Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K972113
    Date Cleared
    1997-09-02

    (89 days)

    Product Code
    Regulation Number
    870.3535
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    30 CC AND 40 CC SUB-9 NITINOL AND 30 AND 40 CC SUB-9 STAINLESS STEEL

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BSC/CA IABC's are indicated for use in patients with the following conditions: Refractory power failure. Cardiogenic shock. Unstable refractory angina. Histodynamically significant mechanical complications secondary to acute MI: Ventricular septal defect. Mitral valve regurgitation. Papillary muscle rupture. Cardiac support for high risk general surgical and coronary angiography/ angioplasty patients. Septic shock. The intended use of the 30 cc and 40 cc SUB-9 remain identical to that of the currently marketed IABC's. All of the devices are intended to provide temporary circulatory ntireded that 3. The or in athrough controlled mechanical displacement of a volume of blood in the aorta. The mechanical action of the IAB catheter therapy lowers the cardiac workload by two reans : Systolic unloading, as noted by a reduction in the patient's systolic pressure, l . which provides reduced myocardial oxygen consumptions (MVO,). Diastolic augmentation which provides an increase in the mean ann aortic pressure and leads to an improvement in systemic and coronary arterial perfusion. Balloon pump therapy is achieved by inserting an intra-aortic balloon catheter into the descending thoracic aorta via the common femoral artery. Balloon inflation is timed to occur during diastole, beginning with the aortic valve closure. The balloon remains inflated until the onser of left ventricular ejection or systole, then rapidly deflates, reducing the aortic pressure, which in turn reduces the afterload.

    Device Description

    The proposed 30 cc and 40 cc SUB-9 Stainless Steel and Nitinol consists of a polyurethane blend symmetrical balloon at the distal end of a polyurethane-covered nylon shaft. The balloons are coated with a thin layer of silicone fluid. A Nitinol central lumen runs throughout the length of the 30 cc and 40 cc SUB-9 Nitinol catheters and terminates at the distal tip. A Stainless Steel central lumen runs throughout the length of the 30 cc and 40 cc SUB-9 Stainless Steel catheters and terminates at the distal Each central lumen may be used to pass the devices over their tip. guidewire. The balloon is supplied prewrapped for insertion utilizing either a hemostasis sheath (provided in each IAB insertion kit) or sheathless.

    AI/ML Overview

    The provided 510(k) summary for the Boston Scientific Corporation/Cardiac Assist (BSC/CA) SUB-9 Intra-Aortic Balloon Catheter describes the device and its substantial equivalence to predicate devices. However, this document does not contain the detailed acceptance criteria, specific device performance metrics, or a formal study report with the requested sample sizes, ground truth establishment, or multi-reader multi-case study information that would typically be found in a comprehensive clinical or non-clinical study report.

    The document focuses on demonstrating substantial equivalence through a series of non-clinical tests comparing the new SUB-9 device to its predicate devices. The "acceptance criteria" in this context are implicitly that the SUB-9 design performs substantially equivalently to, or better than, the predicate devices in various mechanical and functional aspects.

    Here's an attempt to extract and infer the information based on the provided text, with clear indications where information is not available in the document:


    Acceptance Criteria and Device Performance for K912113 SUB-9 Intra-Aortic Balloon Catheter

