510(k) Data Aggregation

    K Number
    K241564
    Date Cleared
    2024-08-02

    (63 days)

    Product Code
    Regulation Number
    888.3045
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Actifuse ABX; Actifuse MIS; Actifuse Shape; ALTAPORE; ALTAPORE SHAPE; ALTAPORE MIS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ACTIFUSE ABX, ACTIFUSE SHAPE and ACTIFUSE MIS:

    Actifuse™ is a bone void filler intended only for orthopaedic applications as a filler for gaps and voids that are not intrinsic to the stability of the bony structure. Actifuse is indicated to be packed gently into bony voids or gaps of the skeletal system, i.e., extremities, pelvis, and spine, including use in posterolateral fusion procedures with appropriate stabilizing hardware and intervertebral disc space. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.

    When used in the intervertebral disc space, Actifuse must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler.

    ALTAPORE:

    ALTAPORE is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine, intervertebral disc space, and pelvis). ALTAPORE can be used by itself, with autograft as a bone graft extender or with autogenous bone marrow aspirate. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPORE resorbs and is replaced with bone during the healing process.

    When used in the intervertebral disc space, ALTAPORE must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler.

    ALTAPORE SHAPE:

    ALTAPORE SHAPE is an implant intended to fill bony voids or gaps of the skeletal system (i.e. extremities, posterolateral spine, intervertebral disc space, and pelvis).

    ALTAPORE SHAPE can be used by itself or with autograft as a bone graft extender in posterolateral spinal fusion procedures. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPE resorbs and is replaced with bone during the healing process.

    When used in the intervertebral disc space. ALTAPORE SHAPE must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler.

    ALTAPORE MIS: ALTAPORE MIS is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine, intervertebral disc space, and pelvis).

    ALTAPORE MIS can be used by itself, with autograft as a bone graft extender or with autogenous bone marrow aspirate. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPORE MIS resorbs and is replaced with bone during the healing process.

    When used in the intervertebral disc space, ALTAPORE MIS must be used with an intervertebral body fusion device cleared by FDA for use with a bone void filler.

    Device Description

    ACTIFUSE ABX, ACTIFUSE SHAPE, ACTIFUSE MIS

    Actifuse is a phase-pure, porous, silicate-substituted calcium phosphate osteoconductive bone void filler, comprising a single-phase calcium phosphate scaffold, either granules or granules delivered in a matrix of resorbable polymer. 0.8wt% silicon (Si) is incorporated into the crystalline structure. The interconnected and open porous structure of Actifuse is similar to human cancellous bone. After it is implanted, ACTIFUSE undergoes physiologically-mediated resorption and is replaced by natural bone

    ACTIFUSE is osteostimulatory based upon in vitro studies that show cellular responses, such as metabolic activity and proliferation, are accelerated when compared to an identical that did not contain 0.8% silicon by weight..

    ACTIFUSE is bioactive based on in vitro studies that show it forms a surface apatite-layer when submerged in simulated body fluid that contains the same ion concentrations as human blood plasma. This apatite layer provides scaffolding onto which the patient's new bone will grow, allowing complete repair of the defect.

    ACTIFUSE is osteoconductive based on in vivo animal studies that show it achieves bone healing in a critical defect model as confirmed with radiographic, histolopathological, histomorphometric, and mechanical analyses.

    The Actifuse bone void filler is presented in multiple product variants:

    ACTIFUSE ABX

    ACTIFUSE ABX presents the ACTIFUSE bioactive and osteoconductive silicate-substituted calcium phosphate bone void filler suspended in an absorbable aqueous gel carrier packaged in a wide-bore applicator. ACTIFUSE ABX is a sterile, single-use device available in volumes of 1.5 ml, 2.5 ml, 5 ml, 10 ml, and 20 ml.

    ACTIFUSE ABX is designed for use as a standalone bone graft substitute or as an autograft extender. While not necessary, the product can be mixed with autologous blood, autologous bone marrow aspirate, or autologous bone at the discretion of the surgeon.

    ACTIFUSE MIS

    ACTIFUSE MIS System is an ACTIFUSE ABX packaging configuration in which the ACTIFUSE ABX implant material is supplied in a sterile 7.5ml cartridge packed with ACTIFUSE ABX implant material is presented with a delivery applicator and as a separate 7.5ml refill cartridge.

