Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K090943
    Date Cleared
    2009-05-27

    (54 days)

    Product Code
    Regulation Number
    862.1660
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro diagnostic use (i.e. for external use only) by healthcare professionals and in the home by people with diabetes mellitus to assess the performance of the Agamatrix WaveSense Presto Blood Glucose Monitor.

    Device Description

    The AbT Glucose Control Solution consists of a viscosity-adjusted, aqueous liquid control solution containing a known quantity of glucose. It is packaged in a plastic dropper tipped bottle for easy application of the control solution to the test strips and a red coloration to aid the user to visually confirm application of the control. The product is nonhazardous and contains no human or animal derived materials.

    AI/ML Overview

    The AbT Glucose Control Solution is a control material for blood glucose monitoring systems. The provided documentation does not detail a study involving human readers or comparative effectiveness for diagnostic image analysis. Instead, it describes a device for quality control in glucose measurement. The performance studies focused on the stability and precision of the control solution itself, not on diagnostic accuracy in the way an AI-powered image analysis tool would be evaluated.

    Here's an analysis based on the provided text, reformatted to address your specific questions about acceptance criteria and study details, albeit adapted for this type of device:


    1. Table of Acceptance Criteria and Reported Device Performance

    For the AbT Glucose Control Solution, the "acceptance criteria" are defined by its target glucose range and its ability to maintain stability and precision within that range. The "reported device performance" reflects how these characteristics compare to predicate devices.

    CharacteristicAcceptance Criteria (Predicate Devices - Target Ranges)Reported Device Performance (New Product - Target Range)
    Glucose Target RangeWaveSense Normal Control Solution: 108 - 159AbT Glucose Control Solution: 95 - 145
    Liberty Normal Control Solution: 94 - 147
    StabilityImplied comparable stability to predicate devicesFound to be acceptable (see "Performance Studies")
    Open Vial StabilityImplied comparable open vial stabilityFound to be acceptable (see "Performance Studies")
    Test PrecisionImplied comparable precision to predicate devicesFound to be acceptable (see "Performance Studies")

    Note: The specific quantitative results for stability and precision are not provided in this summary but are indicated as having been performed and concluded to be satisfactory.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of diagnostic performance on patient data since this device is a quality control solution.

    • Sample Size: Not applicable in the traditional sense of patient data. The "samples" would be aliquots of the control solution itself. The document does not provide the number of replicates or batches tested for stability and precision studies.
    • Data Provenance: Not applicable in the context of patient data. The studies would have been conducted internally by American Biological Technologies, Inc. No country of origin for patient data is relevant here. The studies are prospective in nature, testing the manufactured control solution.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable. "Ground truth" for a control solution relates to its manufactured glucose concentration and its stability and precision characteristics, not diagnostic interpretations by experts. The "ground truth" would be established through analytical chemistry methods during production and quality control testing.

    4. Adjudication Method

    Not applicable. Adjudication methods are relevant for expert review of diagnostic cases, which is not the purpose of this device.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This device is a quality control solution, not a diagnostic tool intended for interpretation by human readers. Therefore, an MRMC study is not relevant.

    6. Standalone Performance

    Yes, in a sense. The "performance studies" (stability, open vial stability, test precision) are evaluations of the AbT Glucose Control Solution itself and its inherent characteristics. This is a form of standalone performance assessment, focusing on the quality and reliability of the control solution as a product. The results of these studies informed the claim of substantial equivalence to predicate devices.

    7. Type of Ground Truth Used

    The "ground truth" for the AbT Glucose Control Solution is based on analytical manufacturing specifications and laboratory testing.

    • Glucose Concentration: The target range (95 - 145) is a defined analytical specification for the product.
    • Stability: This is determined by monitoring the glucose concentration over time under various storage conditions.
    • Precision: This is determined by repeatedly measuring the glucose concentration and analyzing the variability.

    8. Sample Size for the Training Set

    Not applicable. This device is not an AI algorithm that requires a "training set" of data. Its "performance" is based on its physical and chemical properties as a control solution.

    9. How Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for an AI algorithm, there is no ground truth to establish in this context.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1