    This 510(k) summary focuses on demonstrating substantial equivalence of the SUB-9 Intra-Aortic Balloon Catheter to its predicate devices through non-clinical testing. The "acceptance criteria" are implicitly met when the SUB-9 device performs comparably to, or better than, the predicate devices in the described tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test CategoryImplicit Acceptance CriteriaReported Device Performance and Remarks
    1. Sheathed Insertion TestForce required for insertion is substantially equivalent to predicate devices.The force required to insert the SUB-9 designs (30 and 40 cc, Stainless Steel and Nitinol) was demonstrated to be substantially equivalent to the force required to insert the predicate NICATH™ and Sensation™ designs.
    2. Sheathless Insertion TestAbility to pass through restrictions comparable to, or better than, predicate devices.The tightest restriction that the SUB-9 design can pass through was found to be smaller than that required for the predicate BSC/CA IAB's, the Datascope 40 cc 9.5 F Percor-Stat, and the Kontron/Arrow 40 cc 9 F IAB. (SUB-9 performed better)
    3. Maximum Pumping Rate Limit TestMaximum pumping rate at which the balloon fully inflates/deflates is comparable to, or better than, predicate devices.The maximum achievable pumping rate limits for the SUB-9 designs were found to be higher than that for the predicate NICATH™ and Models 940 and 930 on the BSC/CA 3001. It is inferred that SUB-9 will perform faster on any IABP it is labeled for use on. (SUB-9 performed better).
    4. Reliability/Integrity TestReliability and integrity comparable to predicate devices (ability to cycle at least 3.6 million cycles).The reliability of the SUB-9 was found comparable to all BSC/CA predicate devices, all of which were able to cycle for at least 3.6 million cycles. Maximum pumping rate limit post-reliability was identical to pre-reliability rate for 40cc SUB-9. Inflation and deflation times post-reliability were comparable to pre-reliability times.
    5. Aneurization and Burst Pressure TestAneurization and burst pressures are substantially equivalent to predicate devices, and external IABP output pressures are below these limits.Demonstrated substantially equivalent aneurization and burst pressures as compared to the predicate devices. The SUB-9 balloons will not aneurize or burst when used on labeled IABP's, as IABP's cannot provide the volume/pressure required. Peak output pressures of all indicated IABP's are below the pressures required to aneurize and burst the balloons.
    6. Transmembrane Pressure and Volume Measurement TestTransmembrane pressure at a given inflated volume is substantially equivalent to predicate devices.Since the balloons on SUB-9 are the same as predicate Models 940 and 930 and 30/40 cc NICATH™, they will have the same transmembrane pressure at the same inflated volume. The volume of the 30 and 40 cc SUB-9 were found substantially equivalent to the predicate Models 940 and 930, demonstrating equal transmembrane pressures.
    7. Trackability TestAbility to track over guidewire into position in aorta without incident.All SUB-9 test samples (30 cc and 40 cc; Stainless Steel and Nitinol) were able to track their guidewire into position in the aorta without incident. No catheter kinks, guidewire kinks, or problems inserting/removing were encountered.
    8. Kink Resistance TestMean kink radius substantially equivalent to predicate devices.The mean kink radius of the SUB-9 Stainless Steel and Nitinol were demonstrated to be substantially equivalent to the predicate NICATH™, Sensation™, and Models 940 and 930.
    9. Initial Performance, Insertion/Removal Integrity TestProper opening per Directions for Use, and integrity maintained post-insertion/removal.All 30 cc and 40 cc SUB-9 Nitinol and Stainless Steel test samples opened as they should per the Directions for Use. The integrity of the SUB-9 design post insertion and removal was demonstrated.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document repeatedly refers to "the SUB-9 designs (30 and 40 cc and Stainless Steel and Nitinol)" or "SUB-9 test samples (30 cc and 40 cc; Stainless Steel and Nitinol)". However, specific numerical sample sizes (e.g., "n=X devices were tested for each configuration") are not provided.
    • Data Provenance: All tests described are "Nonclinical Tests" performed by Boston Scientific Corporation/Cardiac Assist. The origin of the data is internal laboratory testing and physical characterization. The document states, "No clinical testing was performed by Boston Scientific Corporation/Cardiac Assist in support of this premarket notification." Therefore, the data is retrospective non-clinical bench testing data, not from human subjects, and there is no country of origin for human data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    • Not applicable. The "ground truth" in this context refers to the physical and functional characteristics of the devices (e.g., force required for insertion, pumping rate, burst pressure). These are objective measurements determined by laboratory equipment and procedures, not by expert interpretation. Therefore, experts were not used to establish a ground truth in the sense of clinical interpretations.

    4. Adjudication Method for the Test Set

    • Not applicable. As the ground truth is based on objective physical measurements, there is no need for expert adjudication. The tests likely follow validated laboratory protocols.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC study was done. This device is an Intra-Aortic Balloon Catheter, which is a physical medical device, not an AI or imaging diagnostic tool that would typically involve human readers or AI assistance in interpretation. The document explicitly states: "No clinical testing was performed by Boston Scientific Corporation/Cardiac Assist in support of this premarket notification."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a hardware medical device, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" for the non-clinical tests is established by objective physical measurements and engineering specifications. For example, burst pressure is measured until the balloon bursts, pumping rate until full inflation/deflation cannot be achieved, and insertion force is quantified using mechanical testing equipment. Comparison is made against the performance of legally marketed predicate devices, which serve as the reference standard for "substantially equivalent" performance.

    8. The Sample Size for the Training Set

    • Not applicable. Since this is a hardware device and not an AI or machine learning model, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no training set for an AI/ML model, there is no ground truth established for it.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1