Search Results
Found 2 results
510(k) Data Aggregation
(30 days)
ACUSON Sequoia Diagnostic Ultrasound System, ACUSON SC2000 Diagnostic Ultrasound System, ACUSON Freestyle
Diagnostic Ultrasound System,ACUSON S1000, S2000, S3000 Diagnostic Ultrasound Systems, ACUSON P200 Diagnostic
The ACUSON Sequoia ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Fetal, Abdominal, Pediatric, Neonatal Cephalic, Small Parts, OB/GYN (useful for visualization of the ovaries, follicles, uterus and other pelvic structures), Cardiac, Pelvic, Vascular, Musculoskeletal and Peripheral Vascular applications.
The system also provides the ability to measure anatomical structures for fetal, abdominal, pediatric, small organ, cardiac, transrectal, transvaginal, peripheral vessel, musculoskeletal and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transesophageal, Adult Cephalic, Peripheral Vessel, Abdominal, Intraoperative Abdominal, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
The typical examinations performed using the SC2000 Ultrasound System are: Cardiac Imaging Applications and Analysis, Vascular Imaging Applications and Analysis, Superficial Imaging Applications, Intraoperative Imaging Applications, Transcranial Imaging Applications.
The ACUSON Freestyle Ultrasound System is intended for diagnostic imaging or fluid flow analysis of the human body performed by an appropriately wained healthcare professional in a healthcare setting for the following conditions: Abdominal, Pediatric, Small Organ, Peripheral Vessel, Musculoskeletal (Conventional), Musculoskeletal (Superficial).
The ultrasound imaging systems are intended for the following applications: Fetal. Abdominal. Intraoperative, Pediatric. Small Parts, Transcranial, OB/GYN (including monitoring of the ovarian follicle development), Cardiac. Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
The system also provides the ability to measure anatomical structures {fetal, abdominal, intraoperative, small organ, neonatal cephalic, adult cephalic, trans-esophageal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculo-skeletal (superficial) and neonatal cardiac} and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement: Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging".
The ACUSON AcuNav Ultrasound Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients.
The ACUSON P200 ultrasound imaging system is intended for the following applications: Fetal, Abdominal, Pediatric, Small Parts, OB/GYN (useful for visualization of the ovaries, follicles, uterus, and other pelvic structures), Adult, Pediatric and Neonatal Cardiac, Pelvic, Neonatal Cephalic, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
The system also provides the ability to measure anatomical structures; fetal, abdominal, pediatric, small organ, neonatal cephalic, cardiac (adult, pediatric and neonatal), trans-esophageal, transrectal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculosketal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
The ACUSON P500 ultrasound imaging system is intended for the following applications: Fetal, Abdominal (including live), Pediatric, Small Parts, Transcranial, OB/GYN(useful for visualization of the ovaries, follicles, uterus and other pelvic structures), Pelvic. Neonatal, Cardiac, Vascular(including Peripheral Vessel), Musculoskeletal, Superficial Musculoskeletal, and Urology applications.
The system also provides the ability to measure anatomical structures fetal, abdominal, small organ, transrectal, transvaginal, cardiovascular, peripheral vessel, musculoskeletal (conventional), and musculoskeletal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system.
This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Society of Echocardiography Carotid Intima-Media Thickness Task Force. Endorsed by the Society for Vascular Imaging."
The ACUSON Acunav Ultrasound Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other of adult and pediatric patients.
For ACUSON NX3: The ACUSON NX3 ultrasound imaging system is intended for the following applications: Fetal Abdominal (including liver), Pediatric, Small Parts (Small Organ), Neonatal Cephalic, Transcranial, OB/GYN, Pelvic, Neonatal, Cardiac, Vascular (including Peripheral Vessel), Musculoskeletal and Urology applications. The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
For ACUSON NX3 Elite: The ACUSON NX3 Elite ultrasound imaging system is intended for the following applications: Fetal, Abdominal (including liver, intra-operative), Pediatric, Small Parts (Small Organ including intra-operative), Neonatal Cephalic, Adult Cephalic, Transcranial, OB/GYN, Pelvic, Neonatal, Cardiac(including Transesophageal), Vascular (including Peripheral Vessel, intra-operative), Musculoskeletal, Superficial Musculoskeletal and Urology applications. The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system.
Note: This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Society of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Medicine."
The ACUSON Sequoia Diagnostic Ultrasound System is a multi-purpose mobile, software controlled, diagnostic ultrasound system with an on-screen display of thermal and mechanical indices related to potential bio-effect mechanisms. Its function is to transmit and receive ultrasound echo data and display it in B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Color M Mode, Doppler Tissue Mode, Amplitude Doppler Mode, a combination of modes, Panoramic Imaging, Contrast agent Imaging, Virtual Touch Strain Imaging, Virtual Touch - pSWE Imaging, Virtual Touch - SWE Imaging, syngo Velocity Vector Imaging, Custom Tissue Imaging, 3D/4D Volume Imaging and Harmonic Imaging on a Display.
