Search Results
Found 1 results
510(k) Data Aggregation
(171 days)
8MP COLOR LCD MONITOR, REDIFORCE RX840
The RadiForce RX840-MG is intended to be used in displaying and viewing digital images including those of digital mammography for review and analysis by trained medical practitioners.
The RadiForce RX840-MG is a color LCD monitor for viewing medical images including those of mammography. The color panel employs in-plane switching (IPS) technology allowing wide viewing angles. With the matrix size (or resolution) of 4.096 x 2.160 pixels (8MP), the RX840-MG is an alternate replacement for traditional dual head 5MP display installations.
Here's a breakdown of the acceptance criteria and study information for the EIZO RadiForce RX840-MG, based on the provided document:
Acceptance Criteria and Reported Device Performance
Acceptance Criteria (from TG18 guideline & pre-defined) | Reported Device Performance (RadiForce RX840-MG) |
---|---|
Conformance to DICOM GSDF | Met criteria (Test results showed conformance) |
Angular dependency of luminance response (horizontal, vertical, diagonal) | Measured and found to be equivalent to predicate device, with differences not affecting observer performance. |
Luminance non-uniformity characteristics | Measured and found to be equivalent to predicate device, with differences not affecting observer performance. |
Chromaticity non-uniformity characteristics | Measured and found to be equivalent to predicate device, with differences not affecting observer performance. |
Chromaticity at center of display (5%, 50%, 95% max luminance) | Measured. Specific results not detailed but implied to meet criteria or not affect observer performance. |
Display reflections (specular, diffuse, haze) | Measured. Specific results not detailed but implied to meet criteria or not affect observer performance. |
Small-spot contrast ratio | Measured. Specific results not detailed but implied to meet criteria or not affect observer performance. |
Spatial resolution (Modulation Transfer Function - MTF) | Measured. Specific results not detailed but implied to meet criteria or not affect observer performance. |
Noise (Noise Power Spectrum - NPS) | Measured. Specific results not detailed but implied to meet criteria or not affect observer performance. |
Pixel aperture ratio | Measured. Specific results not detailed but implied to meet criteria or not affect observer performance. |
Absence of miscellaneous artifacts | Visually checked and met criteria (No mention of artifacts affecting performance). |
Temporal response | Measured. Specific results not detailed but implied to meet criteria or not affect observer performance. |
Luminance stability | Performance data on luminance stability reviewed. Implied to meet criteria or not affect observer performance. |
Maximum allowed pixel defects/faults (agreed with panel manufacturer) | Met criteria (Implied, as the device was deemed substantially equivalent). |
Note: The document explicitly states: "The test results showed that the RadiForce RX840-MG has display characteristics equivalent to those of the predicate device, RadiForce GX530 except 4 items, each of which was determined that it would not affect observer's performance. Besides, the display characteristics of the RadiForce RX840-MG meet the pre-defined criteria when criteria are set." The specific 4 items and their exact deviation are not detailed, but the overall conclusion is that any differences do not negatively impact performance or safety.
Study Details
-
Sample Size used for the test set and the data provenance:
- The study involved bench testing of the RadiForce RX840-MG display monitor.
- The "test set" in this context refers to the physical device itself and its components, not a dataset of medical images or patient data.
- Data provenance: Not directly applicable as this is a device performance test, not a clinical study with patient data. The device itself is manufactured by EIZO NANAO CORPORATION in Japan.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This was a bench test of a display monitor, not a study requiring expert readers to establish ground truth for medical diagnoses.
- The "ground truth" was established by objective measurements against established technical guidelines (AAPM Task Group 18, DICOM GSDF) and pre-defined criteria. There is no mention of human experts being used to establish ground truth for the test set itself.
-
Adjudication method for the test set:
- Not applicable. As a bench test of a display monitor, there was no adjudication of discrepant readings or interpretations. The performance was measured directly.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a MRMC comparative effectiveness study was not done.
- This submission is for a medical display monitor, not an AI-powered diagnostic tool. The purpose was to show the display's technical performance characteristics.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, the testing was effectively a "standalone" performance evaluation of the monitor itself, without any human interaction influencing the measurement of its technical specifications. However, it's not an "algorithm-only" performance as it's a hardware device. The device's performance was evaluated independently against technical standards.
-
The type of ground truth used:
- Technical Benchmarks and Standards: The ground truth was based on established quality control guidelines and standards for medical displays, specifically:
- DICOM GSDF (Grayscale Standard Display Function)
- AAPM Task Group 18 (TG18 guideline)
- Pre-defined criteria set by the manufacturer for various display characteristics.
- Technical Benchmarks and Standards: The ground truth was based on established quality control guidelines and standards for medical displays, specifically:
-
The sample size for the training set:
- Not applicable. This is a bench test for a hardware device, not a machine learning model that requires a training set. The device itself is manufactured, not "trained."
-
How the ground truth for the training set was established:
- Not applicable, as there is no training set for a hardware display monitor.
Ask a specific question about this device
Page 1 of 1