Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K093773
    Date Cleared
    2010-03-10

    (92 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    AMSINO I.V. ADMINISTRATION SET, AMSAFE3 I.V ADMINISTRATION SET, EXTENSION SETS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AMSINO® I.V. ADMINISTRATION SET is a device intended to administer fluids from a container to a patient's vascular system through a catheter inserted into a vein.

    Device Description

    The AMSINO® I.V. ADMINISTRATION SET is a single use, latex-free, Non-DEHP, gravity feed, sterile device sterilized with Ethylene Oxide Gas. It is used to administer fluids from a container to a patient's vascular system through a catheter inserted into a vein. It is comprised of various components such as: bag, spike, drip chamber (vented and non-vented), Y-site, burette, tubing, flow controller, drip selector, clamp, check valve, latex-free injection site, needleless injection site, flashbulb, filter, manifold, stopcock, flash bulb, luer connectors and bag hanger.

    This submission is an extension of the original approval (K973107) and covers the following AMSINO Product Line:

    AMSINO® STANDARD I.V. ADMINISTRATION SET AMSafe® I.V. ADMINISTRATION SET (identical to the Amsino Standard I.V. Administration set, except it is marketed for Emergency Medical Services group/personnel)

    AMSafe3® I.V. ADMINISTRATION SET (identical to AmSafe I.V. Administration Set, with the exception of the drip selector option.)

    AMSINO offers both standard and custom sets with tubing of various sizes and lengths with a choice of 10, 15, 20 and 60 drops per ml to meet customer requirements and specifications. Customers may request I.V. Administration sets with varying configuration containing any combination of the parts per the table:

    Tubing
    Drip Chamber
    Needleless injection site
    Split Septum Injection Site
    Bag Spike
    Clamp
    Flow Controller
    Drip Selector
    Filter
    Rotating male luer lock
    Male or Female luer lock
    Check Valve
    Bag Hanger
    Y-site
    stopcock
    manifold

    AI/ML Overview

    The provided document describes a Special 510(k) Premarket Notification for a modification to the AMSINO® I.V. ADMINISTRATION SET. This submission primarily focuses on demonstrating substantial equivalence to a previously cleared device (K973107) rather than a de novo device requiring extensive clinical studies to establish safety and effectiveness.

    Therefore, the "study that proves the device meets the acceptance criteria" in the traditional sense of a human clinical trial is not applicable here. Instead, the submission relies on bench performance testing, biocompatibility, and sterilization validations to show that the modified device performs as intended and is as safe and effective as the predicate device.

    Here's an analysis of the provided information based on your requested categories:

    1. A table of acceptance criteria and the reported device performance

    The document lists several performance criteria the device meets, primarily referencing ISO standards and/or AMSINO's internal testing criteria. However, specific quantitative acceptance criteria values and corresponding reported device performance values are generally not provided in this summary document. Instead, it makes statements of compliance.

    Acceptance Criterion (Test Type)Reported Device Performance
    Bench Performance Testing:Meets requirements according to ISO 8536-4 (when appropriate) and/or AMSINO's testing and acceptance criteria
    Closure Piercing Device (Spike) FeaturesMeets requirements
    Air Inlet Device CharacteristicsMeets requirements
    Connector Performance (to prevent leakage)Meets requirements
    Drip Chamber and Drip Tube PerformanceMeets requirements
    Flow Regulator PerformanceMeets requirements
    Flow characteristicsMeets requirements
    Tensile Strength of ConnectorsMeets requirements
    Self-sealing injection site challenge testMeets requirements
    Injection Port Access to Failure (needleless)Meets requirements
    Biocompatibility & Hemocompatibility:Biocompatible and Hemocompatible; test results indicate neither sensitivity nor toxicity (based on ISO 10993-1 guidelines including Cytotoxicity, Sensitization, Irritation, Systemic Cytotoxicity, and Hemocompatibility testing)
    Sterility:Sterilized by Ethylene Oxide (validated per ISO 11135-1:2007)
    Pyrogenicity:Tested for pyrogenicity (status: meets requirements)
    Microbial Ingress Challenge Test:Demonstrates a 4-log reduction of microorganisms against gram negative and gram positive organisms using proper aseptic technique

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes for the various bench tests, biocompatibility tests, or microbial ingress tests. It only states that these tests were conducted and the device "meets requirements" or "demonstrates a 4-log reduction." The data provenance is internal to Amsino International, Inc. (Pomona, CA, USA) and the testing would be considered prospective for the purpose of this submission, meaning the tests were conducted specifically to support this application.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This section is not applicable to this type of device and submission. The "ground truth" for the performance criteria is established by recognized international standards (like ISO) and internal engineering/quality specifications, not expert consensus in the medical image interpretation sense. The experts involved would be the engineers and quality assurance personnel at Amsino International responsible for designing, testing, and validating the device and its manufacturing processes. Their qualifications are not detailed.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving interpretation of medical data (e.g., images) by multiple readers. This is not applicable to the bench testing, biocompatibility, or sterilization validation performed for this I.V. administration set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is an I.V. administration set, not an AI-powered diagnostic tool requiring human readability studies.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is based on established engineering and materials science principles, international standards (e.g., ISO 8536-4, ISO 10993-1, ISO 11135-1), and Amsino's internal product specifications. This would involve:

    • Physical/Mechanical Properties: Measured values against predefined tolerance ranges.
    • Biocompatibility: Laboratory test results (e.g., cell viability, irritation scores) compared against acceptable limits derived from ISO 10993.
    • Sterility: Sterility assurance levels (SAL) achieved through validated processes per ISO 11135-1.
    • Microbial Ingress: Measured reduction in microbial count against a defined log reduction target.

    8. The sample size for the training set

    This is not applicable. This is a physical medical device, not an AI/machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    This is not applicable for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K973107
    Manufacturer
    Date Cleared
    1997-10-02

    (44 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    AMSINO I.V. ADMINISTRATION SET

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AMSINO INTRAVASCULAR ADMINISTRATION SET IS A DEVICE INTENDED TO ADMINISTER FLUIDS FROM A CONTAINER TO A PATIENT'S VASCULAR SYSTEM THROUGH A CATHETER INSERTED INTO A VEIN.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) premarket notification decision letter from the FDA for the Amsino I.V. Administration Set. It states that the device is substantially equivalent to legally marketed predicate devices.

    This document describes a regulatory approval for an I.V. administration set, not a software device or an AI/ML medical device. Therefore, the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert qualifications, and MRMC studies is not applicable to this submission.

    The letter is a notification of the FDA's finding of substantial equivalence, which allows the company to market the device. It does not contain performance study data or acceptance criteria in the format typically provided for novel medical devices or AI/ML technologies.

    The "Indications For Use" provided in the enclosure merely describes the intended purpose of the device: "AMSINO INTRAVASCULAR ADMINISTRATION SET IS A DEVICE INTENDED TO ADMINISTER FLUIDS FROM A CONTAINER TO A PATIENT'S VASCULAR SYSTEM THROUGH A CATHETER INSERTED INTO A VEIN."

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1