All of the transducers and the catheter-based transducers will follow Track 3 acoustic labeling (AIUM 1004, IEC 2007, AIUM/NEMA 2004a) and remain unchanged from the currently cleared ACUSON systems.
Note: Some ACUSON Diagnostic Ultrasound configurations of components, accessories and/or software may vary from device brands and within device families. There are no modifications, additions or labeling changes to components, accessories and/or software for each device indicated.
The ACUSON Diagnostic Ultrasound systems are intended to be used by trained medical professionals at various locations where patient care, evaluation, monitoring or research is performed.
The provided document is a 510(k) Pre-market Notification for several Siemens ACUSON Diagnostic Ultrasound Systems, seeking clearance based on substantial equivalence to previously marketed devices. The document explicitly states that no new features, modifications, or changes were made to the hardware or software of the devices themselves. The entire submission focuses on an "expanded labeling" for lung ultrasound imaging, supported by existing clinical literature.
Therefore, the acceptance criteria and the study proving the device meets these criteria are not related to a new device or a new AI/algorithm feature being validated. Instead, it is about demonstrating that the existing, cleared devices can be marketed with expanded indications for use, supported by references to previously conducted tests and existing clinical literature.
Given this, I cannot provide information on:
- Reported device performance (as no new performance was claimed or tested)
- Sample size used for the test set or data provenance
- Number of experts used to establish ground truth or their qualifications
- Adjudication method
- MRMC comparative effectiveness study or effect size
- Standalone algorithm performance
- Sample size for training set or how ground truth for training set was established
The document explicitly states: "clinical studies were not required to support substantial equivalence. The expanded Labeling for Lung Ultrasound imaging (LUS) is supported by clinical literature." and "There are no software or hardware modifications to any device under this review and the expanded labeling has no impact on the individual ACUSON ultrasound device, therefore, testing results are not included in this Special 510(k) submission."
Below is a table summarizing the acceptance criteria and "device performance" in the context of this 510(k) submission, emphasizing that the "performance" here refers to the device meeting the previously established safety and effectiveness for its predicate indications, which is being extended to a new recognized clinical use case via labeling.
Description of Acceptance Criteria and Evidence for ACUSON Diagnostic Ultrasound Systems (K202683)
This 510(k) submission (K202683) by Siemens Medical Solutions, USA, Inc., is a Special 510(k). This type of submission is used when changes to a cleared device do not alter its fundamental scientific technology or intended use, but may expand its labeling or functionality within its established safety and effectiveness profile. In this specific case, the submission focuses on expanded labeling for Lung Ultrasound Imaging (LUS), utilizing the existing capabilities of the cleared ACUSON Diagnostic Ultrasound Systems and their associated transducers.
Crucially, the document states that no modifications were made to the software or hardware of the devices. Therefore, the "acceptance criteria" and "proof" are based on demonstrating substantial equivalence to previously cleared predicate devices and leveraging existing validation data for those predicates, rather than presenting new performance metrics for a novel algorithm or device.
1. Acceptance Criteria and Reported Device "Performance" (Meeting Substantial Equivalence)
Acceptance Criteria (from 510(k) perspective) | Reported Device "Performance" / Justification for Meeting Criteria |
---|---|
Identical Intended Use/Indications for Use: The devices, with expanded labeling, align with the fundamental intended use of previously cleared predicate devices. | The Indications for Use statements for all listed ACUSON systems (Sequoia, SC2000, Freestyle, S1000/S2000/S3000, P200, P500, NX3/NX3 Elite) are presented as identical to their respective primary predicate devices, with the exception of the added use case for lung imaging explicitly mentioned in the "Indications for Use" section descriptions (e.g., "Abdominal (including liver, intra-operative)" for NX3 Elite). The core functionality and diagnostic purpose remain the same. The submission argues that lung ultrasound imaging falls within the scope of general diagnostic ultrasound. |
Identical Technological Characteristics: The device hardware, software, and operating principles are unchanged from their predicate versions, ensuring safety and effectiveness. | The document explicitly states: "the subject devices are substantially equivalent to the cleared primary predicate devices and employs the same fundamental scientific technology and intended use/indications for use." |
Detailed comparison tables (Table 2 in the original document, not fully reproduced here due to length) consistently list "Identical" for all compared technological characteristics: Reusability, Duration of Use, Scientific Technology, Operating Principles, Type of Previously Cleared Transducers, Acoustic Outputs, Previously Cleared Imaging Modes, and Biocompatibility (ISO 10993-1). |
| Safety and Effectiveness: There are no new safety or effectiveness concerns introduced by the expanded labeling. | - Software Verification and Validation: The CRS (Customer Requirement Specifications) for each device were verified/validated during original development. No software/hardware modifications for this submission.
