Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K992594
    Manufacturer
    Date Cleared
    1999-09-24

    (52 days)

    Product Code
    Regulation Number
    876.5820
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    A-15 HEMODIALYZER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Hemodialysis with Althin capillary dialyzers are indicated for patients with acute or chronic renal failure when conservative therapy is judged to be inadequate.

    Device Description

    The A-15 Hemodialyzer is a conventional hemodialyzer that is used as an artificial kidney in a hemodialysis system to treat patients with acute or chronic renal failure. The blood from a patient flows through the arterial tubing of the extracorporeal blood circuit to the blood compartment of the hemodialyzer, then returns through the venous tubing of the extracorporeal blood system to the patient. The hemodialyzer has semipermeable hollow fibers which divide the device into two compartments. When the blood passes through the hollow fibers, water and toxic waste products from the blood pass through the semipermeable membrane into the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the hemodialyzer. Each A-15 Hemodialyzer is packaged in a plastic bag and 20 hemodialyzers are packed in a cardboard box. It is sterilized by gamma radiation and intended, and labeled, for single use only.

    AI/ML Overview

    The provided text is a 510(k) summary for the Althin Medical AB A-15 Hemodialyzer, which is a medical device. This document does not describe a study involving performance metrics like accuracy, sensitivity, or specificity, which are typically associated with AI/ML model evaluations. Instead, it describes a
    The acceptance criteria are that the device perform equivalently to the predicate device. The study performed that demonstrates that the A-15 Hemodialyzer meets this criteria is functional testing. The functional testing was based on the specifications of the predicate device, the Altra Nova 140 Hemodialyzer (K945625). The results of the functional testing showed "significant equivalence." Additionally, biocompatibility testing was performed in accordance with ISO 10993, and the device passed all indicated tests.

    Here's a breakdown of the requested information based on the provided text. Please note that many of the categories you requested are typically relevant for AI/ML device evaluations and are not applicable to the traditional medical device submission described here.

    1. Table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Functional performance equivalent to predicate device (Altra Nova 140 Hemodialyzer, K945625)Functional testing showed "significant equivalence" to the predicate device's specifications. The device conforms to its specifications and is suitable for its intended use.
    BiocompatibilityPassed the panel of tests indicated for a hemodialyzer device in accordance with ISO 10993.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size for test set: Not specified in the document. The testing involved "functional testing" and "biocompatibility testing," but the number of units tested is not provided.
    • Data provenance: Not explicitly stated. The company, Althin Medical AB, is based in Sweden. The testing would have been conducted to support the 510(k) submission to the FDA in the USA.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Number of experts: Not applicable. The "ground truth" for this type of device is based on established engineering specifications and industry standards for functional and biocompatibility performance, not expert medical opinion on individual cases.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication method: Not applicable. This does not involve interpretation of data by multiple experts.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC study: No. This is a traditional hemodialyzer, not an AI/ML device, so such a study is not relevant or described.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone performance: Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Type of ground truth: The ground truth for this device's performance is based on established engineering specifications for hemodialyzers (e.g., ultrafiltration coefficient, clearance) and the requirements of biocompatibility standards (ISO 10993). Performance is compared against the specifications of a legally marketed predicate device.

    8. The sample size for the training set

    • Sample size for training set: Not applicable. This is not a machine learning device and therefore does not have a "training set" in the context of AI/ML.

    9. How the ground truth for the training set was established

    • Ground truth for training set: Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1