510(k) Data Aggregation

    K Number
    K113562
    Date Cleared
    2012-07-16

    (228 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    3MP COLOR LCD MONITOR, RADIFORCE RX340

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RadiForce RX340 is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The device does not support the display of mammography images for diagnosis.

    Device Description

    The RadiForce RX340 is a color LCD monitor for viewing medical images other than those of mammography. The matrix size (or resolution) of the panel is 1,536 x 2,048 pixels (3MP) with a pixel pitch of 0.21075 mm. Since factory calibrated display modes, each of which is characterized by a specific tone curve (including DICOM GSDF), a specific luminance range and a specific color temperature, are stored in lookup tables within the monitor, the tone curve is e.g. DICOM compliant regardless of the display controller used. RadiCS is application software to be installed in each workstation offering worry-free quality control of the diagnostic monitors including RX340 based on several QC guidelines. The RadiCS and its subset, RadiCS LE are included in this 510(k) submission as an accessory to the RadiForce RX340.

    AI/ML Overview

    The provided document is a 510(k) summary for the EIZO RadiForce RX340 3MP Color LCD Monitor. It does not contain information about a study proving the device meets specific acceptance criteria in the way a clinical trial for an AI-powered diagnostic device would.

    This document describes a medical display monitor and its substantial equivalence to a predicate device, not an algorithm that generates a diagnosis or interpretation. Therefore, many of the requested items (e.g., sample size for test set, number of experts for ground truth, adjudication method, MRMC study, standalone performance, training set size) are not applicable to this type of device submission.

    Instead, the submission focuses on technical characteristics and performance benchmarks compared to a predicate device to demonstrate substantial equivalence.

    Here's an analysis based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't present a formal table of "acceptance criteria" against which an algorithm's performance is measured. Instead, it compares the proposed device's technical specifications and performance to those of a predicate device (RadiForce RX320). The "acceptance criteria" are implicitly that the proposed device must perform at least as well as the predicate device in key display characteristics, or that any differences do not affect safety or effectiveness.

    CharacteristicPredicate Device (RadiForce RX320) PerformanceProposed Device (RadiForce RX340) PerformanceMeets Implied Acceptance Criteria?
    Panel Matrix Size/Resolution1536 x 2048 pixels (3MP)1536 x 2048 pixels (3MP)Yes (Same)
    Active Area Size323.7 mm x 431.6 mm323.7 mm x 431.6 mmYes (Same)
    Pixel Pitch(Not specified for predicate, assumed same)0.21075 mmYes (Same, as panel is same)
    DICOM Calibrated Luminance400 cd/m²400 cd/m²Yes (Same)
    Typical Maximum Luminance900 cd/m²1000 cd/m²Yes (Higher, which is considered an improvement)
    Backlight TypeCCFLLEDYes (Improved, mercury-free, less power, slower deterioration)
    DICOM GSDF ComplianceYes (via factory calibrated modes)Yes (via factory calibrated modes)Yes (Same)
    Supported Video InputsDigital Visual Interface (DVI)DVI, DisplayPortYes (Adds a new standard without affecting quality)
    QC SoftwareRadiCS / RadiCS LERadiCS / RadiCS LEYes (Same)
    Backlight Sensor (BS)PresentPresentYes (Same functionality)
    Built-in Front Sensor (IFS)Not explicitly mentioned but implies absencePresentYes (Enhance, allows auto grayscale calibration)
    Presence Sensor (PS)Not mentionedPresentNot directly comparable, but an added feature for power saving
    Ambient Light Sensor (ALS)Not mentionedPresentNot directly comparable, but an added feature for ambient light measurement
    International Safety/EMC Standards Compliance(Implied for predicate)Validated by independent and in-house facilitiesYes (Validated)

    2. Sample Size for the Test Set and Data Provenance

    Not applicable in the context of an algorithm's test set. The "test set" here refers to the actual monitor units evaluated. The document states: "EIZO NANAO CORPORATION performed a range of system and performance tests to ensure that the RadiForce RX340 performs in accordance with its specifications." It does not specify the number of units tested, but it is typical for manufacturers to test a representative sample during design validation. Data provenance is internal testing by the manufacturer.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable. "Ground truth" in this context would relate to the physical characteristics of the display itself (e.g., luminance, resolution, color accuracy), which are measured using calibrated instruments, not expert human assessment.

    4. Adjudication Method for the Test Set

    Not applicable. Measurements are objective and do not require expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. This type of study is relevant for comparing the diagnostic performance of human readers with and without AI assistance for interpreting medical images. The RadiForce RX340 is a display monitor, not an AI diagnostic tool, so an MRMC study is not relevant.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a monitor, not an algorithm.

    7. The Type of Ground Truth Used

    The ground truth used for validating the monitor's performance is based on objective physical measurements of its display characteristics (e.g., luminance, color temperature, resolution, DICOM GSDF compliance) using calibrated testing equipment. It also includes compliance with international safety and EMC standards.

    8. The Sample Size for the Training Set

    Not applicable. This is a hardware device (monitor), not a machine learning algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a monitor.