Search Results
Found 1 results
510(k) Data Aggregation
(287 days)
ACESSA GUIDANCE HANDPIECE
The Acessa Guidance Handpiece is indicated for use in percutaneous laparoscopic coagulation and ablation of soft tissue, including treatment of symptomatic uterine fibroids under laparoscopic ultrasound guidance.
Acessa Guidance Handpiece (Model Number 5300):
The Acessa Guidance Handpiece, when used with the tracking system, is meant to guide the tip of the Handpiece up to the uterine serosa. Once the device is advanced into the uterus, standard ultrasound views are used to guide the tip to the fibroid. Ultrasound visualization for fibroid penetration and treatment remain unchanged.
The Acessa Guidance Handpiece is a single-use only, sterile, radiofrequency (RF) applicator that connects to the Acessa Generator by a custom Cable. The Handpiece consists of a handle with a trocar shaft, active needle tip, and seven deployable needle electrodes each with embedded thermocouple sensors for real-time temperature feedback. The Acessa Guidance Handpiece also has control buttons that allow the user to access the User Interface.
This Acessa Guidance Handpiece also contains the magnetic guidance sensor used to determine spatial location. The embedded sensor and SROM can be used with the electromagnetic tracking technologies to track the positions of an Ultrasound Transducer and the Handpiece shaft, and draws virtual representations of them in their spatial relationship, so that a physician can predict the Handpiece shaft's future path in relation to the features in the ultrasound slice. These electromagnetic tracking technologies are considered optional accessories to Acessa System procedure.
Here's a summary of the acceptance criteria and study information for the Acessa Guidance Handpiece based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Test Category | Acceptance Criteria | Reported Device Performance |
---|---|---|
Electrical and Safety Testing | Compliance with EN/IEC 60601-1:2005/2006 (3rd edition) Medical Electrical Equipment Part 1 General Requirements for Safety; EN/IEC 60601-1-2:2007 (3rd edition) Medical Electrical Equipment General Requirements for Basic Safety and Essential Performance Collateral Standard Electromagnetic Compatibility Requirements and Tests; EN/IEC 60601-2-2:2009 (5th edition) Medical Electrical Equipment Particular Requirements for the Safety of High Frequency Surgical Equipment. | Passed all testing. |
Biocompatibility Testing | Compliance with ISO 10993-1:2009 (4th edition) Biological Evaluation of Medical Devices Part 1 Evaluation and Testing within a Risk Management Process; ISO 10993-5:2009 (3rd edition) Biological Evaluation of Medical Devices Tests for In Vitro Cytotoxicity; ISO 10993-10:2010 (2nd edition) Biological Evaluation of Medical Devices Tests for Irritation and Delayed-Type Hypersensitivity; and ISO 10993-11:2006 (2nd edition) Biological Evaluation of Medical Devices Tests for Systemic Toxicity. | Demonstrated safety, suitability, and appropriateness for intended use. |
Flexural Strength Testing | Not explicitly stated in terms of specific numeric criteria, but evaluated against established criteria. | Met all criteria. |
Compatibility with EM Tracking | Not explicitly stated in terms of specific numeric criteria, but evaluated against established criteria for functionality with an electromagnetic tracking system. | Met all criteria. |
2. Sample Size Used for the Test Set and Data Provenance
The provided document describes performance testing of the device itself (electrical, safety, biocompatibility, mechanical, and compatibility with a tracking system), not a clinical study on patient data. Therefore, there isn't a "test set" in the context of patient data, nor is there data provenance in terms of country of origin or retrospective/prospective nature for such a clinical study. The testing was technical in nature, conducted on the device hardware and materials.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. As noted above, this was not a study evaluating performance against a clinical ground truth established by experts.
4. Adjudication Method for the Test Set
Not applicable. This was not a study involving expert adjudication of clinical cases.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document describes the 510(k) submission for a medical device (Acessa Guidance Handpiece) which is a physical instrument, not an AI or software algorithm. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant here.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is not an algorithm or AI device.
7. The Type of Ground Truth Used
The "ground truth" for this device's performance testing was based on:
- Established national and international standards for electrical safety (EN/IEC 60601 series).
- Established national and international standards for biocompatibility (ISO 10993 series).
- Internal criteria for flexural strength and compatibility with the electromagnetic tracking system.
8. The Sample Size for the Training Set
Not applicable. This is not a machine learning or AI device that requires a training set.
9. How the Ground Truth for the Training Set was Established
Not applicable.
Ask a specific question about this device
Page 1 of 1