- Standards Compliance: Conforms to IEC 62304, IEC 62366-1, NEMA PS 3.1 - 3.20 (DICOM).
- Risk Management: Risk analysis (ISO 14971:2007) was performed on predicate devices, and mitigation controls were implemented. The expanded labeling does not impact these.
- Cybersecurity: Cybersecurity measures are addressed and conform to requirements to prevent unauthorized access, modification, misuse, or denial of use.
- Clinical Literature Support: The expanded labeling for Lung Ultrasound Imaging (LUS) is supported by clinical literature, indicating that the use of ultrasound for lung imaging is an established and safe practice. Clinical studies were not required for this 510(k) as the technology and operating principles are unchanged. |
2. Sample Size and Data Provenance for Test Set
- Test Set Description: Not applicable. For this Special 510(k), no new test set data for device performance was generated or presented. The submission relies on existing validation data for the predicate devices.
- Data Provenance: The underlying data for previously cleared predicate devices would have been generated through rigorous internal testing and validation processes during their initial development and clearance. The document implies that this data exists and supports the current device. The document explicitly states: "Non-clinical Testing has been conducted during product development for each of the predicate ACUSON Diagnostic Ultrasound device described in this submission."
3. Number, Qualifications, and Adjudication Method for Experts
- No new expert assessment: Not applicable for this Special 510(k). Since no new AI/algorithm features or changes to device performance were introduced, there was no independent expert review of a new test set for the purpose of this submission. The "ground truth" for the expanded labeling (Lung Ultrasound) is based on established clinical recognition and existing peer-reviewed literature.
- Adjudication method: Not applicable.
4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No new MRMC study: Not applicable. This submission does not involve a new AI or algorithm that would require an MRMC study to demonstrate human improvement with AI assistance.
5. Standalone (Algorithm Only) Performance Study
- No standalone algorithm: Not applicable. The submission is for diagnostic ultrasound systems, not a standalone algorithm. The core functionality is the ultrasound imaging, and the "expansion" is in the labeling for an existing modality.
6. Type of Ground Truth Used for Device Functions
- Existing Validated Performance: For the predicate devices, the ground truth and validation would have been established through a combination of:
- Direct measurement and engineering specifications: For image quality, acoustic output, and system performance.
- Clinical validation: Through various means including phantom studies, retrospective/prospective clinical data, and expert consensus for the initial indications.
- Clinical Literature/Consensus for Expanded Indication: For the new labeling related to Lung Ultrasound Imaging (LUS), the "ground truth" or justification is based on:
- Established clinical practice: Recognition within the medical community for the utility of ultrasound in lung assessment.
- Peer-reviewed studies and professional guidelines: As explicitly stated, the LUS labeling is "supported by clinical literature." This effectively implies that the medical community's consensus on the utility of LUS (as documented in such literature) serves as the "ground truth" for its clinical relevance.
7. Sample Size for Training Set
- No new training set: Not applicable. As no new AI/algorithm components were introduced or modified, there was no "training set" for the purpose of this specific 510(k) submission.
8. How Ground Truth for Training Set was Established
- No new training set: Not applicable.
Ask a specific question about this device
(30 days)
ACUSON Freestyle Diagnostic Ultrasound System
The ACUSON Freestyle Ultrasound System is intended for diagnostic imaging or fluid flow analysis of the human body performed by an appropriately trained healthcare professional in a healthcare setting for the following conditions: Abdominal, Pediatric, Small Organ, Peripheral Vessel, Musculoskeletal (Conventional), Musculoskeletal (Superficial).
The ACUSON Freestyle™ Diagnostic Ultrasound System is a portable ultrasound imaging system where the system operates with linear and curvilinear array transducers. These transducers may be used in a wireless mode or using a probe adaptor cable and is capable of the following modes (these modes below can be operated in combination or individually):
- । B-Mode
- -Color Doppler Mode
- -Amplitude Doppler Mode
The ACUSON Freestyle™ Diagnostic Ultrasound System includes a main unit console with user interface controls, a 15-inch video display, and system electronics. The main unit weighs approximately 10.5 pounds and is 13.2 inches high, 14.7 inches wide, and 5 inches deep. The system may be mounted on a small roll stand, monitor arm, or tabletop. An optional external receiving antenna can be mounted to the unit console, this provides better line of sight to the wireless transducers. The system provides an intuitive and easy-to operate user interface. It has a compact and portable design and offers DICOM archival and networking capability.
The provided text describes the ACUSON Freestyle Diagnostic Ultrasound System (VA41A), a modification of a previously cleared device (ACUSON Freestyle 4.0, VA40A). The document focuses on demonstrating substantial equivalence to the predicate device, rather than providing a detailed study proving the device meets specific acceptance criteria with quantitative performance metrics.
