delivery of fluids to a patient's vascular system through a cannula inserted into a vein.

    Device Description

    The AMSafe® Sure-Lok™ Needle-Free Connector is a Needle-Free connector intended for single patient use, for direct injection, intermittent infusion, continuous infusion or aspiration of drugs, blood and fluids used as an accessory to an Intravascular administration set without the use of a needle, thus eliminating the risk of needlestick injuries. The AMSafe® Sure-Lok™ Needle-Free Connector is a closed, Luer activated device that allows for negative fluid displacement. The AMSafe® Sure-Lok™ Needle-Free Connector individually packaged is supplied as a sterile, non-pyrogenic, single use, disposable device.

    AI/ML Overview

    The provided text is a 510(k) Summary for the AMSafe® Sure-Lok™ Needle-Free Connector, a medical device. This document describes the device, its intended use, and the testing conducted to demonstrate its safety and effectiveness in comparison to a predicate device.

    Here's an analysis of the acceptance criteria and study information, organized as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    Testing ItemsTest Criteria (Acceptance Criteria)Reported Device Performance
    Visual inspectionConform with the internal criteriaMet the acceptance criterion
    Leakage testConform with ISO 8536-4 Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed [Including: Amendment 1 (2013)]; Conform with ISO 8536-8: 2015 Infusion Equipment for Medical Use - Part 8: Infusion Sets for Single Use with Pressure Infusion ApparatusMet the acceptance criterion
    Compatible with MLL fittingSection 3: Dimensions and tolerances in ISO 594-2:1998 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings)Met the acceptance criterion
    Sealing performanceSection 4.2: leakage in ISO 594-2:1998 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings)Met the acceptance criterion
    Free Flow RateConform with ISO 8536-4:2010 Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed [Including: Amendment 1 (2013)Met the acceptance criterion
    Liquid leakageConform with ISO80369-7: 2016 Small-Bore Connectors For Liquids And Gases In Healthcare Applications - Part 7: Connectors For Intravascular or Hypodermic ApplicationsMet the acceptance criterion
    Air leakageConform with ISO80369-7: 2016 Small-Bore Connectors For Liquids And Gases In Healthcare Applications - Part 7: Connectors For Intravascular or Hypodermic ApplicationsMet the acceptance criterion
    Separation forceConform with ISO80369-7: 2016 Small-Bore Connectors For Liquids And Gases In Healthcare Applications - Part 7: Connectors For Intravascular or Hypodermic ApplicationsMet the acceptance criterion
    Stress crackingConform with ISO80369-7: 2016 Small-Bore Connectors For Liquids And Gases In Healthcare Applications - Part 7: Connectors For Intravascular or Hypodermic ApplicationsMet the acceptance criterion
    Unscrewing torqueConform with ISO80369-7: 2016 Small-Bore Connectors For Liquids And Gases In Healthcare Applications - Part 7: Connectors For Intravascular or Hypodermic ApplicationsMet the acceptance criterion
    Ease of assemblySection 5.6: Ease of assembly in ISO 594-2:1998 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings)Met the acceptance criterion
    Resistance to overridingConform with ISO80369-7: 2016 Small-Bore Connectors For Liquids And Gases In Healthcare Applications - Part 7: Connectors For Intravascular or Hypodermic ApplicationsMet the acceptance criterion
    Weld strength testConform with the internal criteriaMet the acceptance criterion
    CytotoxicityISO 10993-5: 2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro CytotoxicityMet the acceptance criterion
    SensitizationISO 10993-10: 2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin SensitizationMet the acceptance criterion
    Intracutaneous Reactivity TestISO 10993-10: 2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin SensitizationMet the acceptance criterion
    Acute systemic toxicityISO 10993-11: 2006 Biological evaluation of medical devices - Part 11: Tests for systemic toxicityMet the acceptance criterion
    HaemolysisASTM F 756-13: Standard Practice for Assessment of Hemolytic Properties of MaterialsMet the acceptance criterion
    PyrogenicityISO 10993-11: 2006 Biological evaluation of medical devices - Part 11: Tests for systemic toxicityMet the acceptance criterion
    Subchronic ToxicityISO 10993-11: 2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicityMet the acceptance criterion
    SterilityISO 11135:2007, Sterilization of Health Care Products-Ethylene Oxide; ISO 10993-7: 1995 and 2008, Biological Evaluation of Medical Devices-Part 7: Ethylene Oxide sterilization residuals.Assessed in accordance with standards and met acceptance criteria. Based on the conclusion, it implies that the device is sterile after processing.
    Microbial ingress testingInternal standardMet the acceptance criterion
    Biocompatibility (Overall Assessment)ISO 10993-1:2009 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process (for "External communicating device, Blood path, indirect (>24 hours to 30 days)")Assessed in accordance with the standard. Results from Cytotoxicity, Sensitization, Intracutaneous Reactivity Test, Acute systemic toxicity, Haemolysis, Pyrogenicity, and Subchronic Toxicity demonstrate no new issues of safety and efficacy.

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states that "Bench testing of the proposed devices is performed," and tests were conducted according to various ISO and ASTM standards. However, the specific sample sizes used for each individual test or for a comprehensive "test set" are not mentioned. The document does not specify data provenance (country of origin) or if the data was retrospective or prospective, as these are bench tests rather than clinical studies with patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This section is not applicable as the described studies are bench tests (physical and chemical testing of the device) rather than clinical studies requiring expert interpretation or ground truth establishment based on clinical observations or diagnoses.

    4. Adjudication Method for the Test Set

    This section is not applicable for the same reason as point 3. Bench testing adherence to standards usually relies on measurement and observation against pre-defined criteria, not expert adjudication of subjective findings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document describes bench testing and biocompatibility assessments, not comparative effectiveness studies with human readers or clinical outcomes. The goal was to demonstrate substantial equivalence to a predicate device through engineering and biological testing.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    This question is not applicable as the device is a physical medical connector, not an AI algorithm.

    7. The Type of Ground Truth Used

    For the bench testing, the "ground truth" is defined by the specific requirements and limits set forth in the referenced international standards (ISO, ASTM) and the manufacturer's own internal criteria. For example, for a leakage test, the ground truth is "no leakage" or leakage within a specified limit according to the standard. For biocompatibility tests, the ground truth is a lack of cytotoxic, sensitizing, or toxic effects as defined by the respective ISO standards.

    8. The Sample Size for the Training Set

    This question is not applicable. This is a physical medical device, not an AI or machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    This question is not applicable as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1