    ACTIFUSE SHAPE

    ACTIFUSE SHAPE supplies the ACTIFUSE bioactive and osteoconductive silicate-substituted calcium phosphate bone void filler suspended in an absorbable Alkylene Oxide co-polymer carrier as a shaped wax-like putty in a sterile pouch. ACTIFUSE SHAPE is available as a 1.6 ml small cylinder, 2.6 ml medium cylinder, 8 ml large cylinder, and 15.8 ml large strip.

    ACTIFUSE SHAPE is designed for use as a standalone bone graft substitute or as an autograft extender. While not necessary, the product can be mixed with autologous blood or autologous bone at the discretion of the surgeon.

    ALTAPORE

    ALTAPORE is a bioactive and osteoconductive silicate-substituted calcium phosphate bone void filler. The interconnected and open porous structure of the silicate-substituted calcium phosphate phase of ALTAPORE is similar to human cancellous bone and is intended to support bone growth with macro and micro- porosity. ALTAPORE is composed solely of elements that exist naturally in normal bone (Ca. P. O. H. Si).

    ALTAPORE is supplied in a sterile applicator and contains ALTAPORE microgranules, sized 1-2 mm, 80-85% total porosity, suspended in an absorbable aqueous gel carrier. ALTAPORE does not set in-situ following implantation. ALTAPORE is available in 1.5 ml, 2.5 ml, 5 ml, 10 ml, and 20 ml configurations.

    ALTAPORE is designed for use as a standalone bone graft substitute or as an autograft extender. While not necessary, the product can be mixed with Bone Marrow Aspirate (BMA) or autologous bone at the discretion of the surgeon.

    ALTAPORE is bioactive based on in vitro studies that show it forms a surface apatite layer when submerged in simulated body fluid that contains the same ion concentrations as human blood plasma. This apatite layer provides scaffolding onto which the patient's new bone will grow, allowing complete repair of the defect.

    ALTAPORE is osteoconductive based on in vivo animal studies that show it achieves bone healing in a critical defect model as confirmed with radiographic, histolopathological, histomorphometric, and mechanical analyses. ALTAPORE undergoes cell-mediated remodeling and is replaced by natural bone.

    ALTAPORE MIS

    ALTAPORE MIS System is an ALTAPORE packaging configuration in which the ALTAPORE implant material is supplied in a sterile 7.5ml cartridge is presented with a delivery applicator and as a separate 7.5ml refill cartridge.

    ALATAPORE MIS System (hereafter referred to as "Altapore MIS) is designed for use as a standalone bone graft substitute or as an autograft extender. While not necessary, the product can be mixed with Bone Marrow Aspirate (BMA) or autologous bone at the discretion of the surgeon.

    ALTAPORE MIS is bioactive based on in vitro studies that show it forms a surface apatite layer when submerged in simulated body fluid that contains the same ion concentrations as human blood plasma. This apatite layer provides scaffolding onto which the patient's new bone will grow, allowing complete repair of the defect.

    ALTAPORE SHAPE

    ALTAPORE SHAPE supplies the ALTAPORE bioactive and osteoconductive silicate-substituted calcium phosphate bone void filler suspended in an absorbable Alkylene Oxide co-polymer carrier as a shaped wax-like putty in a sterile pouch. ALTAPORE SHAPE is available as a 1.6 ml small cylinder, 2.6 ml medium cylinder, 8 ml large cylinder, and 15.8 ml large strip.

    ALTAPORE SHAPE is designed for use as a standalone bone graft substitute or as an autograft extender. While not necessary, the product can be mixed with autologous blood or autologous bone at the discretion of the surgeon.

    ALTAPORE SHAPE is bioactive based on in vitro studies that show it forms a surface apatitelayer when submerged in simulated body fluid that contains the same ion concentrations as human blood plasma. This apatite layer provides scaffolding onto which the patient's new bone will grow, allowing complete repair of the defect.

    ALTAPORE SHAPE is osteoconductive based on in vivo animal studies that show it achieves bone healing in a critical defect model as confirmed with radiographic, histolopathological, histomorphometric, and mechanical analyses. ALTAPORE SHAPE undergoes cell-mediated remodeling and is replaced by natural bone.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for several bone void filler devices (Actifuse ABX, Actifuse MIS, Actifuse Shape; ALTAPORE, ALTAPORE SHAPE; ALTAPORE MIS) by Baxter Healthcare Corporation. The submission aims to expand the indications for use to include intervertebral disc space applications.

