510(k) Data Aggregation

    K Number
    K143340
    Date Cleared
    2015-07-09

    (230 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or fingers to prevent contamination between patient and examiner.

    Device Description

    Powder-free Vinyl Patient Examination Gloves, Yellow Color. A disposable device intended for medical purposes that is worn upon the examiner's hands or finger. Prevent contamination between patient and examiner. Cover the hand and wrist area. Covers have separate sheaths or openings for each finger and the thumb.

    AI/ML Overview

    This document is a 510(k) summary for a medical device: Powder-free Vinyl Patient Examination Gloves, Yellow Color. It focuses on demonstrating substantial equivalence to a predicate device. Therefore, the "device" in this context is the glove itself, not an AI or software-based medical device.

    As a result, most of the requested information regarding AI/ML device studies (such as sample sizes for test/training sets, data provenance, number/qualifications of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and how ground truth for training was established) is not applicable to this type of medical device filing.

    However, I can extract the acceptance criteria and reported performance for the physical characteristics of the gloves based on the provided text.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally "Confirmed with ASTM D5250-06" or "Meet ASTM D5250-06" for physical properties and "Meet ASTM D6124-06" for residual powder. For biocompatibility, the criteria are "Not a sensitizer" and "Not an irritant." The reported device performance indicates that the proposed device meets these standards.

    Acceptance Criteria / Performance MetricAcceptance CriteriaReported Device Performance
    Physical Properties (per ASTM D5250-06):
    Length on Large SizeConfirmed with ASTM D5250-06Confirmed with ASTM D5250-06
    Width of Palm on Large SizeConfirmed with ASTM D5250-06Confirmed with ASTM D5250-06
    Palm ThicknessConfirmed with ASTM D5250-06Confirmed with ASTM D5250-06
    Fingertip ThicknessConfirmed with ASTM D5250-06Confirmed with ASTM D5250-06
    Pinhole ResultsMeet ASTM D5250-06Meet ASTM-D-5250-06
    Before Aging: Tensile Strength(Mpa) and Ultimate ElongationsMeet ASTM D5250-06Meet ASTM D5250-06
    After Aging: Tensile Strength(Mpa) and Ultimate ElongationsMeet ASTM D5250-06Meet ASTM D5250-06
    Other Non-Clinical Results:
    Residual PowderMeet ASTM D6124-06 (or