Search Results
Found 1 results
510(k) Data Aggregation
(203 days)
ACTIVE ANTERIOR CERVICAL PLATE SYSTEM
The Active "10 Anterior Cervical Plate System is intended for the treatment of the cervical spine in skeletally mature patients receiving fusion by autogenous and/or allogenic bone graft. The implants are attached to the anterior cervical spine (C2-11) with removal of the implants after the attainment of a solid fusion mass. The Active ™ Anterior Cervical Plate System is intended for use under the following indications:
· Degenerative disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
- · Spondylolisthesis
- · Trauma (i.e., fracture or dislocation)
- · Spinal stenosis
- · Deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis)
• Tumor
• Pseudoarthrosis
- · Failed previous fusion
The Active " Anterior Cervical Plate System is a fixation system designed to allow for settling of the bone graft during fusion. Plates are provided in various lengths to accommodate fusions from 1 to 4 Ievels. Fixed and variable angle locking unicortical screws are provided in lengths of 12, 14, 16, and 18mm and diameters of 4mm and 4.5mm.
The provided text describes the 510(k) summary for the "Active Anterior Cervical Plate System" (K073500), a medical device intended for spinal fixation. This submission focuses on establishing substantial equivalence to previously marketed predicate devices, primarily through mechanical performance testing and comparison of technological characteristics.
Here's an analysis based on your requested information:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
Mechanical performance attributes per ASTM F1717 | Found to be substantially equivalent to those of predicate devices. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify a separate "test set" in the context of clinical data or patient samples. The evaluation of this device is based on mechanical performance attributes per ASTM F1717. For such testing, the "sample size" would refer to the number of devices or components tested. This information is not provided in the given text.
Similarly, the data provenance (country of origin, retrospective/prospective) is not applicable as the study described is a mechanical engineering test, not a clinical study involving human subjects or their data.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable. The "ground truth" for mechanical testing is typically established by the specifications of the ASTM F1717 standard and the results of the physical tests themselves. It does not involve human expert consensus in the way clinical studies do.
4. Adjudication Method for the Test Set
This is not applicable for mechanical performance testing. Adjudication methods (like 2+1 or 3+1) are used in clinical studies where human interpretation or consensus is required to establish ground truth or resolve discrepancies in evaluations.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Readers Improvement with AI vs. Without AI Assistance
This information is not applicable. The submission describes a spinal fixation system, which is a physical implant, not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study comparing human reader performance with and without AI assistance is irrelevant to this device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This is not applicable. The device is a physical implant, not a software algorithm.
7. The Type of Ground Truth Used
The "ground truth" for this device's evaluation is primarily mechanical performance standards (ASTM F1717). The device's performance was compared to existing predicate devices based on these standards.
8. The Sample Size for the Training Set
This information is not applicable as the evaluation is based on mechanical performance testing, not a machine learning model that requires a "training set."
9. How the Ground Truth for the Training Set Was Established
This is not applicable for the same reasons as #8.
Summary of the Study and Device Performance:
The study described for the Active Anterior Cervical Plate System is largely a technical and mechanical performance evaluation to demonstrate substantial equivalence to legally marketed predicate devices. The primary "study" involved:
- Comparison of Technological Characteristics: The device's features (augmenting stability, segmented plate for screw translation, use of titanium alloys, compression application) were compared and found to be substantially equivalent to features present in various predicate devices.
- Mechanical Testing per ASTM F1717: The device underwent mechanical performance testing according to the ASTM F1717 standard. The results of this testing demonstrated that its mechanical attributes were "substantially equivalent" to those of the predicate devices.
The FDA's review concluded that the device is substantially equivalent to predicate devices for its stated indications for use, primarily based on these comparative analyses and mechanical testing. No clinical studies with human participants, expert consensus for diagnosis, or AI performance evaluations are mentioned in this 510(k) summary.
Ask a specific question about this device
Page 1 of 1