    Crucially, the document explicitly states that "no additional non-clinical data was needed to support the expanded indications proposed in the current submission" because "the subject device is identical to the previously cleared primary predicate device (K232315)." This means that the current submission relies entirely on the performance data of the predicate device (Catalyst Bone Void Filler, OssDsign AB, K232315) and existing reference devices (Actifuse ABX, Actifuse Shape, & Actifuse MIS, Baxter, K082575; Altapore, Baxter, K192363; Altapore Shape, Baxter, K200640; Altapore MIS, Baxter, K201464).

    Therefore, I cannot provide detailed acceptance criteria or study results for the currently submitted devices directly from this document, as no new performance testing specific to this submission's expanded indication for these devices is reported. The information requested below would typically come from the original 510(k) submission for the predicate device (K232315) or the reference devices.

    However, I can extract the available information as specifically related to this submission's approach to demonstrating substantial equivalence.


    1. A table of acceptance criteria and the reported device performance

    As stated above, no new performance data or acceptance criteria are provided in this specific submission for the expanded indications. The submission argues for substantial equivalence based on the device being identical to a predicate (K232315) and similar to reference devices (K082575, K192363, K200640, K201464).

    Acceptance Criteria (Implicit via Substantial Equivalence to Predicate): The device is considered to meet acceptance criteria if its technological characteristics, intended use, and performance are substantially equivalent to the cleared predicate and reference devices, implying it is as safe and effective.

    Reported Device Performance:
    The document describes the characteristics and prior in vitro/in vivo findings of the devices, which would have supported their initial clearances, but does not present new performance test results against new acceptance criteria for the expanded indication.

    • ACTIFUSE ABX, ACTIFUSE SHAPE, ACTIFUSE MIS:
      • Osteostimulatory: Based on in vitro studies showing accelerated cellular responses (metabolic activity and proliferation) compared to an identical product without 0.8% silicon by weight.
      • Bioactive: Based on in vitro studies showing formation of a surface apatite layer in simulated body fluid.
      • Osteoconductive: Based on in vivo animal studies demonstrating bone healing in a critical defect model, confirmed by radiographic, histopathological, histomorphometric, and mechanical analyses.
    • ALTAPORE, ALTAPORE SHAPE, ALTAPORE MIS:
      • Bioactive: Based on in vitro studies showing formation of a surface apatite layer in simulated body fluid.
      • Osteoconductive: Based on in vivo animal studies demonstrating bone healing in a critical defect model, confirmed by radiographic, histopathological, histomorphometric, and mechanical analyses. Undergoes cell-mediated remodeling and is replaced by natural bone.

    The remaining points cannot be fully addressed from the provided text for this specific K-submission, as it explicitly states that no additional non-clinical data was needed. This implies that the current submission relies on previous data and FDA's prior determinations of substantial equivalence for the predicate and reference devices.

    To answer points 2-9, one would need to access the original 510(k) submissions for:

    • Primary Predicate: Catalyst Bone Void Filler, OssDsign AB, K232315
    • Reference Devices:
      • Actifuse ABX, Actifuse Shape, & Actifuse MIS, Baxter, K082575
      • Altapore, Baxter, K192363
      • Altapore Shape, Baxter, K200640
      • Altapore MIS, Baxter, K201464

    Without those specific predicate/reference 510(k) summaries, I cannot provide the requested details regarding sample sizes, data provenance, ground truth establishment, or specific study methodologies.

    Summary based on information provided for the current K241564 submission:

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    * N/A: No new test set data was reported as needed for this submission. Reliance on previously cleared predicate (K232315) and reference devices (K082575, K192363, K200640, K201464).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    * N/A: No new ground truth establishment reported as needed for this submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    * N/A: No new test set reported as needed for this submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    * N/A: This device is a bone void filler, not an AI/imaging device, so MRMC studies are not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    * N/A: This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    * N/A: No new ground truth reported as needed for this submission. However, for the original clearances of the predicate/reference devices, the descriptions mention "radiographic, histopathological, histomorphometric, and mechanical analyses" in animal studies, which serve as forms of ground truth for osteoconductivity.