However, based on the information provided, we can infer the acceptance criteria relate to maintaining equivalence with the predicate device and adhering to established medical device standards. The "study" proving acceptance criteria is a substantial equivalence determination based on comparison and non-clinical testing.
Here's an breakdown based on your request, as much as can be extracted from the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria here are implicitly stated as being substantially equivalent to the predicate device in terms of:
- Intended Use and Indications for Use
- Technological Characteristics
- Safety and Effectiveness
The document does not provide quantitative performance metrics against specific, pre-defined acceptance criteria (e.g., minimum sensitivity, specificity, or image resolution values). Instead, it asserts that the new device, VA41A, maintains the performance and safety profile of the predicate device, VA40A, which was already cleared.
Acceptance Criteria (Inferred from Substantial Equivalence Claim) | Reported Device Performance (Summary of Comparison) |
---|---|
Intended Use | Identical: The device maintains the same intended use: "diagnostic imaging or fluid flow analysis of the human body performed by an appropriately trained healthcare professional in a healthcare setting." |
Indications for Use | Modified but Substantially Equivalent: Existing indications (Abdominal, Pediatric, Small Organ, Peripheral Vessel, Musculoskeletal Conventional, Musculoskeletal Superficial) are identical. Some indications (Fetal, Cardiac, Neonatal Cephalic, Intraoperative Neurological) were deleted. The addition of the L17-5 transducer expands capabilities for existing indications but does not introduce new types of clinical applications. |
Technological Characteristics | Largely Identical with Minor Modifications/Enhancements: |
- Imaging modes (B-Mode, Color Doppler, Amplitude Doppler, M-mode, PWD) are identical.
- Transducer frequencies are identical for existing transducers (L8-3, L13-5, C5-2).
- New Transducer: Introduction of L17-5 (5.0 - 17.0 MHz) with higher frequency for improved near-field resolution.
- Imaging functions (Multi-Hertz, Beam-free, Spatial Compounding, Speckle Filter, TGC, Auto Send, Auto Study Management, Needle V, Measurements, Trapezoidal Imaging, Cine Capture) are identical, with new Dual Mode and Split Mode.
- Wireless features, FCC compliance, battery operation, connectivity, monitoring, OS, output display, patient contact materials, and UL certification are identical. |
| Safety and Effectiveness | Confirmed by Compliance: Evaluated for acoustic output, cleaning, disinfection, thermal, electromagnetic, and mechanical safety. Found to conform to applicable medical device safety standards (AIUM/NEMA UD-2, UD-3, ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62359, IEC 60601-1-6, IEC 62304, ISO 10993-1, ISO 14971). No new issues of safety or effectiveness are raised. |
2. Sample Size Used for the Test Set and Data Provenance
The document explicitly states: "Since the ACUSON Freestyle Diagnostic Ultrasound System is a class II device and uses the same technology and operating principles as existing predicate device, ACUSON Freestyle Ultrasound System (K162417), therefore clinical studies were not required to support substantial equivalence."
This indicates that there was no dedicated clinical test set for evaluating the performance of the VA41A in human subjects. The assessment relies on a comparison to the predicate device and non-clinical engineering and safety testing. Therefore, details regarding sample size, country of origin, or retrospective/prospective nature of a clinical test set are not applicable here because such a study was not performed.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
As there was no clinical test set requiring ground truth establishment from human readings, this information is not applicable.
4. Adjudication Method for the Test Set
As there was no clinical test set requiring ground truth establishment from human readings, this information is not applicable.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
The device described is a diagnostic ultrasound system, not an AI-powered diagnostic tool. Therefore, an MRMC comparative effectiveness study comparing human readers with and without AI assistance is not applicable.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This refers to an AI algorithm's performance in isolation. As the device is an ultrasound system and not an AI algorithm for image interpretation, this is not applicable. The "device" itself is the imaging system, and its performance is evaluated through engineering and safety standards, and equivalence to a predicate, not as an AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the non-clinical testing and comparison to standards, the "ground truth" would be established by:
- Engineering specifications and measurements: For acoustic output, thermal properties, electromagnetic compatibility, and mechanical safety.
- Standard compliance: Adherence to recognized national and international standards (e.g., AIUM/NEMA, ANSI AAMI, IEC, ISO).
- Predicate device characteristics: The established performance and safety profile of the legally marketed predicate device (ACUSON Freestyle 4.0, VA40A).
8. The Sample Size for the Training Set
As this is not an AI/ML device, the concept of a "training set" is not applicable.
9. How the Ground Truth for the Training Set was Established
As this is not an AI/ML device, the concept of a "training set" and its ground truth establishment is not applicable.
Ask a specific question about this device
Page 1 of 1