    8. The sample size for the training set
    * N/A: This is a physical medical device, not an AI/ML algorithm requiring a training set.

    9. How the ground truth for the training set was established
    * N/A: Not applicable for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200640
    Device Name
    ALTAPORE SHAPE
    Date Cleared
    2020-05-08

    (58 days)

    Product Code
    Regulation Number
    888.3045
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ALTAPORE SHAPE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ALTAPORE SHAPE is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine and pelvis).

    ALTAPORE SHAPE can be used by itself or with autograft as a bone graft extender in posterolateral spinal fusion procedures. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPORE SHAPE resorbs and is replaced with bone during the healing process.

    Device Description

    ALTAPORE SHAPE is a bioactive and osteoconductive silicate-substituted calcium phosphate bone void filler. The interconnected and open porous structure of the silicatesubstituted calcium phosphate phase of ALTAPORE SHAPE is similar to human cancellous bone and is intended to support bone growth with macro and micro-porosity. The microgranule phase ALTAPORE SHAPE is composed solely of elements that exist naturally in normal bone (Ca, P, O, H, Si).

    ALTAPORE SHAPE is supplied as a shaped wax-like putty in a sterile pouch containing ALTAPORE microgranules, sized 1-2 mm, 80-85% total porosity, suspended in an absorbable Alkylene Oxide co-polymer carrier. ALTAPORE does not set in-situ following implantation. ALTAPORE is available as a 1.6 ml small cylinder, 2.6 ml medium cylinder, 8 ml large cylinder, and 15.8 ml large strip.

    ALTAPORE SHAPE is designed for use as a standalone bone graft substitute or as an autograft extender. While not necessary, the product can be mixed with autologous blood or autologous bone at the discretion of the surgeon.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Altapore Shape device, which is a resorbable calcium salt bone void filler. The submission focuses on aligning the indications for use of Altapore Shape for posterolateral spinal fusion procedures with a previously cleared predicate device, Altapore (K192363).

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance results for this submission (K200640). Instead, it relies on the substantial equivalence to previously cleared predicate devices. The key "performance criterion" for this submission is demonstrating that the change in indications (specifically referring to its use in posterolateral spinal fusion as a stand-alone graft) does not alter the safety and effectiveness, leveraging prior studies.

    However, the document mentions that for the initial clearance of Altapore Shape (K191513) and Altapore (K192363), various tests were conducted and met acceptance criteria. Based on the provided text, the relevant performance characteristics and implicitly met acceptance criteria are:

    Acceptance Criteria (Implied)Reported Device Performance (Reference to K191513 and K192363)
    Sterility: Achieve a 10-6 Sterility Assurance Level (SAL).Established and validated at the manufacturing facility as described in ANSI/AAMI/ISO ISO 11137-2. Labels "Sterile". Package verification based on Visual Inspection, Seal Strength, and Bubble Leak testing.
    Shelf Life: Maintain stability over time.A shelf life claim of 5 years is substantiated by stability results presented under K191513.
    Bioactivity: Forms a surface apatite-layer.Bioactive based on in vitro studies that show it forms a surface apatite-layer when submerged in simulated body fluid (contains same ion concentrations as human blood plasma).
    Osteoconductivity and Bone Healing: Supports bone growth and is replaced by natural bone in a critical defect model.Osteoconductive based on in vivo animal studies that show it achieves bone healing in a critical defect model, confirmed with radiographic, histopathologic, histomorphometric, and mechanical analyses. Undergoes cell-mediated remodeling and is replaced by natural bone. This was demonstrated for Altapore Shape (K191513) and for Altapore (K192363) for posterolateral spine fusion, and these findings are leveraged here.
    Biocompatibility: No adverse biological reactions.Biocompatibility assessments (Cytotoxicity, Sensitization, Irritation, System Toxicity, Pyrogen, Genotoxicity, Implantation) were conducted based on ISO-10993-1 and FDA guidance for K191513 and are unchanged.
    (For this submission) Equivalence of Posterolateral Spine Fusion Indication: Safe and effective for stand-alone use in posterolateral spinal fusion.The preclinical animal study conducted for Altapore (K192363) on posterolateral spine fusion, which demonstrated effectiveness as a bone graft extender, is "presented again to support the performance of ALTAPORE SHAPE as a stand-alone graft substitute in posterolateral spinal fusion procedures." The hydroxyapatite granules are identical to ALTAPORE, and carrier phase differences do not alter therapeutic use or safety/effectiveness.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a distinct "test set" in the context of this 510(k) submission (K200640) for human clinical trials. The submission focuses on leveraging previous substantial equivalence.

    • Animal Studies: A preclinical animal study was conducted to evaluate the effectiveness of Altapore (or similar material) for posterolateral spinal fusion. The sample size for this animal study is not specified in the provided text. The data provenance would be prospective (animal study) from wherever the animal studies were conducted (country of origin not specified).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the text. For animal studies, the "ground truth" (e.g., bone healing confirmation) would typically be established by veterinary pathologists or histomorphometrists.

    4. Adjudication Method

    This information is not provided in the text. For animal studies, typically, the assessment of outcomes (e.g., histopathology, radiological findings) is performed by qualified individuals, but a formal "adjudication method" like 2+1 or 3+1 among experts is usually not described in such summaries unless there's a specific need for it in complex interpretations.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not mentioned or implied in the provided text. This type of study is more common for diagnostic AI devices, not directly for a bone void filler material itself, especially in a 510(k) submission focused on material equivalence and a slight indication expansion based on prior animal data.

    6. Standalone Performance Study

    Yes, a standalone performance evaluation was done through the preclinical animal study and the in vitro studies.

    • In vitro studies: Demonstrated bioactive properties (apatite layer formation).
    • Animal study: Evaluated effectiveness for use in posterolateral fusion. This study directly assessed the material's ability to achieve bone healing. While the exact details of how "performance" was measured (e.g., fusion rates, bone density) are not given, the study was conducted on the device material itself.

    7. Type of Ground Truth Used

    For the animal studies mentioned:

    • Histopathology: Assessing tissue changes at a microscopic level.
    • Histomorphometry: Quantitative analysis of tissue structure (e.g., amount of new bone formation).
    • Radiographic analysis: Imaging techniques to visualize bone healing.
    • Mechanical analyses: Testing the strength of the healed bone segment.

    These methods collectively served as the "ground truth" to determine bone healing and fusion in the animal model.

    8. Sample Size for the Training Set

    The document does not mention a training set. This is because Altapore Shape is a medical device (a material), not an AI algorithm that requires a training set for model development. The performance is based on its material properties and biological interactions demonstrated in non-clinical studies.

    9. How the Ground Truth for the Training Set Was Established

    As there is no mention of a training set (relevant for AI), this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191513
    Device Name
    Altapore Shape
    Date Cleared
    2019-10-19

    (134 days)

    Product Code
    Regulation Number
    888.3045
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Altapore Shape

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ALTAPORE SHAPE is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, pelvis and posterolateral spine).

    ALTAPORE SHAPE must be used in combination with autograft as a bone graft extender in posterolateral spinal fusion procedures. These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPORE resorbs and is replaced with bone during the healing process.

    Device Description

    ALTAPORE SHAPE is a bioactive and osteoconductive silicate-substituted calcium phosphate bone void filler. The interconnected and open porous structure of the silicatesubstituted calcium phosphate phase of ALTAPORE SHAPE is similar to human cancellous bone and is intended to support bone growth with macro and micro-porosity. The microgranule phase ALTAPORE SHAPE is composed solely of elements that exist naturally in normal bone (Ca, P, O, H, Si).

    ALTAPORE SHAPE is supplied as a shaped wax-like putty in a sterile pouch containing ALTAPORE microgranules, sized 1-2 mm, 80-85% total porosity, suspended in an absorbable Alkylene Oxide co-polymer carrier. ALTAPORE does not set in-situ following implantation. ALTAPORE is available as a 1.6 ml small cylinder, 2.6 ml medium cylinder, 8 ml large cylinder, and 15.8 ml large strip.

    ALTAPORE SHAPE is designed for use as a standalone bone graft substitute or as an autograft extender. While not necessary, the product can be mixed with autologous blood or autologous bone at the discretion of the surgeon.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Altapore Shape device, a resorbable calcium salt bone void filler. The document focuses on demonstrating substantial equivalence to predicate and reference devices, outlining the device's composition, intended use, and nonclinical testing.

    Here's an analysis of the acceptance criteria and supporting studies based on the provided text:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly present a table of "acceptance criteria" with corresponding "reported device performance" in the traditional sense of a clinical study with primary endpoints. Instead, the "acceptance criteria" are implied by the various nonclinical tests performed to demonstrate substantial equivalence to legally marketed predicate devices. The "reported device performance" is the conclusion that the device met these implied criteria, making it substantially equivalent.

    Acceptance Criteria (Implied)Reported Device Performance
    Sterility: Achieve a 10-6 Sterility Assurance Level (SAL) via radiation sterilization. Package integrity maintained.Sterility: The device is sterilized with radiation, and the minimum sterilizing dose (MSD) for a 10-6 SAL was established and validated as described in ANSI/AAMI/ISO ISO 11137-2. Package Verification testing (Visual Inspection, Seal Strength, Bubble Leak testing) supports sterility.
    Shelf Life: Maintain stability and performance over time, ideally matching or exceeding predicate/reference devices.Shelf Life: A shelf life claim of 5 years is substantiated by stability results of the reference device ACTIFUSE SHAPE (K082575), as stability-indicating parameters are identical due to shared materials and packaging.
    Performance - Bench: Demonstrate bioactive and osteoconductive properties, and similar physical structure/porosity to predicate.Performance - Bench: No additional bench testing was conducted as the microgranule phase has identical chemical composition, physical structure, and manufacturing process to the predicate (K181225), and the carrier phase is identical to the reference device (K082575). Previous verification/validation testing for K181225 and K082575 are considered applicable. In vitro studies from the predicate submission evaluated bioactive properties (forming a surface apatite-layer in simulated body fluid).
    Performance - Animal (Effectiveness): Demonstrate effectiveness for use in posterolateral fusion and extremities.Performance - Animal: A preclinical animal study evaluated effectiveness for use in posterolateral fusion (with autogenous iliac crest bone graft (ICBG)). Another preclinical animal study assessed effectiveness for use in extremities (with and without autologous blood). The text states that "ALTAPORE SHAPE is osteoconductive based on in vivo animal studies that show it achieves bone healing in a critical defect model as confirmed with radiographic, histolopathological, histomorphometric, and mechanical analyses. ALTAPORE SHAPE undergoes cell-mediated remodeling and is replaced by natural bone."
    Biocompatibility: Meet ISO-10993-1 and FDA guidance recommendations for biological safety.Biocompatibility: Assessments were conducted based on ISO-10993-1 and FDA guidance. Testing included Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Pyrogen, Genotoxicity, and Implantation. The components of Altapore Shape (hydroxyapatite microgranule and alkylene oxide copolymer carrier) have been used in previously cleared predicate/reference devices with the same intended use and contact duration.
    Substantial Equivalence: All non-clinical data demonstrates the device is substantially equivalent to predicates.Substantial Equivalence: The non-clinical data demonstrates that the subject device is substantially equivalent and performs comparably to the predicate and reference devices for the same intended use.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document refers to preclinical animal studies for performance testing. However, it does not provide details on the sample size for these animal studies.
    The provenance of the data (e.g., country of origin) is also not specified. The studies appear to be prospective in nature, as they were conducted to evaluate the device's performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. The animal studies involved radiographic, histopathological, histomorphometric, and mechanical analyses, which would typically involve expert interpretation. However, the number and qualifications of these experts are not detailed.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not describe any adjudication methods for the test set. Animal study results would typically be analyzed by the study researchers.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This document describes the clearance of a medical implant (bone void filler), not an AI-powered diagnostic or assistive device. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. As stated above, this is an implantable medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the animal performance studies, the "ground truth" (or evidence of performance) was established through:

    • Radiographic analyses
    • Histolopathological analyses
    • Histomorphometric analyses
    • Mechanical analyses

    These are objective measures of bone healing and integration in animal models.

    For biocompatibility, the ground truth was based on the results of the specified in vitro and in vivo tests (Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Pyrogen, Genotoxicity, Implantation) as per ISO-10993-1 and FDA guidance.

    8. The sample size for the training set

    The document does not mention a "training set" as it pertains to AI/ML development. This is a traditional medical device submission based on substantial equivalence to existing devices and nonclinical testing.

    9. How the ground truth for the training set was established

    Not applicable, as there is no mention of an AI/ML training set in the document.